Detecting imipenem resistance in Acinetobacter baumannii by automated systems (BD Phoenix, Microscan WalkAway, Vitek 2); high error rates with Microscan WalkAway

Abstract

Background

Increasing reports of carbapenem resistant Acinetobacter baumannii infections are of serious concern. Reliable susceptibility testing results remains a critical issue for the clinical outcome. Automated systems are increasingly used for species identification and susceptibility testing. This study was organized to evaluate the accuracies of three widely used automated susceptibility testing methods for testing the imipenem susceptibilities of A. baumannii isolates, by comparing to the validated test methods.

Methods

Selected 112 clinical isolates of A. baumanii collected between January 2003 and May 2006 were tested to confirm imipenem susceptibility results. Strains were tested against imipenem by the reference broth microdilution (BMD), disk diffusion (DD), Etest, BD Phoenix, MicroScan WalkAway and Vitek 2 automated systems. Data were analysed by comparing the results from each test method to those produced by the reference BMD test.

Results

MicroScan performed true identification of all A. baumannii strains while Vitek 2 unidentified one strain, Phoenix unidentified two strains and misidentified two strains. Eighty seven of the strains (78%) were resistant to imipenem by BMD. Etest, Vitek 2 and BD Phoenix produced acceptable error rates when tested against imipenem. Etest showed the best performance with only two minor errors (1.8%). Vitek 2 produced eight minor errors(7.2%). BD Phoenix produced three major errors (2.8%). DD produced two very major errors (1.8%) (slightly higher (0.3%) than the acceptable limit) and three major errors (2.7%). MicroScan showed the worst performance in susceptibility testing with unacceptable error rates; 28 very major (25%) and 50 minor errors (44.6%).

Conclusion

Reporting errors for A. baumannii against imipenem do exist in susceptibility testing systems. We suggest clinical laboratories using MicroScan system for routine use should consider using a second, independent antimicrobial susceptibility testing method to validate imipenem susceptibility. Etest, whereever available, may be used as an easy method to confirm imipenem susceptibility.

Peer Review reports

Background

Acinetobacter baumannii infections mostly affect debilitated patients in intensive care units and are associated with high mortality rates [1, 2]. A. baumannii is difficult to control and treat because of its prolonged environmental survival and its ability to develop resistance to multiple antimicrobial agents [3]. Carbapenems are considered the gold standard treatment for multidrug resistant (MDR) A. baumannii infections, imipenem being the most active agent [1, 3]. However, reports of imipenem-resistant A. baumannii strains have been increasing during the past few years, and these isolates are often multidrug resistant MDR [3–7]. Inappropriate antimicrobial treatment increases mortality in patients with bacteraemia [8]. Reliable susceptibility testing results remains a critical issue for the clinical outcome for these patients.

Automated systems have been increasingly used in many clinical laboratories for species identification of the bacteria and susceptibility testing. These systems are able to decrease the labour and in-laboratory time compared to that required for standardized methods and help the physicians guide efficient antimicrobial therapy based on rapid and convenient results [9]. On the other hand, probable errors reported by test system can have serious implications for the clinical outcome for patients.

Currently MDR Acinetobacters are among the most frequent nosocomial pathogens in our institution. After the introduction of the MicroScan WalkAway system for susceptibility testing in 2006, we noted a considerable proportion of discordant results in imipenem susceptibilities of Acinetobacters compared to disk diffusion results. Upon this observation, this study was organized to evaluate the accuracies of three widely used automated susceptibility testing methods for testing the imipenem susceptibilities of Acinetobacter baumannii isolates, by comparing to the validated test methods.

