            <p>Monitoring of people and workers exposure to the electric, magnetic and electromagnetic fields in an Italian national cancer Institute</p>
         </title>
         <aug>
            <au id="A1" ca="yes" ce="yes">
               <snm>Di Nallo</snm>
               <mnm>Maria</mnm>
               <fnm>Anna</fnm>
               <insr iid="I1"/>
               <email>dinallo@ifo.it</email>
            </au>
            <au id="A2" ce="yes">
               <snm>Strigari</snm>
               <fnm>Lidia</fnm>
               <insr iid="I1"/>
               <email>strigari@ifo.it</email>
            </au>
            <au id="A3" ce="yes">
               <snm>Giliberti</snm>
               <fnm>Claudia</fnm>
               <insr iid="I2"/>
               <email>Claudia.Giliberti@ispesl.it</email>
            </au>
            <au id="A4" ce="yes">
               <snm>Bedini</snm>
               <fnm>Angelico</fnm>
               <insr iid="I2"/>
               <email>Angelico.Bedini@ispesl.it</email>
            </au>
            <au id="A5" ce="yes">
               <snm>Palomba</snm>
               <fnm>Raffaele</fnm>
               <insr iid="I2"/>
               <email>Raffaele.Palomba@ispesl.it</email>
            </au>
            <au id="A6" ce="yes">
               <snm>Benassi</snm>
               <fnm>Marcello</fnm>
               <insr iid="I1"/>
               <email>benassi@ifo.it</email>
            </au>
         </aug>
         <insg>
            <ins id="I1">
               <p>Laboratorio di Fisica Medica e Sistemi Esperti Regina Elena National Cancer Institute, Rome, Italy</p>
            </ins>
            <ins id="I2">
               <p>Dipartimento Insediamenti Produttivi ed Interazione con l'Ambiente Italian National Institute for Health and Safety at Work, Rome, Italy</p>
            </ins>
         </insg>
         <source>Journal of Experimental &amp; Clinical Cancer Research</source>
         <issn>1756-9966</issn>
         <pubdate>2008</pubdate>
         <volume>27</volume>
         <issue>1</issue>
         <fpage>16</fpage>
         <url>http://www.jeccr.com/content/27/1/16</url>
         <xrefbib>
            <pubidlist>
               <pubid idtype="pmpid">18598357</pubid>
               <pubid idtype="doi">10.1186/1756-9966-27-16</pubid>
            </pubidlist>
         </xrefbib>
      </bibl>
      <history>
         <rec>
            <date>
               <day>21</day>
               <month>5</month>
               <year>2008</year>
            </date>
         </rec>
         <acc>
            <date>
               <day>03</day>
               <month>7</month>
               <year>2008</year>
            </date>
         </acc>
         <pub>
            <date>
               <day>03</day>
               <month>7</month>
               <year>2008</year>
            </date>
         </pub>
      </history>
      <cpyrt>
         <year>2008</year>
         <collab>Di Nallo et al; licensee BioMed Central Ltd.</collab>
         <note>This is an Open Access article distributed under the terms of the Creative Commons Attribution License (<url>http://creativecommons.org/licenses/by/2.0</url>), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.</note>
      </cpyrt>
      <abs>
         <sec>
            <st>
               <p>Abstract</p>
            </st>
            <sec>
               <st>
                  <p>Background</p>
               </st>
               <p>The paper reports the electric, magnetic and electromagnetic fields (<it>emf</it>) measurements carried out in the <it>Regina Elena National Cancer Institute (NCI)</it>. Several devices, used in diagnostics and in medical cures, can represent sources of <it>emf </it>for the workers and for the public subjected to the treatments. The aim is to evaluate their exposition, in order to assess the compliance with the law.</p>
            </sec>
            <sec>
               <st>
                  <p>Methods</p>
               </st>
               <p>The investigations have been carried out in the departments of: intensive care, physiotherapy, MR presstherapy and in the surgical rooms. The measurements have been performed using broad band probes in the frequency ranges 5 Hz&#247;30 kHz and 100 kHz-3 GHz.</p>
            </sec>
            <sec>
               <st>
                  <p>Results</p>
               </st>
               <p>The variability of the magnetic induction (B(&#956;T)) levels is between 0,05 &#956;T and 80 &#956;T. The statistical distribution shows that most of the measurements are in the range 0,05&lt;B = 0,5 &#956;T and the 89% of the B(&#956;T) levels are within the 3 &#956;T.</p>
            </sec>
            <sec>
               <st>
                  <p>Conclusion</p>
               </st>
               <p>The measurement of the <it>emf </it>levels in the <it>NCI </it>is recommended because of the presence of the oncological patients; their long stay near the equipments and their day-long exposure represent additional risk factors for which a prudent avoidance strategy have to de adopted.</p>
            </sec>
         </sec>
      </abs>
   </fm>
   <bdy>
      <sec>
         <st>
            <p>Background</p>
         </st>
         <p>In the last years, the increase in the number of systems for telecommunications and for the electric energy transport, has produced a huge debate both in the scientific community and among the public on the potential risks on health due to long time exposition to the electric field (<it>ef</it>), magnetic field (<it>mf</it>) and electromagnetic field (<it>emf</it>) in the working and living environments. The effects produced by the interaction of these physical agents with the biological systems depend on frequency, amplitude of the fields and on the time of the exposition. The effects can be classified in <it>thermal </it>and <it>non-thermal </it><abbrgrp><abbr bid="B1">1</abbr></abbrgrp>. The former are scientifically well-know and consist in heat production in the biological system; the exposure limits recommended by the national and international legislation protect from this injury. The existence of the <it>non-thermal </it>effects represents an open issue; they concern the evidence of biological changes without an effective increase of temperature. The scientific debate is centred on the issue if a long exposition to <it>ef-mf-emf </it>levels lower than the limits could cause harmful effects on health. Many national and international research projects are involved in this investigation by <it>in vivo</it>, <it>in vitro </it>and epidemiological studies. <abbrgrp><abbr bid="B2">2</abbr><abbr bid="B3">3</abbr><abbr bid="B4">4</abbr><abbr bid="B5">5</abbr><abbr bid="B6">6</abbr><abbr bid="B7">7</abbr></abbrgrp>. Until now, conclusive results are still not available.</p>
         <p>In Italy, the issue of people exposition to the <it>ef-mf-emf </it>is regulated by two decrees promulgated in 2003, one for fields produced by sources with frequencies between 100 kHz and 300 GHz <abbrgrp><abbr bid="B8">8</abbr></abbrgrp>, the other for power lines at the frequency of 50 Hz <abbrgrp><abbr bid="B9">9</abbr></abbrgrp>. Their distinctive feature is to establish people long term exposure levels to <it>ef-mf-emf </it>lower than the limits recommended by the international standards fixed by the <it>International Commission on non-ionizing radiation protection (ICNIRP) </it><abbrgrp><abbr bid="B10">10</abbr></abbrgrp>. For sources not referable to power lines in the frequency range 0 Hz-100 kHz, the whole of the restrictions fixed by the <it>European Council Recommendation of 12 July 1999 </it><abbrgrp><abbr bid="B11">11</abbr></abbrgrp> has to be considered. About the workers, the directive 2004/40/EC <abbrgrp><abbr bid="B12">12</abbr></abbrgrp>, recently incorporated into the Italian legislation system with the decree n. 257/2007 <abbrgrp><abbr bid="B13">13</abbr></abbrgrp>, has to be applied.</p>
         <p>The main aim of this paper is to report the <it>ef-mf-emf </it>measurements carried out in the <it>Regina Elena National Cancer Institute </it>(<it>NCI</it>). Several electric and electronic devices, used in diagnostics and in medical cures, have been investigated, which can represent potential sources of <it>ef-mf-emf </it>for the health staff, for the public subjected to the treatments and also for their accompanists. The main peculiarity of this study is then to verify the exposition to these physical agents for a class of particularly vulnerable patients, as the oncological ones, in order to assess the compliance with the law. The long stay of these patients near the equipments and their day-long exposure can represent additional risk factors for which a prudent avoidance strategy has to be adopted.</p>
