Comparison of a portable, pneumotach flow-sensor–based spirometer (Spirofy™) with the vitalograph alpha Touch™ spirometer in evaluating lung function in healthy individuals, asthmatics, and COPD patients—a randomized, crossover study

Abstract

Background

Spirofy™ is India’s first portable, pneumotach flow-sensor-based digital spirometer developed to diagnose asthma and chronic obstructive pulmonary disease (COPD). In this study, we compared the performance of the Spirofy™ device with that of the Vitalograph Alpha Touch™ spirometer in measuring the lung capacities of healthy individuals, asthmatics, and COPD patients. We also assessed the inter-device variability between two Spirofy™ devices.

Methods

In a randomized, three-way crossover, open-label study, we measured the differences in forced expiratory volume in the first second (FEV1) and forced vital capacity (FVC) between the Spirofy™ and Vitalograph Alpha Touch™ spirometers. A proportion of the FEV1/FVC ratio distribution of < 0.7 was used to compare the diagnostic accuracies of the Spirofy™ with Vitalograph™ Alpha Touch™ spirometers.

Results

Ninety subjects participated in this study. The mean ± SD FVC values obtained from the Spirofy™ 1, Spirofy™ 2, and Vitalograph Alpha Touch™ devices were 2.60 ± 1.05 L, 2.64 ± 1.04 L, and 2.67 ± 1.04 L, respectively. The mean ± SD FEV1 values obtained from the Spirofy™ 1, Spirofy™ 2, and Vitalograph Alpha Touch™ devices were 1.87 ± 0.92 (L), 1.88 ± 0.92 (L), and 1.93 ± 0.93 (L), respectively. A significant positive correlation was found between the FVC and FEV1 values recorded by Vitalograph Alpha Touch™, Spirofy™ 1, and Spirofy™ 2. As compared to Vitalograph Alpha Touch™, the Spirofy™ device showed good sensitivity (97%), specificity (90%), and overall accuracy (93.3%) at an FEV1/FVC ratio < 0.7. No inter-device variability was observed between the two Spirofy™ devices.

Conclusion

Spirofy™ is a portable and easy-to-use device and is as accurate as the standard Vitalograph Alpha Touch™ spirometer for the diagnosis of COPD and asthma.

Trial registration

CTRI/2021/09/036492 (Clinical Trials Registry - India)

Peer Review reports

Introduction

Globally, chronic respiratory diseases are the most common causes of morbidity and mortality [1] and lead to significant healthcare expenses [2]. Obstructive airway diseases (OAD), which include chronic obstructive pulmonary disease (COPD) and asthma, are a subset of chronic respiratory diseases that share common features [1]. Globally, asthma is the 16th cause of years-lived-with-disability and the 28th most common source of burden of disease [3]. In 2019, COPD was the 6th major cause of years-lived-with-disability for all ages and the third leading cause of death [4]. There are an estimated 50 million COPD cases and 34 million asthma cases in India and COPD is the second leading cause of death and disability-adjusted life years in India [57].

The gold standard test for the diagnosis of obstructive airway diseases is spirometry [8]. It is primarily used in patients with asthma and COPD [9]. It is a safe, practical, and reproducible test to evaluate lung function in primary care settings [10]. Under-use of spirometry is the most important cause of under-diagnosis of asthma and COPD; this occurs because of several barriers associated with its use in the primary care setup, viz., high cost, challenges in performing the test, and the extended time required to perform the test. It is believed that specialized training in performing the spirometry procedure and interpreting its results as well as improving the affordability and accessibility of the device could enhance its use [11, 12].

Cipla Limited has developed a new, digital point-of-care spirometer (Spirofy™), which is portable, easy to use, and based on pneumotach flow-sensing technology. Spirofy™ has undergone laboratory validation under different testing conditions. A randomized, open-label, three-way, crossover study was conducted to validate the performances of two Spirofy™ devices (Cipla, pneumotach sensor-based spirometer). The main objective of this study was to compare the performances of two Spirofy™ devices (device 1 and 2) with that of the Vitalograph Alpha Touch™ spirometer in measuring the lung capacities of healthy individuals, asthmatics, and patients with COPD. The secondary objective of this study was to assess the inter-device variability between the Spirofy™ devices.