Methods

One hundred twelve non dublicate clinical isolates of A. baumanii collected between January 2003 and May 2006 were tested to confirm imipenem susceptibility results. These isolates were selected retrospectively among the strain collection derived from the clinical samples submitted to the microbiology laboratory for routine diagnostic procedures. A. baumannii isolates were previously characterized molecularly by PFGE method, and the selected strains were predominantly representing different A. baumannii genotypes found in this institution including both epidemic and sporadic clones. This research has been performed with the approval of Zonguldak Karaelmas University, Application and Research Hospital Ethics Committee. Identification of A. baumannii was performed according to conventional microbiological methods and confirmed by API 20 NE (bioMerieux Inc, France) [10]. Strains were tested against imipenem by broth dilution, disk diffusion, Etest (AB Biodisk, Solna, Sweden) and by the automated systems BD Phoenix (Becton Dickinson Diagnostic Systems, Sparks, MD, USA), MicroScan WalkAway (Dade Behring INC. West Sacramento, CA, USA) and Vitek 2 (bioMe'rieux, Marcy l'Étoile, France). The reference broth microdilution (BMD) test was performed using in-house prepared panels according to CLSI [11, 12]. Imipenem was kindly provided by the manufacturer (Merck Sharp & Dohme, Madrid, Spain). Ouality control strains included P. aeruginosa ATCC 27853 and E. coli ATCC 25922 as recommended by CLSI [11]. Susceptibility of the isolates to the following antibacterial agents was tested by disk-diffusion (DD) method using discs (Oxoid, UK), and interpreted as recommended by CLSI [11]; imipenem (IPM, 10 μg), meropenem (MEM, 10 μg), ceftazidime (CAZ, 30 μg), cefepime (FEP, 30 μg), piperacillin (PRL, 100 μg), cefotaxime (CTX, 30 μg), ciprofloxacin (CIP, 5 μg), levofloxacin (LEV, 5 μg), gentamicin (CN, 10 μg), tetracyclin (TE, 30 μg), trimethoprim-sulfamethoxazole (SXT, 1.25/23.75 μg). The study was performed in three laboratories; for MicroScan WalkAway in Zonguldak Karaelmas University Hospital, Zonguldak, for BD Phoenix in Acibadem Labmed Clinical Laboratory, Istanbul, and for Vitek2 in bioMe'rieux Laboratory, Istanbul. The organisms were tested in a blinded fashion in each laboratory according to the procedures recommended by the manufacturers. All systems were tested with inocula from the same subculture. In addition to the WalkAway automated reading, manual readings were performed on all MicroScan panels and each isolate was tested at least twice, at different times.

Data were analysed by comparing the results from each test method to those produced by BMD test.

SPSS (ver. 11.5) programme was used for all calculations. Kappa tests was used for agreement. If P values less than 0.05 it was accepted statisticaly significant.

Very major errors were considered when an organism was defined as resistant by the reference method but was categorized as susceptible with the tested system. Major errors were defined when an organism found to be susceptible by the reference method was considered resistant with the system. Minor errors occurred when an organism was considered susceptible or resistant either by the reference microdilution method or with the tested system but intermediate by the other method. An overall category error rate of < 10% was considered for an acceptable performance of susceptibility tests, including ≤1.5% of very major errors and ≤3.0% major errors [13].

Results

MicroScan performed correct identification of all (n = 112) A. baumannii strains while Vitek 2 unidentified one strain, and Phoenix unidentified two strains and misidentified two strains.

The BMD testing showed that 25 (22%) strains were susceptible to imipenem while 87 (78%) were resistant. MIC50 and MIC90 for imipenem was 32 μg/ml and 64 μg/ml respectively. Compared to the BMD results all other test systems in the study produced errors when A. baumannii was tested against imipenem. Etest showed the best performance with only two minor error (1.8%). Vitek2 produced eight minor errors(7.2%). BD Phoenix produced three major errors (2.8%). DD produced two very major errors (1.8%) and three major errors (2.7%). MicroScan showed the worst performance in susceptibility testing with 28 very major errors (25%) and 50 minor errors (44.6%). The error rates resulted by the three commercial automated systems and two validated methods are listed in Table 1. When the results were evaluated upon the acceptable performance criteria for susceptibility tests; the error rates of BD Phoenix and Vitek2 were in acceptable limits while MicroScan performed unacceptable results with high very major and minor error rates in detecting imipenem susceptibility.