      </sec>
      <sec>
         <st>
            <p>Methods</p>
         </st>
         <p>The investigations have been carried out in the intensive care division, the physiotherapy department, the surgical rooms, the Magnetic Resonance (MR) department and the press-therapy room. For each division, the investigated equipments are reported in Table <tblr tid="T1">1</tblr>. The study has been extended also to a wireless equipment for internal communications and to external sources.</p>
         <tbl id="T1">
            <title>
               <p>Table 1</p>
            </title>
            <caption>
               <p>Equipments investigated in each division of the <it>NCI-</it>IRE</p>
            </caption>
            <tblbdy cols="3">
               <r>
                  <c ca="center">
                     <p>
                        <b>Divisions</b>
                     </p>
                  </c>
                  <c ca="center">
                     <p>
                        <b>Equipments</b>
                     </p>
                  </c>
                  <c ca="center">
                     <p>
                        <b>Manufacturer</b>
                     </p>
                  </c>
               </r>
               <r>
                  <c cspan="3">
                     <hr/>
                  </c>
               </r>
               <r>
                  <c ca="center">
                     <p>
                        <b>Physiotherapy division</b>
                     </p>
                  </c>
                  <c ca="left">
                     <p>Electro-stimulators</p>
                  </c>
                  <c ca="left">
                     <p>Biorem TENS</p>
                  </c>
               </r>
               <r>
                  <c>
                     <p/>
                  </c>
                  <c ca="left">
                     <p>Electro-stimulators</p>
                  </c>
                  <c ca="left">
                     <p>Biorem ionophoresis (IP)</p>
                  </c>
               </r>
               <r>
                  <c>
                     <p/>
                  </c>
                  <c ca="left">
                     <p>Laser therapy</p>
                  </c>
                  <c ca="left">
                     <p>Leve Laser</p>
                  </c>
               </r>
               <r>
                  <c>
                     <p/>
                  </c>
                  <c ca="left">
                     <p>Ultrasound therapy</p>
                  </c>
                  <c ca="left">
                     <p>Radar Biorem</p>
                  </c>
               </r>
               <r>
                  <c>
                     <p/>
                  </c>
                  <c ca="left">
                     <p>Cyclette</p>
                  </c>
                  <c ca="left">
                     <p>Technogym</p>
                  </c>
               </r>
               <r>
                  <c>
                     <p/>
                  </c>
                  <c ca="left">
                     <p>Pressotherapy</p>
                  </c>
                  <c ca="left">
                     <p>Linfopress Fisioline</p>
                  </c>
               </r>
               <r>
                  <c ca="center">
                     <p>
                        <b>Press-therapy room</b>
                     </p>
                  </c>
                  <c ca="left">
                     <p>Lymphatic Drainage</p>
                  </c>
                  <c ca="left">
                     <p>Linfopress</p>
                  </c>
               </r>
               <r>
                  <c>
                     <p/>
                  </c>
                  <c ca="left">
                     <p>Lymphatic Drainage</p>
                  </c>
                  <c ca="left">
                     <p>Linfopress studio</p>
                  </c>
               </r>
               <r>
                  <c ca="center">
                     <p>
                        <b>Intensive care division</b>
                     </p>
                  </c>
                  <c ca="left">
                     <p>Workplaces &#8211; PC</p>
                  </c>
                  <c ca="left">
                     <p>Philips</p>
                  </c>
               </r>
               <r>
                  <c>
                     <p/>
                  </c>
                  <c ca="left">
                     <p>Electric panel</p>
                  </c>
                  <c ca="left">
                     <p>FAAE</p>
                  </c>
               </r>
               <r>
                  <c>
                     <p/>
                  </c>
                  <c ca="left">
                     <p>Monitor</p>
                  </c>
                  <c ca="left">
                     <p>Philips</p>
                  </c>
               </r>
               <r>
                  <c>
                     <p/>
                  </c>
                  <c ca="left">
                     <p>Ventilator</p>
                  </c>
                  <c ca="left">
                     <p>Dr&#228;ger</p>
                  </c>
               </r>
               <r>
                  <c>
                     <p/>
                  </c>
                  <c ca="left">
                     <p>Pump</p>
                  </c>
                  <c ca="left">
                     <p>Abbott</p>
                  </c>
               </r>
               <r>
                  <c ca="center">
                     <p>
                        <b>Surgical rooms</b>
                     </p>
                  </c>
                  <c ca="left">
                     <p>Ventilator</p>
                  </c>
                  <c ca="left">
                     <p>Dr&#228;ger</p>
                  </c>
               </r>
               <r>
                  <c>
                     <p/>
                  </c>
                  <c ca="left">
                     <p>Electric scalpel</p>
                  </c>
                  <c ca="left">
                     <p>Martin/Conmed</p>
                  </c>
               </r>
               <r>
                  <c ca="center">
                     <p>
                        <b>MR</b>
                     </p>
                  </c>
                  <c ca="left">
                     <p>Magnetic Resonance</p>
                  </c>
                  <c ca="left">
                     <p>Philips (0,5 T)</p>
                  </c>
               </r>
               <r>
                  <c ca="center">
                     <p>
                        <b>Other departments</b>
                     </p>
                  </c>
                  <c ca="left">
                     <p>DECT</p>
                  </c>
                  <c ca="left">
                     <p>Alcatel</p>
                  </c>
               </r>
            </tblbdy>
         </tbl>
         <p>Initially, the technical characteristics of each equipment have been collected. Moreover, their operating modalities during diagnostic and therapy sessions, and their localizations in each area have been acquired, in order to detect potential interactions.</p>
         <p>The measurements have been performed using a broad band probe (EMR 300 Wandel &amp; Goltermann) providing the total electric field for the frequency range 100 kHz &#8211; 3 GHz. According with the Italian decrees <abbrgrp><abbr bid="B8">8</abbr><abbr bid="B9">9</abbr></abbrgrp>, the length of time of each measurement was six minutes. For the frequencies range 5 Hz&#247;30 kHz, the EFA-3 Wandel &amp; Goltermann, equipped with an external magnetic probe, has been used. All the measurements have been carried out in the operating clinical conditions.</p>
         <p>The check on the external sources has been executed using the narrow band device SRM3000 Narda, equipped with an isotropic probe working in the frequency range 75 MHz&#8211;3 GHz.</p>
      </sec>
      <sec>
         <st>