Methodology

Study design

In this randomized, open-label, three-way crossover study, we compared the performance of two Spirofy™ devices to that of the reference Vitalograph Alpha Touch™ standard spirometer (pneumotach sensor-based spirometer) in measuring the lung capacities of healthy subjects, asthmatics, and COPD patients. Vitalograph Alpha Touch™ was chosen as the reference standard for evaluating Spirofy because it is widely used and is based on pneumotach technology, which is also utilized in the Spirofy™ device. The study was conducted over a period spanning from October 2021 to February 2022.

Study population

The study population was comprised of randomly selected healthy subjects, asthmatics, and COPD patients visiting three respiratory clinics in India. The Independent Ethics Committee/Institutional Review Board (IEC/IRB) approved the informed consent, study protocol, and subject information sheets before the initiation of the study. A total of 90 subjects were included, comprising 29 healthy individuals, 31 asthmatics, and 30 COPD patients. The study was done as per the approved protocol requirement. The study protocol can be made available on request with justification. Requests for access should be directed to the corresponding author.

Sample size

The sample size was calculated assuming an intra-subject standard deviation of 150 ml with a mean difference of 45 ml for FEV1 measured by 2 spirometers, with a power of 80% and 5% level of significance. Based on the calculations, a total of 90 subjects were required.

Inclusion criteria

Healthy subjects, stable asthma patients, or stable COPD patients aged ≥ 18 years with medical history/records showing evidence of diagnosed asthma or COPD, no history of disease exacerbations in the preceding 4 weeks, and no other active or chronic respiratory complaints were included in the study. The other criteria for inclusion for healthy individuals were a declaration of normalcy after a clinical examination and a negative history of respiratory tract infection in the previous 4 weeks. The subjects had to test negative for the severe acute respiratory syndrome coronavirus (SARS-CoV-2) within the last 24–48 h or as per the respiratory clinic’s policy.

Exclusion criteria

Patients with lung/airway diseases other than asthma or COPD were excluded from the study. Patients with (a) contraindications for spirometry as per the American Thoracic Society/European Respiratory Society (ATS/ERS) 2019 guidelines; (b) any significant deformity of the thorax or vertebral column that may cause persistent airflow limitations or altered lung volumes; (c) uncontrolled or unstable asthma/COPD status; and (d) severe concomitant diseases such as cardiovascular, pulmonary, abdominal, neurological, or endocrine disorders were also excluded. Pregnant women were also excluded from this study.

Study devices

The devices used in the study were: (1) Two Spirofy™ pneumotach sensor-based spirometers (Spirofy™ 1 and 2) manufactured by Cipla Ltd., India as test medical devices (Figs. 1) and (2) a Vitalograph Alpha Touch™ spirometer (pneumotach sensor-based spirometer) manufactured by Vitalograph Inc., USA as a standard for comparison.

Fig. 1
figure 1

Image of Spirofy™ pneumotach sensor-based spirometer used in the study

Screening and spirometry

Each subject underwent one screening and one study visit. Both visits were conducted on the same day on which the subject tested negative for the SARS-CoV-2 test or within 24–48 h (or as per site policy) of the negative result and only if the subject was willing to participate in the study. After the screening, informed consent was obtained, and the demographic characteristics and routine history of the individual were recorded (Fig. 2).

Fig. 2
figure 2

The schematic representing the study protocol. PIS: Patient information sheet; QA: Quality assurance; RT-PCR: Real-time-polymerase chain reaction; SARS-CoV-2: Severe acute respiratory syndrome coronavirus 2

Three spirometry tests were performed on the participants (one test each with the two Spirofy™ spirometers of Cipla Ltd. and one with the Vitalograph Alpha Touch™ spirometer) in a pre-determined, randomized sequence generated for each of the three devices. Due to the random allocations of the subjects to all three device sequences, the study provided unbiased estimates for the differences in measured values between the devices. The cross-over design of the study allowed each subject to serve as his/her control to minimize the influence of potential confounders such as inter-subject variability. Details regarding clinical information and results from index/reference standard tests were available to the personnel administering the tests.

Spirometry was performed according to the ATS/ERS 2019 guidelines [13]. The study protocol included specific measures to control for the use of bronchodilators among participants on the day of spirometry testing. Prior to conducting spirometry, all participants were instructed to withhold any ongoing respiratory medication for a period consistent with the defined washout periods recommended by the ATS/ERS 2019 guidelines. Three to eight spirometry maneuvers were recorded with each spirometer to ensure that the subject had performed three acceptable and two reproducible maneuvers. Participants were requested to rest for at least 10 min between two spirometry tests, with the rest period not exceeding 1 h. Three sets of spirometry reports were generated for each subject on the same day during the same session of the day. The highest values obtained by the participants from each of the three acceptable and reproducible spirometry reports were used for further analysis.