Table 1 Types of errors produced when testing imipenem susceptibilities of A. baummanii isolates by three commercial automated systems and two validated methods.

Susceptibility rates of the isolates against the given antibiotics; by DD method and the three automated sytems, are shown in Table 2. Variations in susceptibility rates when tested by DD and three automated sytems are also displayed in the table. Discordant susceptibility rates were observed mostly in imipenem, meropenem, cefepime and trimethoprim-sulfamethoxazole.

Table 2 Susceptibilities by disk diffusion and three automated sytems against the antibiotics tested

Numbers of discordances between the results of DD and the three automated systems in susceptibility testing to the given antibiotics are exhibited in Table 3. For trimethoprim-sulfamethoxazole susceptibilities; high numbers of major discordances were displayed between the DD and all three automated systems and also between the automated systems. Minor discordances were mostly observed for meropenem, levofloxacin, cefepime and tetracyclin.between the three automated systems.

Table 3 Correlation of the results of disk diffusion (DD) and automated systems in susceptibility testing to the antibiotics tested: Numbers of discordances.

Discussion

Treatment of severe nosocomial infections increasingly depends on carbapenems. Despite the high level resistance in Acinetobacter spp., carbapenems are still considered as the most active agents [14–17]. On the other hand, emergence of resistance to carbapenems is of serious concern [3]. Outbreaks of imipenem-resistant Acinetobacter have been reported worldwide and carbapenem resistance among A. baumannii can be endemic in certain hospitals as well [3, 18]. Thus, besides the clinical outcome and cost, reliable susceptibility testing results remains a critical issue for hospital infection control and surveillance programs.

Reproducibility, the ability to track results, potential impact on the workflow and the availability of rapid results favor the use of automatic systems in the microbiology laboratory [9]. Rapid identification and susceptibility testing can have a significant impact on the management of infections, especially those caused by antibiotic-resistant bacteria. The development of expert automated systems has allowed an increase both in the reproducibility and in the reliability of the results. Unfortunately numerous studies have reported errors of various automated systems when several organism-antimicrobial combinations tested[9, 19–22].

Discrepancies of resistance to imipenem by automated systems have been described in enteric species, previously [19, 23]. False resistance to imipenem was documented in Pseudomonas aeruginosa, A. baumanni, Proteus mirabilis whereas false susceptibility was reported in Klebsiella pneumoniae [2, 24, 25]. It has been shown that false resistance was due to the imipenem degradation over time, which resulted false increases in MICs of imipenem [24]. Whereas, false susceptible results by the automated systems in K. pneumoniae isolates were attributed in part to low inoculum size [26, 27].

Treatment of A. baumannii infections is guided foremost by in vitro antimicrobial susceptibility assays. Although the BD Phoenix, MicroScan Walkaway and Vitek are commonly used automated test systems in antimicrobial susceptibility, few studies have been reported for discrepancies of resistance to imipenem in A. baumannii [2, 28]. It was also supposed that the prediction of resistance in A. baumannii based upon susceptibility assays might be less certain than for other bacteria [18].

According to our results, it can be stated that, overall, the Vitek2 and Phoenix system reliably carried out susceptibility testing for imipenem against A. baumanni. The MicroScan WalkAway system generally failed to accurately detect imipenem resistance among our collection of carbapenem resistant A. baumanni isolates. Up to our knowledge this is the first description of considerable false susceptibility results to imipenem in A. baumannii by an automated system.

This study puts forth discordant results between the three widely used automated susceptibility testing methods for testing the imipenem susceptibilities of A. baumannii isolates for consideration. According to the fact that the study was performed in three laboratories; standardization difficulties might be questioned. However, all systems were tested with inocula from the same subculture and procedures were performed stringly according to the manufacturers recommendations for both inoculum preparation and technical details at every laboratory. All strains for which the commercial MIC results were discrepant with MIC results from the BMD reference method were retested by using all methods. Besides, the BMD, Etest, DD methods were performed at the same laboratory with the automated system which displayed unacceptable error rates.

