Table 1 |
||
|
NEMA multiple window spatial registration (MWSR) error for the 3 camera heads |
||
|
MWSR Error (mm) |
||
|
|
||
|
Detector Thickness (mm) |
||
|
|
||
|
67Ga |
133Ba |
|
|
|
||
|
9.5 |
0.7 |
0.4 |
|
|
||
|
15.9 |
1.6 |
1.0 |
|
|
||
|
25.4 |
0.7 |
1.0 |
|
|
||
|
NEMA multiple window spatial registration (MWSR) error for the 3 camera heads, given as the maximum deviation over the 5 point source positions. The 133Ba MWSR errors are obtained for the 5 mm collimator hole size configuration. |
||
|
Bergmann et al. BMC Medical Physics 2008 8:6 doi:10.1186/1756-6649-8-6 |
||