            <p>Results and discussion</p>
         </st>
         <sec>
            <st>
               <p>(a) Physiotherapy department</p>
            </st>
            <p>The physiotherapy department (<it>12x3m) </it>includes an <it>open space</it>, where <it>cyclettes </it>and <it>tapis roulant </it>are positioned, and n. 4 boxes where the devices for specific therapies (Table <tblr tid="T1">1</tblr>) can be placed on demand. Broad band electric field measurements, carried out in the <it>open space</it>, with all the equipments working in operating conditions, have supplied values lower than the sensitivity of the electric field probe (1,0 V/m). In table <tblr tid="T2">2</tblr> and <tblr tid="T3">3</tblr> are reported the results of the magnetic induction (B(&#956;T)) measurements respectively near the <it>tapis roulant </it>(Fig. <figr fid="F1">1</figr>) and specific equipments, conveniently placed in one box (Fig. <figr fid="F2">2</figr>). B(&#956;T) levels are generally lower than 1 &#956;T, with few exceptions, near the <it>tapis roulant </it>(4,5 &#956;T), and near the press-therapy equipment, where a maximum B(&#956;T) level of 1,6 &#956;T has been registered. Effective electric field (E<sub>eff</sub>(V/m)) levels lower than 2,3 V/m have been registered near the equipment for ultrasound therapy. Further measurements, performed near the head of the Laser Power Diode, have provided E<sub>eff</sub>(V/m) measurements in the range 1,0 -3,0 V/m.</p>
            <tbl id="T2">
               <title>
                  <p>Table 2</p>
               </title>
               <caption>
                  <p>The magnetic induction measurements near the <it>tapis roulant </it>(fig.1)</p>
               </caption>
               <tblbdy cols="2">
                  <r>
                     <c ca="center">
                        <p>
                           <b>Points of measurements</b>
                        </p>
                     </c>
                     <c ca="center">
                        <p>
                           <b>B (&#956;T)</b>
                        </p>
                     </c>
                  </r>
                  <r>
                     <c cspan="2">
                        <hr/>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>1</p>
                     </c>
                     <c ca="center">
                        <p>
                           <b>4,5</b>
                        </p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>2</p>
                     </c>
                     <c ca="center">
                        <p>0,03</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>3</p>
                     </c>
                     <c ca="center">
                        <p>0,01</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>4</p>
                     </c>
                     <c ca="center">
                        <p>0,03</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>5</p>
                     </c>
                     <c ca="center">
                        <p>0,02</p>
                     </c>
                  </r>
               </tblbdy>
            </tbl>
            <tbl id="T3">
               <title>
                  <p>Table 3</p>
               </title>
               <caption>
                  <p>The magnetic induction B(&#956;T) and effective electric field E<sub>eff</sub>(V/m) levels in the box described in fig. 2</p>
               </caption>
               <tblbdy cols="6">
                  <r>
                     <c ca="center">
                        <p>
                           <b>Points</b>
                        </p>
                     </c>
                     <c cspan="2" ca="center">
                        <p>
                           <b>Electro stimulator equipment</b>
                        </p>
                     </c>
                     <c ca="center">
                        <p>
                           <b>Press-therapy equipment</b>
                        </p>
                     </c>
                     <c cspan="2" ca="center">
                        <p>
                           <b>Ultrasounds therapy equipment</b>
                        </p>
                     </c>
                  </r>
                  <r>
                     <c>
                        <p/>
                     </c>
                     <c cspan="2">
                        <hr/>
                     </c>
                     <c>
                        <p/>
                     </c>
                     <c>
                        <p/>
                     </c>
                     <c>
                        <p/>
                     </c>
                  </r>
                  <r>
                     <c>
                        <p/>
                     </c>
                     <c ca="center">
                        <p>
                           <b>TENS</b>
                        </p>
                     </c>
                     <c ca="center">
                        <p>
                           <b>IP</b>
                        </p>
                     </c>
                     <c>
                        <p/>
                     </c>
                     <c>
                        <p/>
                     </c>
                     <c>
                        <p/>
                     </c>
                  </r>
                  <r>
                     <c>
                        <p/>
                     </c>
                     <c cspan="5">
                        <hr/>
                     </c>
                  </r>
                  <r>
                     <c>
                        <p/>
                     </c>
                     <c ca="center">
                        <p>
                           <b>B (&#956;T)</b>
                        </p>
                     </c>
                     <c ca="center">
                        <p>
                           <b>B (&#956;T)</b>
                        </p>
                     </c>
                     <c ca="center">
                        <p>
                           <b>B (&#956;T)</b>
                        </p>
                     </c>
                     <c ca="center">
                        <p>
                           <b>B (&#956;T)</b>
                        </p>
                     </c>
                     <c ca="center">
                        <p>
                           <b>E (V/m) (100 kHz&#247;3 GHz)</b>
                        </p>
                     </c>
                  </r>
                  <r>
                     <c cspan="6">
                        <hr/>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>1</p>
                     </c>
                     <c ca="center">
                        <p>0,2</p>
                     </c>
                     <c ca="center">
                        <p>0,2</p>
                     </c>
                     <c ca="center">
                        <p>0,5</p>
                     </c>
                     <c ca="center">
                        <p>0,1</p>
                     </c>
                     <c ca="center">
                        <p>&lt;1</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>2</p>
                     </c>
                     <c ca="center">
                        <p>0,1</p>
                     </c>
                     <c ca="center">
                        <p>0,1</p>
                     </c>
                     <c ca="center">
                        <p>0,5</p>
                     </c>
                     <c ca="center">
                        <p>0,2</p>
                     </c>
                     <c ca="center">
                        <p>2,3</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>3</p>
                     </c>
                     <c ca="center">
                        <p>0,07</p>
                     </c>
                     <c ca="center">
                        <p>0,1</p>
                     </c>
                     <c ca="center">
                        <p>1,1</p>
                     </c>
                     <c ca="center">
                        <p>0,1</p>
                     </c>
                     <c ca="center">
                        <p>3</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>4</p>
                     </c>
                     <c ca="center">
                        <p>0,1</p>
                     </c>