Although our results clearly displayed that the MicroScan system failed to accurately detect imipenem resistance among our A. baumanni isolates; it was not possible to attribute the invalid MIC results to a certain factor in our case. Previously it was suggested that the variability in detecting imipenem resistance by automated systems was partly a result of underinoculating the panels and was emphasised that appropriate inoculum size was a critical factor for achieving accurate results especially when cell wall-active antimicrobials were tested [19, 26]. False-susceptible results based on these factors were noted with both VITEK and MicroScan WalkAway systems [26, 27]. Based on this data; our initial effort was careful attention to inoculum. Prompt Inoculation System-D (3 M Company, St. Paul, MN) is generally used practically for inoculum preparation in MicroScan systems. We repeated imipenem susceptibility testing of all discordant isolates in our study using the MicroScan System, with inocula prepared using both Prompt Inoculation System-D and the CLSI recommended protocol. However, repeat testing of isolates with careful attention to inoculum appeared not to improve results. On the other hand, the same bacterial inoculum prepared per each isolate according to CLSI was used concurrently for the Etest and BMD and also retested by the MicroScan WalkAway system, which gave high error rates. According to our results, it should be stated that inoculum size, indeed, cannot be attributed to the inaccurate results in our case.

As recommended by the manufacturer, manual readings were performed on all MicroScan panels as well. It was previously noted that MicroScan yielded several very major errors by the automated instrument readings but only minor or no errors when the tests were read visually[23]. The probable reason for manuel reading was based on the instrument's threshold for identifying bacterial growth [23]. At the same time growth control wells were also checked on each panel. Technical maintenance was carried out by the manufacturer's service and a selected group of the discordant A. baumanni strains was also tested by the application specialist. Nevertheless these efforts appeared not to be able to improve the results performed by MicroScan WalkAway system. Yet we remain unable to explain the high rate of very major and minor errors by MicroScan system Speculating other probable reasons; high false sensitivity of A. baumannii to imipenem might be due to a higher-than-expected or inconsistent concentration of imipenem used in the system, which remains to be further illuminated by the manufacturer. Improper plates and other probable technical errors are other issues needs to be addressed. However it could be noted that our results may only apply under the circumstances in our laboratory as well as the shipping conditions during the study.

According to our results; DD method produced an unacceptable rate of very major errors (1.8%), slightly higher than the acceptable limit (1.5%), that can be reconsidered accounting the total number of the isolates. In consideration of low cost and requirement of no special equipment; DD method, available in most laboratories, seems to be a useful method for susceptibility testing of A. baumannii to imipenem. However, our data suggest that traditional disk diffusion method may result in very low rates of very major errors. Etest method, whereever available, may be used as an accurate testing method to confirm questionable results generated by automated methods and DD for susceptibility testing of A. baumannii to imipenem.

It is noteworthy that; inconsistent results with DD and by all automated systems in susceptibilities to antibiotics other than imipenem were also observed in this study; mainly in susceptibilities to trimethoprim-sulfamethoxazole, meropenem, cefepime and levofloxacin; consecutively. Besides when compared to each other; discordances between the three automated systems were also encountered mostly with these antibiotics. Previously high minor error levels were already noted for automated systems when testing β-lactam antimicrobials in P. aeruginosa [19, 22, 23]. However, the accuracy of the automated systems in susceptibility testing of A. baumannii against the aforementioned antibiotics should be further evaluted with other studies by using the reference BMD method.

Conclusion

Testing difficulties in susceptibility testing do exist in automated susceptibility testing systems. Reporting errors can have serious implications for the clinical outcome for patients. We suggest clinical laboratories using MicroScan automated system for routine use should consider using a second, independent antimicrobial susceptibility testing method to validate imipenem susceptibility. Disk diffusion method may result in very low rates of very major errors. Etest as an easy method may be used to confirm imipenem susceptibility. If treatment failure with carbapenems is observed for isolates of A. baumanni that were previously reported as susceptible to carbapenems, repeat testing with a nonautomated method should be warranted.