Primary endpoints

The primary endpoints were the differences in forced vital capacity (ΔFVC) and the differences in forced expiratory volume in the first second (ΔFEV1) between the Spirofy™ and Vitalograph Alpha Touch™ spirometers. The co-primary endpoint was the difference in FEV1/FVC ratio of < 0.7 between the Spirofy™ and Vitalograph Alpha Touch™ spirometers. The distribution of the FEV1/FVC ratio < 0.7 was used to assess the diagnostic accuracies of the Spirofy™ devices by comparing their performance with that of the Vitalograph in being able to differentiate subjects with OAD from healthy individuals; the results have been presented as the sensitivity, specificity, and overall accuracy of each device.

Secondary endpoints

The secondary endpoints were ΔFVC and ΔFEV1 between the Spirofy™ devices 1 and 2 to test the inter-device variability.

Statistical methods

Data on the baseline characteristics and other test parameters were summarized using descriptive statistics. Continuous variables are described using mean ± SD and minimum and maximum values; categorical variables are described using percentages and frequencies. The Kolmogorov–Smirnov (K–S) test was used to check if the data were normally distributed before data analysis was performed. The mean differences in spirometry measures between the Spirofy™ and Vitalograph™ Alpha Touch devices were analyzed with the Wilcoxon test or paired sample t-test depending upon the distribution of the data. The correlation coefficient was calculated to test if the measurements of both devices correlated with each other.

The coefficient of variation between two measurements obtained by the Spirofy™ was used to evaluate reproducibility (coefficient of variation = SD/mean) and the 95% limits of agreement (1.96*SD). The limits of agreement were calculated as mean difference ± 1.96*SD and Bland–Altman plots were created. The Bland–Altman method was used to validate if any systematic differences were present in the flow profile measurements taken from Spirofy™ 1 and 2. For comparison, the acceptable limits of agreement were considered as 0.50 L for FVC and 0.35 L for FEV1. There were no indeterminate test results or missing data from index or reference standard tests. No exploratory analysis was carried out. All statistical analyses were carried out using the Statistical Packages for Social Science (SPSS) version 27.0.

Results

Study population distribution

The study enrolled 102 subjects randomly from a total of 107 subjects, with 35 healthy subjects, 34 subjects with asthma, and 33 subjects with COPD; these 107 subjects constituted the intention-to-treat (ITT) population. The enrolled subjects were randomized to device sequences 123, 231, or 312, where device 1 represented Spirofy™ 1, device 2 represented Spirofy™ 2, and device 3 represented the Vitalograph Alpha Touch™. Ninety subjects with acceptable spirometry test results, per protocol (PP) population, completed the study; these included 29 healthy subjects, 31 subjects with asthma, and 30 subjects with COPD (Fig. 3). The distribution of severity of the target condition is provided as supplementary information (SI-1). There were no alternate diagnoses in subjects without the target condition.

Fig. 3
figure 3

Consolidated Standards of Reporting Trials (CONSORT) diagram depicting study population distribution. COPD: Chronic obstructive pulmonary disease; ITT: Intention to treat; PP: Per protocol

Study population demographic characteristics

Among the 102 Indian subjects enrolled in the study and analyzed for the ITT population, 65.7% were men. The subjects had a mean age (± SD) of 49.8 (± 17.0) years. Of the 90 subjects analyzed for the PP population, 63% were men and the subjects had an average age (± SD) of 50.4 (± 16.7) years. Amongst the enrolled subjects, 35.6% reported chronic respiratory symptoms, 11% had a history of allergic rhinitis, and 24% had been exposed to tobacco smoke in the past. For subjects with asthma, the mean (± SD) disease duration was 4.19 (± 3.36) years; for subjects with COPD, this value was 6.04 (± 4.37) years (Table 1).