                     <c ca="center">
                        <p>0,1</p>
                     </c>
                     <c ca="center">
                        <p>0,5</p>
                     </c>
                     <c ca="center">
                        <p>0,3</p>
                     </c>
                     <c ca="center">
                        <p>&lt; 1</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>5</p>
                     </c>
                     <c ca="center">
                        <p>0,3</p>
                     </c>
                     <c ca="center">
                        <p>1,0</p>
                     </c>
                     <c ca="center">
                        <p>
                           <b>1,6</b>
                        </p>
                     </c>
                     <c ca="center">
                        <p>0,4</p>
                     </c>
                     <c ca="center">
                        <p>2,2</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>6</p>
                     </c>
                     <c ca="center">
                        <p>0,02</p>
                     </c>
                     <c ca="center">
                        <p>0,03</p>
                     </c>
                     <c ca="center">
                        <p>0,3</p>
                     </c>
                     <c ca="center">
                        <p>0,5</p>
                     </c>
                     <c ca="center">
                        <p>&lt; 1</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>7</p>
                     </c>
                     <c ca="center">
                        <p>0,02</p>
                     </c>
                     <c ca="center">
                        <p>0,02</p>
                     </c>
                     <c ca="center">
                        <p>0,03</p>
                     </c>
                     <c ca="center">
                        <p>0,4</p>
                     </c>
                     <c ca="center">
                        <p>&lt; 1</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>8</p>
                     </c>
                     <c ca="center">
                        <p>0,03</p>
                     </c>
                     <c ca="center">
                        <p>0,02</p>
                     </c>
                     <c ca="center">
                        <p>0,07</p>
                     </c>
                     <c ca="center">
                        <p>0,3</p>
                     </c>
                     <c ca="center">
                        <p>&lt; 1</p>
                     </c>
                  </r>
               </tblbdy>
            </tbl>
            <fig id="F1">
               <title>
                  <p>Figure 1</p>
               </title>
               <caption>
                  <p>Points of measurements near the <it>tapis roulant </it>(detected frequency: 50 Hz)</p>
               </caption>
               <text>
                  <p>
                     <b>Points of measurements near the <it>tapis roulant </it>(detected frequency: 50 Hz).</b>
                  </p>
               </text>
               <graphic file="1756-9966-27-16-1"/>
            </fig>
            <fig id="F2">
               <title>
                  <p>Figure 2</p>
               </title>
               <caption>
                  <p>Specific equipments in a physiotherapy box</p>
               </caption>
               <text>
                  <p>
                     <b>Specific equipments in a physiotherapy box.</b>
                  </p>
               </text>
               <graphic file="1756-9966-27-16-2"/>
            </fig>
         </sec>
         <sec>
            <st>
               <p>(b) Press-therapy room</p>
            </st>
            <p>The rehabilitation area consists of: the reception, the press-therapy room and a medical office. In particular, in the press-therapy room, two equipments for the arms <it>Lymphatic Drainage </it>are placed. The duration of this therapeutic treatment varies from 2 hours up to 4 hours and is generally performed for four patients at the same time. The B(&#956;T) levels, measured in the points reported in Fig. <figr fid="F3">3</figr>, are reported in Table <tblr tid="T4">4</tblr>.</p>
            <tbl id="T4">
               <title>
                  <p>Table 4</p>
               </title>
               <caption>
                  <p>The B(&#956;T) levels measured in the points reported in Fig. 3</p>
               </caption>
               <tblbdy cols="2">
                  <r>
                     <c ca="center">
                        <p>
                           <b>Points of measurements</b>
                        </p>
                     </c>
                     <c ca="center">
                        <p>
                           <b>B (&#956;T)</b>
                        </p>
                     </c>
                  </r>
                  <r>
                     <c cspan="2">
                        <hr/>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>1</p>
                     </c>
                     <c ca="center">
                        <p>
                           <b>80,0</b>
                        </p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>2</p>
                     </c>
                     <c ca="center">
                        <p>5,3</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>3</p>
                     </c>
                     <c ca="center">
                        <p>2,0</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>4</p>
                     </c>
                     <c ca="center">
                        <p>4,8</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>5</p>
                     </c>
                     <c ca="center">
                        <p>
                           <b>11,1</b>
                        </p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>6</p>
                     </c>
                     <c ca="center">
                        <p>2,3</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>7</p>
                     </c>
                     <c ca="center">
                        <p>1,1</p>
                     </c>
                  </r>
               </tblbdy>
            </tbl>
            <fig id="F3">
               <title>
                  <p>Figure 3</p>
               </title>
               <caption>
                  <p>Results of the magnetic induction measurements near the press-therapy equipments</p>
               </caption>
               <text>
                  <p>
                     <b>Results of the magnetic induction measurements near the press-therapy equipments.</b>
                  </p>
               </text>
               <graphic file="1756-9966-27-16-3"/>
            </fig>
            <p>During the treatments, the B(&#956;T) levels detected near the patients are lower than 1 &#956;T. Around the equipments, in positions where only the staff could eventually stay, B(&#956;T) levels between 1,1 and 5,0 &#956;T have been registered. Higher levels (up to 80 &#956;T) have been obtained near the air compressors and the electronics of the equipments, in the positions reported in Table <tblr tid="T4">4</tblr>. Additional measurements, performed at a height of about 10 cm from the floor, have provided B(&#956;T) levels lower than 1,4 &#956;T. Further investigations, carried out in the contiguous rooms, during the equipments standard working conditions, provided B(&#956;T) values between 0,1 and 0,2 &#956;T.</p>
         </sec>
         <sec>
            <st>
               <p>(c) Intensive care division</p>
            </st>
            <p>The intensive care division consists of an <it>open space</it>, where three patients' beds and two monitoring workplaces are located, and three rooms. In Table <tblr tid="T5">5</tblr>, the results of the B(&#956;T) measurements in the <it>open space </it>(Fig. <figr fid="F4">4</figr>) near the sources listed in Table <tblr tid="T1">1</tblr>, are reported; the magnetic induction measurements, carried out at a height of 1,0 m and 1,9 m from the floor, near the electric panel and near the intensive care staff workplaces, are lower than 1,9 &#956;T. Magnetic field background measurements in the <it>open space </it>provided B(&#956;T) levels equal to 70 nT. The high frequencies E<sub>eff</sub>(V/m) values, measured near a patient's bed, resulted lower than 1,0 V/m. In Table <tblr tid="T5">5</tblr>, the results of the B(&#956;T) measurements around a bed (Fig. <figr fid="F5">5</figr>) are reported. The B(&#956;T) levels close to the patients are lower than 1 &#956;T, while levels up to 7,0 &#956;T have been detected at <it>1 m </it>from the bed, at a distance of few centimetres from the volumetric and syringe infusion pumps. Higher B(&#956;T) levels (equal to 14 &#956;T) have been measured at the bottom of the bed near the engine of the anti-bedsore mattress.</p>
