Table 3 |
||||||||
|
Test-retest reliability of the goniometer and ETS for hip ROM measurement in patients with FAI and healthy subjects |
||||||||
|
ICC2,1 |
CV (%) |
SEM (°) |
LOA (°) |
|||||
|
|
||||||||
|
Gonio |
ETS |
Gonio |
ETS |
Gonio |
ETS |
Gonio |
ETS |
|
|
|
||||||||
|
Flexion |
0.916 |
0.943 |
3.12 |
2.66 |
3.94 |
2.96 |
0.72 ± 11.15 |
0.27 ± 8.42 |
|
|
||||||||
|
Abduction |
0.924 |
0.947 |
5.84 |
5.66 |
2.36 |
2.01 |
0.14 ± 6.67 |
-0.08 ± 5.72 |
|
|
||||||||
|
Adduction |
0.842 |
0.823 |
6.73 |
6.34 |
2.36 |
1.59 |
-0.54 ± 6.60 |
1.07 ± 4.01 |
|
|
||||||||
|
Internal rotation |
0.950 |
0.902 |
7.74 |
10.19 |
2.42 |
2.93 |
-0.24 ± 6.89 |
1.32 ± 7.90 |
|
|
||||||||
|
External rotation |
0.914 |
0.934 |
5.23 |
5.10 |
2.53 |
1.81 |
0.02 ± 7.16 |
0.25 ± 5.14 |
|
|
||||||||
|
ETS: electromagnetic tracking system, ROM: range of motion, FAI: femoroacetabular impingement, ICC: intraclass correlation coefficient, CV: coefficient of variation, SEM: standard error of measurement, LOA: limits of agreement, Gonio: goniometer. Note that bias (retest lower than test) was significant only for adduction measured with the ETS (P < 0.01). |
||||||||
|
Nussbaumer et al. BMC Musculoskeletal Disorders 2010 11:194 doi:10.1186/1471-2474-11-194 |
||||||||