Table 1 Demographic characteristics of intention-to-treat and per-protocol population

Medication use among the study population

The 31 asthma patients who completed the study were on inhaled corticosteroid (ICS)/long-acting beta-agonist (LABA) (28), ICS/LABA/long-acting muscarinic antagonist (LAMA) (1),xanthine derivatives (6), leukotriene receptor antagonist (LTRA) (6), antihistamine (5), short-acting muscarinic antagonist (SAMA)/short-acting beta-agonist (SABA) (1), oral corticosteroids (2), and N-acetylcysteine (1). The 30 COPD patients were on ICS/LABA (9), ICS/LABA/LAMA (14), LABA/LAMA (5), LAMA (1), ICS/LABA + LABA/LAMA (2), xanthine derivatives (18), LTRA (4), SAMA/SABA (1), oral corticosteroids (2), N-acetylcysteine (1). A list of all medications used by the asthma and COPD patient groups is provided in the supplementary information (SI-2).

Primary endpoints

The mean ± SD FVC values reported for the Spirofy™ 1, Spirofy™ 2, and Vitalograph Alpha Touch™ were 2.60 ± 1.05, 2.64 ± 1.04, and 2.67 ± 1.04 L, respectively (Table 2). The mean differences in FVC values between the Vitalograph Alpha Touch™ and Spirofy™ 1 were 71.44, 75.17, 27.74, and 113 mL for overall, healthy, asthma, and COPD groups, respectively (Table 3). The mean differences in FVC values between the Vitalograph Alpha Touch™ and Spirofy™ 2 for the overall, healthy, asthma, and COPD groups were 35.89, 72.76, -15.16, and 53 mL, respectively (Table 3).

Table 2 Mean values of FVC and FEV1 in the per-protocol population as measured by the Spirofy™ 1, Spirofy™ 2, and Vitalograph Alpha Touch™ devices
Table 3 The comparative limits of agreement and correlation coefficients for FVC and FEV1 in the per-protocol population measured by Spirofy™ 1, Spirofy™ 2, and Vitalograph Alpha Touch™ devices

Strong positive correlations in the FEV1 values measured by the Vitalograph Alpha Touch™, Spirofy™ 1, and Spirofy™ 2 devices for the overall, healthy, asthma, and COPD groups were detected (Table 3). The mean ± SD FEV1 values reported for the Spirofy™ 1, Spirofy™ 2, and Vitalograph Alpha Touch™ were 1.87 ± 0.92, 1.88 ± 0.92, and 1.93 ± 0.93 L, respectively. The mean differences in FEV1 values between Vitalograph Alpha Touch™ and Spirofy™ 1 were 65.44 mL, 83.10 mL, 58.39 mL, and 55.67 mL for overall, healthy, asthma, and COPD groups, respectively (Table 3). The mean difference in FEV1 values between the Vitalograph Alpha Touch™ and Spirofy™ 2 for the overall, healthy, asthma, and COPD groups were 46.67, 72.76, 31.61, and 37.0 mL, respectively. (Table 3).

Comparison of limits of agreement

The FVC values measured by Spirofy™ 1 and Spirofy™ 2 were within the pre-defined limits of agreement of those measured by the Vitalograph Alpha Touch™. The 95% FVC values for Spirofy™ 1 and Vitalograph Alpha Touch™ and those for Spirofy™ 2 and Vitalograph Alpha Touch™ were within the ± 500 mL range [14]. For the differences between the FEV1 values measured by Spirofy™ 1 and Vitalograph Alpha Touch™, 92.2% of values were within the predefined limits of ± 350 mL; in the case of Spirofy™ 2 and Vitalograph Alpha Touch™, 91.1% of these values were within this limit (Figs. 4a and b and 5a and b) [14].

Fig. 4
figure 4

Bland-Altman plot for FVC overall population a)Vitalograph™ vs. Spirofy™ 1 and b) Vitalograph™ vs.Spirofy™ 2 COPD: Chronic obstructive pulmonary disease; FVC: Forced vital capacity

Fig. 5
figure 5

Bland-Altman plot for FEV1 overall population (a) Vitalograph™ vs. Spirofy™ 1 and (b) Vitalograph™ vs. Spirofy™ 2. COPD: Chronic obstructive pulmonary disease; FEV1: Forced expiratory volume in one second

Co-primary endpoint

Diagnostic validation of Spirofy™ 1 using the Vitalograph Alpha Touch™ as a reference at an FEV1/FVC ratio < 0.7 showed that Spirofy™ 1 had good sensitivity (97.2%), specificity (90.1%), and overall accuracy (93.3%) in distinguishing airway obstructions from non-airway obstructions in the PP population (which included the healthy, asthma, and COPD groups). Similarly, Spirofy™ 2 also had good scores in sensitivity (97.2%), specificity (88.9%), and overall accuracy (92.2%) (Table 4).