            <tbl id="T5">
               <title>
                  <p>Table 5</p>
               </title>
               <caption>
                  <p>The B(&#956;T) measurements in the open space reported in Fig. 4</p>
               </caption>
               <tblbdy cols="3">
                  <r>
                     <c ca="center">
                        <p>
                           <b>Points of measurements</b>
                        </p>
                     </c>
                     <c cspan="2" ca="center">
                        <p>
                           <b>B (&#956;T)</b>
                        </p>
                     </c>
                  </r>
                  <r>
                     <c>
                        <p/>
                     </c>
                     <c cspan="2">
                        <hr/>
                     </c>
                  </r>
                  <r>
                     <c>
                        <p/>
                     </c>
                     <c ca="center">
                        <p>H 1 m</p>
                     </c>
                     <c ca="center">
                        <p>H 1.9 m</p>
                     </c>
                  </r>
                  <r>
                     <c cspan="3">
                        <hr/>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>1</p>
                     </c>
                     <c ca="center">
                        <p>1,7</p>
                     </c>
                     <c ca="center">
                        <p>1,2</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>2</p>
                     </c>
                     <c ca="center">
                        <p>1,9</p>
                     </c>
                     <c ca="center">
                        <p>1,1</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>3</p>
                     </c>
                     <c ca="center">
                        <p>0,9</p>
                     </c>
                     <c ca="center">
                        <p>0,7</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>4</p>
                     </c>
                     <c ca="center">
                        <p>0,3</p>
                     </c>
                     <c ca="center">
                        <p>0,2</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>5</p>
                     </c>
                     <c ca="center">
                        <p>0,3</p>
                     </c>
                     <c ca="center">
                        <p>0,1</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>6</p>
                     </c>
                     <c ca="center">
                        <p>0,1</p>
                     </c>
                     <c ca="center">
                        <p>0,1</p>
                     </c>
                  </r>
                  <r>
                     <c ca="center">
                        <p>7</p>
                     </c>
                     <c ca="center">
                        <p>0,07</p>
                     </c>
                     <c ca="center">
                        <p>0,07</p>
                     </c>
                  </r>
               </tblbdy>
            </tbl>
            <fig id="F4">
               <title>
                  <p>Figure 4</p>
               </title>
               <caption>
                  <p>Results of the magnetic induction measurements B(&#956;T) in the intensive care division</p>
               </caption>
               <text>
                  <p>
                     <b>Results of the magnetic induction measurements B(&#956;T) in the intensive care division.</b>
                  </p>
               </text>
               <graphic file="1756-9966-27-16-4"/>
            </fig>
            <fig id="F5">
               <title>
                  <p>Figure 5</p>
               </title>
               <caption>
                  <p>Measurements near a patient's bed in the intensive care division</p>
               </caption>
               <text>
                  <p>
                     <b>Measurements near a patient's bed in the intensive care division.</b>
                  </p>
               </text>
               <graphic file="1756-9966-27-16-5"/>
            </fig>
         </sec>
         <sec>
            <st>
               <p>(d) Surgical room</p>
            </st>
            <p>In the surgical room (<it>6x12 </it>m), an operating table and two trolleys equipped with ventilators and electric scalpels are placed. B(&#956;T) measurements, performed with the equipments in working conditions, without patients, provided levels of about 50 nT, comparable with the background. Levels up to 24,0 &#956;T have been measured near the electric power of the electric scalpels, at 20 cm from the floor. Broad band E<sub>eff</sub>(V/m) measurements, with all the equipments in working conditions, resulted close to the sensitivity of the electric field probe (1,0 V/m).</p>
         </sec>
         <sec>
            <st>
               <p>(e) MR room and other internal sources</p>
            </st>
            <p>High frequencies E<sub>eff</sub>(V/m) measurements near the MR and into the gate room during the sequences activation, showed levels lower than 1,0 V/m.</p>
            <p>Further investigations have been performed in other departments where the indoor DECT base stations for wireless communication are placed. These apparatus are generally located in the passageways, on the wall at a height of 2 m from the floor. High frequencies E<sub>eff</sub>(V/m) measurements have been carried out at a height of 1,0 m and 1,9 m from the floor close to the DECT systems. The E<sub>eff</sub>(V/m) levels within 1 m from the equipments, at a height of 1,9 m are comprised between 1,0 and 1,3 V/m.</p>
         </sec>
         <sec>
            <st>
               <p>(f) External sources</p>
            </st>
            <p>Further investigations have been carried out outside the <it>NCI</it>. In a radius of about 300 m a radiobase station for cellular communications is placed on the roof of a building. In the distance of a few km from the <it>NCI</it>, a tower with several broadcasting systems for radio, TV, and cellular radiobase stations is located. Investigations with the broad and the narrow band tools in the external area of the <it>NCI </it>show that these sources don't give a significant contribution to the electromagnetic field background in the area. Additional investigations inside the <it>NCI</it>, in particular in the physiotherapy division, allowed to detect a weak electric field at the frequency of 1,9 GHz produced by a radiobase station for the cellular networks (Fig. <figr fid="F6">6</figr>).</p>
            <fig id="F6">
               <title>
                  <p>Figure 6</p>
               </title>
               <caption>
                  <p>Results of the spectrum analysis carried out in the physiotherapy division</p>
               </caption>
               <text>
                  <p>
                     <b>Results of the spectrum analysis carried out in the physiotherapy division.</b>
                  </p>
               </text>
               <graphic file="1756-9966-27-16-6"/>
            </fig>
            <p>In Italy, people exposure to the <it>ef-mf-emf </it>is defined by two decrees <abbrgrp><abbr bid="B8">8</abbr><abbr bid="B9">9</abbr></abbrgrp> in terms of three levels: (i) <it>limits</it>, levels that haven't to be exceeded in any circumstances for the protection from short terms effects; (ii) <it>attention levels </it>that have not to be exceeded into gambling areas, schools and buildings where people stay for more than 4 hours, as a precaution from the potential risks on health due to long time exposition; (iii) <it>quality targets</it>, levels to achieve in a long period in order to minimize people exposure. The respective levels for the protection of people exposure at high and low frequencies are reported in Tables <tblr tid="T7">7</tblr> and <tblr tid="T8">8</tblr>.</p>