About the authors

Canan Kulah: Ass. Asc. Prof. Dr. of Medical Microbiology

Elif Aktas: Ass. Asc. Prof. Dr. of Medical Microbiology

Fusun Comert: Asc. Prof. Dr. of Medical Microbiology

Nagihan Ozlu: Specialist Dr. in Medical Microbiology

Isin Akyar: Specialist Dr. in Medical Microbiology

Handan Ankarali: Asc. Prof. Dr. of Biostatistics

References

  1. Bergogne-Berezin E, Towner KJ: Acinetobacter spp. as nosocomial pathogens: microbiological, clinical, and epidemiological features. Clin Microbiol Rev. 1996, 9: 148-165.

    CAS  PubMed  PubMed Central  Google Scholar 

  2. Tsakris A, Pantazi A, Pournaras S, Maniatis A, Polyzou A, Sofianou D: Pseudo-outbreak of imipenem-resistant Acinetobacter baumannii resulting from false susceptibility testing by a rapid automated system. J Clin Microbiol. 2000, 38 (9): 3505-7.

    CAS  PubMed  PubMed Central  Google Scholar 

  3. Poirel L, Nordmann P: Carbapenem resistance in Acinetobacter baumannii: mechanisms and epidemiology. Clin Microbiol Infect. 2006, 12 (9): 826-836. 10.1111/j.1469-0691.2006.01456.x.

    Article  CAS  PubMed  Google Scholar 

  4. Bou G, Cervero G, Dominguez MA, Quereda C, Martinez-Beltran J: PCR-based DNA fingerprinting (REP-PCR, AP-PCR) and pulsed-field gel electrophoresis characterization of a nosocomial outbreak caused by imipenem- and meropenem-resistant Acinetobacter baumannii. Clin Microbiol Infect. 2000, 6 (12): 635-43. 10.1046/j.1469-0691.2000.00181.x.

    Article  CAS  PubMed  Google Scholar 

  5. Corbella X, Montero A, Pujol M, Domínguez MA, Ayats J, Argerich MJ, Garrigosa F, Ariza J, Gudiol F: Emergence and rapid spread of carbapenem resistance during a large and sustained hospital outbreak of multiresistant Acinetobacter baumannii. J Clin Microbiol. 2000, 38 (11): 4086-95.

    CAS  PubMed  PubMed Central  Google Scholar 

  6. He'ritier C, Dubouix A, Poirel L, Marty N, Nordmann P: A nosocomial outbreak of Acinetobacter baumannii isolates expressing the carbapenem-hydrolysing oxacillinase OXA-58. J Antimicrob Chemother. 2005, 55: 115-118. 10.1093/jac/dkh500.

    Article  Google Scholar 

  7. Pournaras S, Markogiannakis A, Ikonomidis A, Kondyli L, Bethimouti K, Maniatis AN, Legakis NJ, Tsakris A: Outbreak of multiple clones of imipenem-resistant Acinetobacter baumannii isolates expressing OXA-58 carbapenemase in an intensive care unit. J Antimicrob Chemother. 2006, 57: 557-561. 10.1093/jac/dkl004.

    Article  CAS  PubMed  Google Scholar 

  8. Cisneros JM, Reyes MJ, Pachon J, Becerril B, Caballero FJ, García-Garmendía JL, Ortiz C, Cobacho AR: Bacteremia due to Acinetobacter baumannii: epidemiology, clinical and prognostic features. Clin Infect Dis. 1996, 22: 1026-1032.