Table 4 Distribution of FEV1/FVC < 0.7 between Spirofy™ 1 and Vitalograph Alpha Touch™ and Spirofy™ 2 and Vitalograph Alpha Touch™ in the overall group of the per-protocol population

Secondary endpoints

FVC

All FVC (100%) values measured by Spirofy™ 1 and Spirofy™ 2 were within the pre-defined limits of agreement with the values of ΔFVC between Spirofy™ 1 and Spirofy™ 2 being within the ± 500 mL range.

FEV1

For the overall group, the mean difference in FEV1 values between Spirofy™ 1 and Spirofy™ 2 was 18.8 mL, with the 95% limits of agreement ranging from − 237.2 to 199.6 mL. A strong positive correlation was observed in the FEV1 values measured between the Spirofy™ 1 and Spirofy™ 2 devices in the overall group. For the differences between the FEV1 values measured by Spirofy™ 1 and Spirofy™ 2, 98.8% of values were within the pre-defined limits of ± 350 mL.

Safety findings

No adverse events were reported from performing the index test or the reference standard test.

Discussion

Despite being common, obstructive airway diseases such as COPD and asthma are frequently underdiagnosed in clinical practice [15]. Previous studies have reported underdiagnosed asthma in 19–73% of individuals [16]. According to the Global Asthma Network (GAN) study conducted across multiple centers in urban parts of India, 75–82% of patients who had wheezing and 68–70% of subjects with severe asthma had not been clinically diagnosed with asthma [17]. The Global Initiative for Chronic Obstructive Lung Disease summit on COPD estimated that 95–98% of COPD cases are underdiagnosed in India as symptomatic patients seek healthcare advice late due to a lack of awareness; in addition, most COPD cases are diagnosed based on symptoms rather than spirometry [18].

Spirometry is an important and highly recommended test for the diagnosis of OAD [12]. It also helps in evaluating the severity of the disease and monitoring airflow obstruction [8]. Spirometry measurements obtained in childhood, adolescence, and adulthood can identify individuals at risk for non-communicable respiratory and non-respiratory diseases [19]. Despite these benefits, spirometry is still underutilized in clinical settings [12]. The main causes of this underutilization include the lack of precise, portable, and reasonably priced spirometers, particularly in primary care settings; difficulties in maintaining such equipment; and difficulties in conducting and interpreting spirometry tests [12, 14]. Most modern spirometers that are manufactured and sold globally are accurate and several are portable; however, they are expensive [2022]. To address the rising burden of OAD in India, an indigenously manufactured, easy to use, accurate, inexpensive, and portable spirometer is critical.

In this study, we examined the clinical validity and accuracy of spirometry measurements obtained using the newly developed Spirofy™ spirometer, and compared it with the traditional Vitalograph Alpha Touch™ spirometer. Spirofy™ is a portable, pneumotach flow sensor-based spirometer with software designed with the capability for continuous improvements. The process to upgrade the device with the latest ATS/ERS algorithm has been initiated and will be completed within a year. A total of 90 participants were enrolled, including healthy individuals as well as people with asthma and COPD. The spirometry tests were performed at random with two Spirofy™ test devices (Spirofy™ 1 and Spirofy™ 2) and one Vitalograph Alpha Touch™ comparator device. This study demonstrates that the primary endpoint was attained by both test devices (Spirofy™ 1 and Spirofy™ 2), with > 95% of FVC and > 91% of FEV1 values falling within the acceptable limits of agreement. The limits of agreement for FVC and FEV1 values (±500 mL and ±350 mL) have been described previously for comparing two different types of spirometers [14]. This result is consistent with the findings of a case–control study that evaluated the accuracy of the Vitalograph® lung monitor and compared it to post-bronchodilator confirmatory spirometry tests used to screen COPD patients; the study revealed that the FEV1 values from the two devices were strongly correlated (r = 0.97) [22]. Another study that compared a portable spirometer (AioCare®) with a reference desktop spirometer demonstrated good correlation between the two devices in measuring peak expiratory flow readings, FEV1, FVC, and FEV1/FVC in patients with asthma or COPD [23]. A meta-analysis of 31 studies has shown that portable spirometers are highly accurate in diagnosing COPD [24]. Among the three most widely used portable spirometers, PIKO-6 was the most accurate, followed by COPD-6 and peak expiratory flow [24]. Conversely, portable turbine spirometers (CareFusion, Yorba Linda) and desktop laboratory spirometers (Vmax Encore System) were found to be uninterchangeable for use in children for diagnosing/monitoring cystic fibrosis owing to repeatability discrepancies in FEV1 values [25]. Furthermore, as Carpenter et al. (2018) have pointed out in their review, differences exist in the accuracy and quality of the assessment of lung function with portable spirometers, and not all are equally accurate [26].