            <p>The European recommendation <abbrgrp><abbr bid="B11">11</abbr></abbrgrp> establishes <it>basic restrictions </it>and <it>reference levels </it>for limiting exposure. <it>Basic restrictions </it>are restrictions on exposure to time-varying electric, magnetic, and electromagnetic fields which are based directly on established health effects and biological considerations. Depending upon the frequency of the field, the physical quantities used to specify these restrictions are magnetic flux density (B), current density (J), specific energy absorption rate (SAR), and power density (S). The <it>reference levels </it>are electric field strength (E), magnetic field strength (H), B, and S. <it>Reference levels </it>and <it>basic restrictions </it>have been developed following a thorough review of all published scientific literature. The criteria applied in the course of the review were designed to evaluate the credibility of the various reported findings; only established effects were used as a basis for the proposed exposure restrictions. Induction of cancer from long-term electromagnetic fields exposure was not considered. However, since there are safety factors of about 50 between the threshold values for acute effects and the basis restrictions, this recommendation implicitly covers also possible long-term effects in the whole frequency range. The respective levels are reported in Table <tblr tid="T9">9</tblr>.</p>
            <p>The directive 2004/40/EC on the minimum health and safety requirements regarding the exposure of workers to the risks arising from electromagnetic fields, has been issued in the European Official Journal on May 25, 2004, together with its recent application in the Italian regulation <abbrgrp><abbr bid="B13">13</abbr></abbrgrp>. It introduces measures protecting workers from the risks associated with <it>emf</it>, owing to their effects on the health and safety. The long-term effects, including possible carcinogenic effects due to exposure to time-varying <it>ef-mf-emf </it>for which there is no conclusive scientific evidence establishing a causal relationship, are not addressed. <it>Exposure limit values </it>and <it>action values </it>have been defined. The former are based directly on established health effects and biological considerations and compliance with these limits ensures that workers exposed to <it>emf </it>are protected against all known adverse health effects. The <it>action values </it>refer to the magnitude of directly measurable parameters, provided in terms of E, H, B and S; the observance with these values ensure compliance with the relevant exposure limit. <it>Action values </it>are obtained from the exposure limit values according to the rationale used by the <it>International Commission on Non-ionising Radiation Protection (ICNIRP) </it>in its guidelines on limiting exposure to non-ionising radiation <abbrgrp><abbr bid="B13">13</abbr></abbrgrp>.</p>
            <p>For the health staff in the <it>NCI</it>, the exposure levels are in agreement with the recommendations fixed by the international standards. However, for the health staff, the more restrictive Italian people long term exposure levels have been established as investigation levels. The results of the monitoring in the working areas frequented by patients and accompanies, near the equipments reported in Table <tblr tid="T1">1</tblr> show a variability of the B(&#956;T) measurements between 0,01 and 80 &#956;T. The global statistical distribution of the B(&#956;T) levels shows that most of the measurements are in the range 0,05&lt;B = 0,5 &#956;T, with 25% of the levels registered in the surgical room, 9% in the press-therapy room and 15% in the physiotherapy division. The 89% of the B(&#956;T) levels measured near electric and electronic devices used in diagnostics and in medical cures, in positions normally occupied by the patients during the treatments are within 3 &#956;T, while 8% are within 10 &#956;T. Only 3% of the measurements is higher than this level but within the restrictive <it>limits </it>imposed by the Italian law to regulate the exposure of the public to the <it>mf</it>. The last levels have been measured typically near the engine of the tools and near the electric power of the instrumentations in areas where the presence of the population is not scheduled. In order to minimize or reduce exposures, it is then sufficient to set the area around the devices, with the indications to the staff not to stay close to the devices in their operative conditions.</p>
            <p>The exposure conditions of the patients for the extremely low frequencies and the <it>emf </it>in the intensive care are in accordance with the measurements carried out by Petrucci <abbrgrp><abbr bid="B14">14</abbr></abbrgrp>. The results show that the type of equipments, their number and also their orientation respect to the patient is important to minimize <it>mf </it>levels; in general, a collection of equipments should be avoided and a safety distance have to be adopted to avoid unnecessary exposures.</p>
            <p>About the possible electromagnetic interference (EMI) with electronic equipments by radio waves coming from outside the hospital, we find E<sub>eff</sub>(V/m) sensibly lower than the values registered by Hanada et others <abbrgrp><abbr bid="B15">15</abbr></abbrgrp>. However, measurements of the electromagnetic environment should be performed by each hospital specially in urban areas <abbrgrp><abbr bid="B16">16</abbr></abbrgrp> where some sources could potentially induce strong electric field intensity causing malfunctions.</p>
         </sec>
      </sec>
      <sec>
         <st>
            <p>Conclusion</p>
         </st>
         <p>The monitoring campaign of <it>ef-mf-emf </it>levels is suggested in areas equipped with a large number of electric and electronic devices used during the treatments, which can represent potential sources of <it>ef-mf-emf</it>. This investigation is also important for the safety of the workers and also for the patients subjected to the treatments and particularly vulnerable, as the oncological patients and also for the accompanists exposures.</p>
         <p>The results of the electric and magnetic field measurements, carried out in the <it>NCI</it>, near devices normally used in diagnostics and in medical cures, in the departments of intensive care, physiotherapy, MR, press-therapy and in the surgical rooms, show that the exposure levels are lower than that established for the protection of people and workers.</p>
         <p>However the monitoring campaign in the hospital environments is recommended when E<sub>eff</sub>(V/m) values beyond the threshold for the electromagnetic compatibility are registered <abbrgrp><abbr bid="B17">17</abbr><abbr bid="B18">18</abbr><abbr bid="B19">19</abbr></abbrgrp>.</p>
         <p>The proper evaluation of the <it>ef-mf-emf </it>in the hospital environments allow to map their levels around the equipments, to choose tools with low emission of <it>ef-mf-emf</it>, and to design accurately and correctly the working areas, detecting potential interactions, reducing and minimizing the exposures.</p>
      </sec>
      <sec>
         <st>
            <p>Competing interests</p>
         </st>
         <p>The authors declare that they have no competing interests.</p>
      </sec>
      <sec>
         <st>
            <p>Authors' contributions</p>
         </st>
         <p>All authors contributed equally to this work, read and approved the final manuscript.</p>
         <tbl id="T6">
            <title>
               <p>Table 6</p>
            </title>
            <caption>
               <p>The B(&#956;T) measurements around the bed reported in Fig. 5</p>
            </caption>
            <tblbdy cols="2">
               <r>
                  <c ca="center">
                     <p>
                        <b>Points of measurements</b>
                     </p>
                  </c>
                  <c ca="center">
                     <p>
                        <b>B (&#956;T)</b>
                     </p>
                  </c>
               </r>
               <r>
                  <c cspan="2">
                     <hr/>
                  </c>