    Article  CAS  PubMed  Google Scholar 

  9. Donay JL, Mathieu D, Fernandes P, Prégermain C, Bruel P, Wargnier A, Casin I, Weill FX, Lagrange PH, Herrmann JL: Evaluation of the automated Phoenix system for potential routine use in the clinical microbiology laboratory. J Clin Microbiol. 2004, 42 (4): 1542-6. 10.1128/JCM.42.4.1542-1546.2004.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  10. Schreckenberger PC, Daneshvar MI, Weyant RS, Hollis DG: Acinetobacter, Achromobacter, Chryseobacterium, Moraxella, and other nonfermentative gram-negative rods. Manual of Clinical Microbiology. Edited by: Murray PR, Baron EJ, Jorgensen JH, Pfaller MA, Yolken RH. 2003, Washington (DC): ASM Pres, 749-779.

    Google Scholar 

  11. Clinical and Laboratory Standards Institute: Approved standart M100-S16. Performance standards for antimicrobial susceptibility testing. 2006, Clinical and Laboratory Standards Institute, Wayne, Pennsylvania, USA

    Google Scholar 

  12. Clinical and Laboratory Standards Institute: Approved Standart M7-A6. Methods for dilution antimicrobial susceptibility tests for bacteria that growth aerobically. 2005, Clinical and Laboratory Standards Institute, Wayne, Pennsylvania, USA, 6

    Google Scholar 

  13. National Committee for Clinical Laboratory Standards: Approved guideline M23-A2. Development of in vitro susceptibility testing criteria and quality control parameters. 2001, National Committee for Clinical Laboratory Standards, Wayne, Pennsylvania, USA, 2

    Google Scholar 

  14. Goossens H: MYSTIC (Meropenem Yearly Susceptibility Test Information Collection) results from Europe: comparison of antibiotic susceptibilities between countries and centre types. MYSTIC Study Group (European centres only). J Antimicrob Chemother. 2000, 46 (Supp 2): 39-52. 10.1093/jac/46.suppl_2.39.

    Article  CAS  Google Scholar 

  15. Jones RN, Mendes C, Turner PJ, Masterton R: An overview of the Meropenem Yearly Susceptibility Test Information Collection (MYSTIC) Program: 1997–2004. Diagn Microbiol Infect Dis. 2005, 53: 247-256. 10.1016/j.diagmicrobio.2005.10.006.

    Article  PubMed  Google Scholar 

  16. Korten V, Ulusoy S, Zarakolu P, Mete B: Turkish MYSTIC Study Group. Antibiotic resistance surveillance over a 4-year period (2000–2003) in Turkey: results of the MYSTIC Program. Diagn Microbiol Infect Dis. 2007, 59 (4): 453-7. 10.1016/j.diagmicrobio.2007.06.016.

    Article  CAS  PubMed  Google Scholar 

  17. Eraksoy H, Basustaoglu A, Korten V, Kurt H, Ozturk R, Ulusoy S, Yaman A, Yuce A, Zarakolu P, Turkish MYSTIC Study Group: Susceptibility of bacterial isolates from Turkey – a report from the Meropenem Yearly Susceptibility Test Information Collection (MYSTIC) Program. J Chemother. 2007, 19 (6): 650-7.

    Article  CAS  PubMed  Google Scholar 

  18. Perez F, Hujer AM, Hujer KM, Decker BK, Rather PN, Bonomo RA: Global challenge of multidrug-resistant Acinetobacter baumannii. Antimicrob Agents Chemother. 2007, 51 (10): 3471-84. 10.1128/AAC.01464-06.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  19. Doern GV, Brueggemann AB, Perla R, Daly J, Halkias D, Jones RN, Saubolle MA: Multicenter laboratory evaluation of the bioMérieux Vitek antimicrobial susceptibility testing system with 11 antimicrobial agents versus members of the family Enterobacteriaceae and Pseudomonas aeruginosa. J Clin Microbiol. 1997, 35 (8): 2115-2119.

    CAS  PubMed  PubMed Central  Google Scholar 

  20. Biedenbach DJ, Jones RN: Interpretive errors using an automated system for the susceptibility testing of imipenem and aztreonam. Diagn Microbiol Infect Dis. 1995, 21: 57-60. 10.1016/0732-8893(94)00069-9.