Portable spirometers are utilized for lung function measurements in community settings and for tracking lung function over time in epidemiological studies. However, in large multicenter studies, different spirometers may be used across study sites and old models may need to be substituted with newer ones while a study is being conducted. This makes it necessary that the variability between different portable spirometers be determined so that the data collected across different devices are comparable [27]. In our study, all FVC values recorded by the Spirofy™ 1 and Spirofy™ 2 were within the pre-defined levels of agreement with the ΔFVC values lying within the ± 500 mL range. For the ΔFEV1 values measured by Spirofy™ 1 and Spirofy™ 2, 98.8% of the values were within the predefined limits of ± 350 mL. This indicates that both test devices meet the secondary endpoint and that there is no device-specific heterogeneity between the two test devices.

Another important objective of this study was to assess whether the Spirofy™ test devices were as good as the Vitalograph Alpha Touch™ in their ability to identify airway obstructions. Both the Spirofy™ devices exhibited high sensitivity (97%), specificity (about 90%), and overall accuracy ( 93.3%) in the diagnosis of and differentiation between different types of airway obstructions in obstructive pulmonary diseases. From the sensitivity and specificity exhibited by the Spirofy™ devices, it can be inferred that these devices successfully identified airflow obstruction in 97% of cases and failed to do so in 3% of the cases using a fixed cut-off value of 0.7 for the FEV1/FVC ratio. Compared to the Vitalograph Alpha Touch™, the Spirofy™ devices over-identified airway obstruction in approximately 10% of cases. These findings suggest that the Spirofy™ device is as effective as the Vitalograph Alpha Touch™ device in diagnosing airway obstructions in clinical settings. A similar efficacy has been observed in another study involving spirometers using an FEV1/FVC ratio < 0.7 in detecting obstruction; in this study, the Air Smart Spirometer was found to have a sensitivity of 90.4% and a specificity of 97.2% [15].

For people with asthma and COPD, the FVC values measured by Spirofy™ 1 and Spirofy™ 2 were within the pre-defined limits of agreement with those measured by the Vitalograph Alpha Touch™. Similarly, the FEV1 values measured by all three devices for people with COPD and asthma were also within the 95% limits of agreement. Therefore, the Spirofy™ was as capable as the Vitalograph Alpha Touch™ in identifying people with asthma and COPD.

Overall, the Spirofy™ is a portable, easy-to-use, pneumotach sensor-based spirometer that is as capable as the conventional Vitalograph Alpha Touch™ spirometer in detecting airway obstruction in patients with asthma and COPD. This device, which provides reliable and reproducible results, can also be a useful and effective diagnostic tool for OAD when combined with clinical history and examination results. This may improve accessibility for respiratory disease diagnoses in primary care settings.

The main strength of this study was its open-label, randomized, cross-over design. The open-label design was deemed suitable because the spirometry variables were objectively measured results that are unlikely to be influenced by subject or investigator bias. The order of use of the spirometers was randomized to ensure that the usage sequences of the devices were balanced within the study population. This study provides the most accurate and unbiased estimations for the differences in performance across devices by randomly assigning the subjects to each of the three device sequences. The cross-over design, in which each subject acts as a control for his/her tests, reduced the effects of potential confounders like intra-subject variability.

While the study provides valuable insights into the performance and clinical utility of Spirofy™, there are some limitations that need to be taken into consideration to contextualize the findings and guide further research. The number of participants were limited and the demographic was specific to India, which may restrict the generalizability of findings across different global populations. Assessment of operational variables, including ease of use and performance in the hands of individuals with varying levels of expertise, will provide insights into the wide applicability of the device. These limitations highlight the need for further research, including larger-scale, diverse population studies, and long-term reliability assessments to fully understand Spirofy™’s utility and applicability in broader clinical and non-clinical settings.

Conclusion

This study demonstrates that Spirofy™, an easy-to-use and portable spirometer, is as accurate as the gold standard Vitalograph Alpha Touch™ spirometer for detecting airflow obstruction, with a sensitivity of 97%, specificity of 90%, and overall accuracy of 93.3%. Spirofy™ can thus be an alternative spirometer that can be used in clinical settings for the diagnosis of OAD.