               </r>
               <r>
                  <c ca="center">
                     <p>8</p>
                  </c>
                  <c ca="center">
                     <p>0,6</p>
                  </c>
               </r>
               <r>
                  <c ca="center">
                     <p>9</p>
                  </c>
                  <c ca="center">
                     <p>0,4</p>
                  </c>
               </r>
               <r>
                  <c ca="center">
                     <p>10</p>
                  </c>
                  <c ca="center">
                     <p>0,1</p>
                  </c>
               </r>
               <r>
                  <c ca="center">
                     <p>11</p>
                  </c>
                  <c ca="center">
                     <p>
                        <b>7,0</b>
                     </p>
                  </c>
               </r>
               <r>
                  <c ca="center">
                     <p>12</p>
                  </c>
                  <c ca="center">
                     <p>1,0</p>
                  </c>
               </r>
               <r>
                  <c ca="center">
                     <p>13</p>
                  </c>
                  <c ca="center">
                     <p>0,1</p>
                  </c>
               </r>
               <r>
                  <c ca="center">
                     <p>14</p>
                  </c>
                  <c ca="center">
                     <p>6,0</p>
                  </c>
               </r>
               <r>
                  <c ca="center">
                     <p>15</p>
                  </c>
                  <c ca="center">
                     <p>0,7</p>
                  </c>
               </r>
               <r>
                  <c ca="center">
                     <p>16</p>
                  </c>
                  <c ca="center">
                     <p>
                        <b>14,0</b>
                     </p>
                  </c>
               </r>
            </tblbdy>
         </tbl>
         <tbl id="T7">
            <title>
               <p>Table 7</p>
            </title>
            <caption>
               <p>Limits, levels and quality targets for people exposure to 100 kHz-300 GHz <abbrgrp><abbr bid="B8">8</abbr></abbrgrp></p>
            </caption>
            <tblbdy cols="4">
               <r>
                  <c ca="left">
                     <p>
                        <b>Exposure limits</b>
                     </p>
                  </c>
                  <c ca="center">
                     <p>
                        <b>Electric field Intensity E (V/m)</b>
                     </p>
                  </c>
                  <c ca="center">
                     <p>
                        <b>Magnetic field intensity H (A/m)</b>
                     </p>
                  </c>
                  <c ca="center">
                     <p>
                        <b>Power density D (W/m<sup>2</sup>)</b>
                     </p>
                  </c>
               </r>
               <r>
                  <c cspan="4">
                     <hr/>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>0,1 &lt; f &#8804; 3 MHz</p>
                  </c>
                  <c ca="center">
                     <p>60</p>
                  </c>
                  <c ca="center">
                     <p>0,2</p>
                  </c>
                  <c ca="center">
                     <p>-</p>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>3 &lt; f &#8804; 3000 MHz</p>
                  </c>
                  <c ca="center">
                     <p>20</p>
                  </c>
                  <c ca="center">
                     <p>0,05</p>
                  </c>
                  <c ca="center">
                     <p>1</p>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>3 &lt; f &#8804; 300 GHz</p>
                  </c>
                  <c ca="center">
                     <p>40</p>
                  </c>
                  <c ca="center">
                     <p>0,01</p>
                  </c>
                  <c ca="center">
                     <p>4</p>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>
                        <b>Attention level</b>
                     </p>
                  </c>
                  <c>
                     <p/>
                  </c>
                  <c>
                     <p/>
                  </c>
                  <c>
                     <p/>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>0,1 MHz&lt; f &#8804; 300 GHz</p>
                  </c>
                  <c ca="center">
                     <p>6</p>
                  </c>
                  <c ca="center">
                     <p>0,016</p>
                  </c>
                  <c ca="center">
                     <p>0,10 (3 MHz-300 GHz)</p>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>
                        <b>Quality targets</b>
                     </p>
                  </c>
                  <c>
                     <p/>
                  </c>
                  <c>
                     <p/>
                  </c>
                  <c>
                     <p/>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>0,1 MHz&lt; f &#8804; 300 GHz</p>
                  </c>
                  <c ca="center">
                     <p>6</p>
                  </c>
                  <c ca="center">
                     <p>0,016</p>
                  </c>
                  <c ca="center">
                     <p>0,10 (3 MHz-300 GHz)</p>
                  </c>
               </r>
            </tblbdy>
         </tbl>
         <tbl id="T8">
            <title>
               <p>Table 8</p>
            </title>
            <caption>
               <p>Limits, levels and quality targets for people exposure to power lines at 50 Hz <abbrgrp><abbr bid="B9">9</abbr></abbrgrp></p>
            </caption>
            <tblbdy cols="4">
               <r>
                  <c>
                     <p/>
                  </c>
                  <c ca="center">
                     <p>
                        <b>Exposure limits</b>
                     </p>
                  </c>
                  <c ca="center">
                     <p>
                        <b>Attention levels</b>
                     </p>
                  </c>
                  <c ca="center">
                     <p>
                        <b>Quality targets</b>
                     </p>
                  </c>
               </r>
               <r>
                  <c cspan="4">
                     <hr/>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>E<sub>eff </sub>(kV/m)</p>
                  </c>
                  <c ca="center">
                     <p>5</p>
                  </c>
                  <c ca="center">
                     <p>-</p>
                  </c>
                  <c ca="center">
                     <p>-</p>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>B<sub>eff </sub>(&#956;T)</p>
                  </c>
                  <c ca="center">
                     <p>100</p>
                  </c>
                  <c ca="center">
                     <p>10</p>
                  </c>
                  <c ca="center">
                     <p>3</p>
                  </c>
               </r>
            </tblbdy>
         </tbl>
         <tbl id="T9">
            <title>
               <p>Table 9</p>
            </title>
            <caption>
               <p>Reference levels for the general public from <abbrgrp><abbr bid="B11">11</abbr></abbrgrp>.</p>
            </caption>
            <tblbdy cols="5">
               <r>
                  <c cspan="5" ca="center">
                     <p>
                        <b>Council Recommendation of 12 July 1999 Reference levels for electric, magnetic and electromagnetic fields (0 Hz to 300 GHz, unperturbed rms values)</b>
                     </p>
                  </c>
               </r>
               <r>
                  <c cspan="5">
                     <hr/>
                  </c>
               </r>
               <r>
                  <c ca="center">
                     <p>
                        <b>Frequency range</b>
                     </p>
                  </c>
                  <c ca="center">
                     <p>
                        <b>E-field strength (V/m)</b>