    Article  CAS  PubMed  Google Scholar 

  21. Marshall SA: Serious interpretive error among three commercial systems for susceptibility testing of aztreonam. Aztreonam Study Group. Diagn Microbiol Infect Dis. 1995, 22 (3): 249-51. 10.1016/0732-8893(95)00064-H.

    Article  CAS  PubMed  Google Scholar 

  22. Sader HS, Fritsche TR, Jones RN: Accuracy of three automated systems (MicroScan WalkAway, VITEK, and VITEK 2) for susceptibility testing of Pseudomonas aeruginosa against five broad-spectrum beta-lactam agents. J Clin Microbiol. 2006, 44 (3): 1101-4. 10.1128/JCM.44.3.1101-1104.2006.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  23. Steward CD, Mohammed JM, Swenson JM, Stocker SA, Williams PP, Gaynes RP, et al: Antimicrobial susceptibility testing of carbapenems: multicenter validity testing and accuracy levels of five antimicrobial test methods for detecting resistance in Enterobacteriaceae and Pseudomonas aeruginosa isolates. J Clin Microbiol. 2003, 41: 351-8. 10.1128/JCM.41.1.351-358.2003.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  24. White RL, Kays MB, Friedrich LV, Brown EW, Koonce JR: Pseudoresistance of Pseudomonas aeruginosa resulting from degradation of imipenem in an automated susceptibility testing system with predried panels. J Clin Microbiol. 1991, 29: 398-400.

    CAS  PubMed  PubMed Central  Google Scholar 

  25. Tenover FC, Kalsi RK, Williams PP, Carey RB, Stocker S, Lonsway D, Rasheed JK, Biddle JW, McGowan JE, Hanna B: Carbapenem resistance in Klebsiella pneumoniae not detected by automated susceptibility testing. Emerg Infect Dis. 2006, 12 (8): 1209-13.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  26. Bratu S, Mooty M, Nichani S, Landman D, Gullans C, Pettinato B, Karumudi U, Tolaney P, Quale J: Emergence of KPC-possessing Klebsiella pneumoniae in Brooklyn, New York: epidemiology and recommendations for detection. Antimicrob Agents Chemother. 2005, 49: 3018-20. 10.1128/AAC.49.7.3018-3020.2005.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  27. Bratu S, Landman D, Haag R, Recco R, Eramo A, Alam M, Quale J: Rapid spread of carbapenem-resistant Klebsiella pneumoniae in New York City: a new threat to our antibiotic armamentarium. Arch Intern Med. 2005, 165: 1430-5. 10.1001/archinte.165.12.1430.

    Article  CAS  PubMed  Google Scholar 

  28. Endimiani A, Luzzaro F, Tamborini A, Lombardi G, Elia V, Belloni R, Toniolo A: Identification and antimicrobial susceptibility testing of clinical isolates of nonfermenting gram-negative bacteria by the Phoenix Automated Microbiology System. New Microbiol. 2002, 25 (3): 323-9.

    CAS  PubMed  Google Scholar 

Pre-publication history

Download references

Acknowledgements

We are grateful to Istanbul bioMe'rieux Laboratory for providing the Vitek 2 consumables and instruments. We thank Hakan Filizel (Sante Medical Istanbul) for providing Phoenix tests and to Dade Behring Istanbul for providing Etests.

This study was partly presented as a poster at the European Congress of Clinical Microbiology and Infectious Diseases (ECCMID) 2008, Barcelona.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Canan Kulah.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

CK conceived the study, and participated in its design, coordination, carrying out and drafted the manuscript. EA carried out MIC testing and participated in the design of the study. FB participated in coordination and the MIC testing. NO carried out the automated tests and participated in the design. IA carried out BD Phoenix test. HA participated in the design of the study and performed the statistical analysis. All authors read and approved the final manuscript.

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.