Data availability

The datasets used and/or analyzed during the current study are available from the corresponding author upon reasonable request.

Abbreviations

ATS/ERS:

American Thoracic Society/European Respiratory Society

CI:

Confidence interval

CONSORT:

Consolidated Standards of Reporting Trials

COPD:

Chronic obstructive pulmonary disease

COVID:

Coronavirus disease

FVC:

Forced vital capacity

FEV1 :

Forced expiratory volume

GAN:

Global Asthma Network

IEC/IRB:

Independent Ethics Committee/Institutional Review Board

ITT:

Intention to Treat

K-S:

Kolmogorov–Smirnov

N:

Total number of randomized subjects

OAD:

Obstructive airway diseases

PIS:

Patient information sheet

PP:

Per protocol

%:

Percentage of subjects calculated relative to the total number of randomized subjects

QA:

Quality assurance

RT-PCR:

Real-time-polymerase chain reaction

SARS-CoV-2:

Severe acute respiratory syndrome coronavirus

SD:

Standard deviation

SPSS:

Statistical Packages for Social Science

References

  1. Cukic V, Lovre V, Dragisic D, et al. Asthma and chronic obstructive pulmonary disease (COPD) - differences and similarities. Mater Sociomed. 2012;24:100–5.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Soler X, Ramsdell JW. Are asthma and COPD a continuum of the same disease? J Allergy Clin Immunol Pract. 2015;3:489–95.

    Article  PubMed  Google Scholar 

  3. Global Asthma Report. Available at: http://globalasthmareport.org/2018/resources/Global_Asthma_Report_2018.pdf. Accessed 1 Apr 2022.

  4. GBD 2019 Diseases and Injuries Collaborators. Global burden of 369 diseases and injuries in 204 countries and territories, 1990–2019: a systematic analysis for the global burden of disease study 2019. Lancet. 2020;396:1204–22.

  5. Poongadan MN, Gupta N, Kumar R. Lifestyle factors and asthma in India: a case-control study. Adv Respir Med. 2016;84:104–8.

  6. Daniel RA, Aggarwal P, Kalaivani M, et al. Prevalence of chronic obstructive pulmonary disease in India: a systematic review and meta-analysis. Lung India. 2021;38:506–13.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Salvi S, Madas S, Ghorpade D, et al. Is underuse of inhaled corticosteroids for asthma in India contributing to 42% of global asthma deaths? Lung India. 2022;39:331–6.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Spirometry. Thinking beyond the COPD gold standard. Available at: https://www.patientcareonline.com/view/spirometry-thinking-beyond-copd-gold-standard. Accessed on: 09 December 2022.

  9. Lamb K, Theodore D, Bhutta BS. Spirometry. [Updated 2022 Jul 18]. In: StatPearls [Internet]. Treasure Island (FL): StatPearls Publishing; 2022 Jan–2022 Sep 6. Available at: https://www.ncbi.nlm.nih.gov/books/NBK560526/. Accessed on: 21 March 2023.

  10. Johns DP, Walters JA, Walters EH. Diagnosis and early detection of COPD using spirometry. J Thorac Dis. 2014;6:1557–69.

    PubMed  PubMed Central  Google Scholar 

  11. Wells CD, Joo MJ. COPD and asthma: diagnostic accuracy requires spirometry. J Fam Pract. 2019;68:76–81.

    PubMed  Google Scholar 

  12. Graham BL, Steenbruggen I, Miller MR, et al. Standardization of spirometry 2019 update. An official American thoracic society and European respiratory society technical statement. Am J Respir Crit Care Med. 2019;200:e70–88.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Miller MR, Hankinson J, Brusasco V. ATS/ERS task force: standardization of lung function testing. Eur Respir J. 2005;26:319–38.

    Article  CAS  PubMed  Google Scholar 

  14. Barr RG, Stemple KJ, Mesia-Vela S, et al. Reproducibility and validity of a handheld spirometer. Respir Care. 2009;53(4):433–41.

    Google Scholar 

  15. HernaÂndez CR, FernaÂndez MN, SanmartõÂn AP, et al. Validation of the portable air-smart spirometer. PLoS ONE. 2018;13:e0192789–99.

    Article  Google Scholar 

  16. Kavanagh J, Jackson DJ, Kent BD. Over- and under-diagnosis in asthma. Breathe. 2019;15:e20–7.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Singh S, Salvi S, Mangal DK, et al. Prevalence, time trends and treatment practices of asthma in India: The Global Asthma Network study. ERJ Open Res. 2022;8:00528–2021.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Halpin DMG, Celli BR, Criner GJ, et al. The GOLD Summit on chronic obstructive pulmonary disease in low- and middle-income countries. Int J Tuberc Lung Dis. 2019;23(11):1131–41.

    Article  CAS  PubMed  Google Scholar 

  19. Agusti A, Fabbri LM, Baraldi E, et al. Spirometry: a practical lifespan predictor of global health and chronic respiratory and non-respiratory diseases. Eur J Intern Med. 2021;89:3–9.

    Article  PubMed  Google Scholar 

  20. Milanzi EB, Koppelman GH, Oldenwening M, et al. Considerations in the use of different spirometers in epidemiological studies. Environ Health. 2019;18:1–8.

    Article  Google Scholar 

  21. Ring B, Burbank AJ, Mills K, et al. Validation of an app-based portable spirometer in adolescents with asthma. J Asthma. 2021;58:497–504.

    Article  PubMed  Google Scholar 

  22. Dickens AP, Fitzmaurice DA, Adab P, et al. Accuracy of Vitalograph lung monitor as a screening test for COPD in primary care. Care Respir Med. 2020;30:2–9.

    CAS  Google Scholar 

  23. Boros PW, Maciejewski A, Nowicki MM, et al. Comparability of portable and desktop spirometry: a randomized, parallel assignment, open-label clinical trial. Adv Respir Med. 2022;90:60–7.

    Article  Google Scholar 

  24. Zhou J, Li X, Wang X, et al. Accuracy of portable spirometers in the diagnosis of chronic obstructive pulmonary disease: a meta-analysis. Care Respir Med. 2022;32:15–27.

    CAS  Google Scholar 

  25. Avdimiretz N, Wilson D, Grasemann H. Comparison of a handheld turbine spirometer to conventional spirometry in children with cystic fibrosis. Pediatr Pulmonol. 2020;55:1394–9.

    Article  PubMed  Google Scholar 

  26. Carpenter DM, Jurdi R, Roberts CA, et al. A review of portable electronic spirometers: implications for asthma self-management. Curr Allergy Asthma Rep. 2018;18:53–62.

    Article  PubMed  Google Scholar 

  27. Duong M, Rangarajan S, Zaman M, et al. Differences and agreement between two portable hand-held spirometers across diverse community-based populations in the prospective Urban Rural Epidemiology (PURE) study. PLoS Glob Public Health. 2022;2(2):e0000141.

Download references

Acknowledgements

We would like to thank BioQuest Solutions for providing editorial assistance.

Funding

The study was funded by Cipla Ltd.

Author information

Authors and Affiliations

Authors

Contributions

All authors have contributed equally to the concept, design, draft development, review, and finalization of the manuscript.

Corresponding author

Correspondence to Meena Lopez.

Ethics declarations

Ethics approval and consent to participate

Each participant provided a written informed consent. The study protocol, subject information sheet, and informed consent were approved by the Independent Ethics Committee/Institutional Review Board (IEC/IRB; Dr. Balamurugan: Institutional Ethics Committee of Sri Muthukumaran Medical College, Chennai; Dr. Deepak Talwar: Metro Ethical Review Board for Metro Hospital and heart institute, Noida; Dr. Mahavir Modi: Royal Pune Independent Ethics committee for Modi Clinic, Pune.) before the initiation of the study. The study was performed in accordance with the approved protocol. All the study related activities and methods were performed in accordance with the relevant guidelines and regulations.

Consent for publication

Not Applicable.

Competing interests

Dr. Meena Lopez, Dr. Rushika Shah, Abhijit Vaidya, Nandan Kulkarni, Dr. Sandesh Sawant, and Dr. Jaideep Gogtay are employees of Cipla Ltd. All other authors do not possess any competing interest.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary Material 1

Supplementary Material 2

Supplementary Material 3

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Talwar, D., Balamurugan, S., Modi, M. et al. Comparison of a portable, pneumotach flow-sensor–based spirometer (Spirofy™) with the vitalograph alpha Touch™ spirometer in evaluating lung function in healthy individuals, asthmatics, and COPD patients—a randomized, crossover study. BMC Pulm Med 24, 230 (2024). https://doi.org/10.1186/s12890-024-02972-4