                     </p>
                  </c>
                  <c ca="center">
                     <p>
                        <b>H-field strenght (A/m)</b>
                     </p>
                  </c>
                  <c ca="center">
                     <p>
                        <b>B-field (&#956;T)</b>
                     </p>
                  </c>
                  <c ca="center">
                     <p>
                        <b>Equivalent plane wave power density Seq (W/m<sup>2</sup>)</b>
                     </p>
                  </c>
               </r>
               <r>
                  <c cspan="5">
                     <hr/>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>0&#8211;1 Hz</p>
                  </c>
                  <c ca="center">
                     <p>-</p>
                  </c>
                  <c ca="center">
                     <p>3,2 &#215; 104</p>
                  </c>
                  <c ca="center">
                     <p>4 &#215; 104</p>
                  </c>
                  <c ca="center">
                     <p>-</p>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>1&#8211;8 Hz</p>
                  </c>
                  <c ca="center">
                     <p>10 000</p>
                  </c>
                  <c ca="center">
                     <p>3,2 &#215; 104/f<sup>2</sup></p>
                  </c>
                  <c ca="center">
                     <p>4 &#215; 104/f<sup>2</sup></p>
                  </c>
                  <c ca="center">
                     <p>-</p>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>8&#8211;25 Hz</p>
                  </c>
                  <c ca="center">
                     <p>10 000</p>
                  </c>
                  <c ca="center">
                     <p>4 000/f</p>
                  </c>
                  <c ca="center">
                     <p>5000/f</p>
                  </c>
                  <c ca="center">
                     <p>-</p>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>0,025&#8211;0,8 Hz</p>
                  </c>
                  <c ca="center">
                     <p>250/f</p>
                  </c>
                  <c ca="center">
                     <p>4/f</p>
                  </c>
                  <c ca="center">
                     <p>5/f</p>
                  </c>
                  <c ca="center">
                     <p>-</p>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>0,8&#8211;3 kHz</p>
                  </c>
                  <c ca="center">
                     <p>250/f</p>
                  </c>
                  <c ca="center">
                     <p>5</p>
                  </c>
                  <c ca="center">
                     <p>6,25</p>
                  </c>
                  <c ca="center">
                     <p>-</p>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>3&#8211;150 kHz</p>
                  </c>
                  <c ca="center">
                     <p>87</p>
                  </c>
                  <c ca="center">
                     <p>5</p>
                  </c>
                  <c ca="center">
                     <p>6,25</p>
                  </c>
                  <c ca="center">
                     <p>-</p>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>0,15&#8211;1 MHz</p>
                  </c>
                  <c ca="center">
                     <p>87</p>
                  </c>
                  <c ca="center">
                     <p>0,73/f</p>
                  </c>
                  <c ca="center">
                     <p>0,92/f</p>
                  </c>
                  <c ca="center">
                     <p>-</p>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>1&#8211;10 MHz</p>
                  </c>
                  <c ca="center">
                     <p>87/f<sup>1/2</sup></p>
                  </c>
                  <c ca="center">
                     <p>0,73/f</p>
                  </c>
                  <c ca="center">
                     <p>0,92/f</p>
                  </c>
                  <c ca="center">
                     <p>-</p>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>10&#8211;400 MHz</p>
                  </c>
                  <c ca="center">
                     <p>28</p>
                  </c>
                  <c ca="center">
                     <p>0,073</p>
                  </c>
                  <c ca="center">
                     <p>0,092</p>
                  </c>
                  <c ca="center">
                     <p>2</p>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>400&#8211;2 000 MHz</p>
                  </c>
                  <c ca="center">
                     <p>1,375 f<sup>1/2</sup></p>
                  </c>
                  <c ca="center">
                     <p>0,0037 f<sup>1/2</sup></p>
                  </c>
                  <c ca="center">
                     <p>0,0046 f<sup>1/2</sup></p>
                  </c>
                  <c ca="center">
                     <p>f/200</p>
                  </c>
               </r>
               <r>
                  <c ca="left">
                     <p>2&#8211;300 GHz</p>
                  </c>
                  <c ca="center">
                     <p>61</p>
                  </c>
                  <c ca="center">
                     <p>0,16</p>
                  </c>
                  <c ca="center">
                     <p>0,20</p>
                  </c>
                  <c ca="center">
                     <p>10</p>
                  </c>
               </r>
            </tblbdy>
         </tbl>
      </sec>
   </bdy>
   <bm>
      <ack>
         <sec>
            <st>
               <p>Acknowledgements</p>
            </st>
            <p>The authors are indebted to the Head and staff of Surgical and Intensive care, Physiotherapy and Press-therapy, Radiology Departments of IRE for their continued general support.</p>
         </sec>
      </ack>
      <refgrp>
         <bibl id="B1">
            <title>
               <p>WHO Electromagnetic fields</p>
            </title>
            <url>http://www.who.int/peh-emf/en/</url>
         </bibl>
         <bibl id="B2">
            <title>
               <p>EMF and health</p>
            </title>
            <aug>
               <au>
                  <snm>Feychting</snm>
                  <fnm>M</fnm>
               </au>
               <au>
                  <snm>Ahlbom</snm>
                  <fnm>A</fnm>
               </au>
               <au>
                  <snm>Kheifets</snm>
                  <fnm>L</fnm>
               </au>
            </aug>
            <source>Annu Rev Public Health</source>
            <pubdate>2005</pubdate>
            <volume>26</volume>
            <fpage>165</fpage>
            <lpage>89</lpage>
            <xrefbib>
               <pubidlist>
                  <pubid idtype="doi">10.1146/annurev.publhealth.26.021304.144445</pubid>
                  <pubid idtype="pmpid" link="fulltext">15760285</pubid>
               </pubidlist>
            </xrefbib>
         </bibl>
         <bibl id="B3">
            <title>
               <p>Public Health Impact of Extremely Low-Frequency Electromagnetic Fields</p>
            </title>
            <aug>
               <au>
                  <snm>Kheifets</snm>
                  <fnm>L</fnm>
               </au>
               <au>
                  <snm>Afifi</snm>
                  <fnm>AA</fnm>
               </au>
               <au>
                  <snm>Shimkhada</snm>
                  <fnm>R</fnm>
               </au>
            </aug>
            <source>Environmental Health Perspectives</source>
            <pubdate>2006</pubdate>
            <volume>114</volume>
            <issue>10</issue>
            <fpage>1532</fpage>
            <lpage>1537</lpage>
            <xrefbib>
               <pubidlist>
                  <pubid idtype="pmcid">1626420</pubid>
                  <pubid idtype="pmpid" link="fulltext">17035138</pubid>
               </pubidlist>
            </xrefbib>
         </bibl>
         <bibl id="B4">
            <title>
               <p>Source of Funding and Results of Studies of Health Effects of Mobile Phone Use Systematic Review of Experimental Studies</p>
            </title>
            <aug>
               <au>
                  <snm>Huss</snm>
                  <fnm>A</fnm>
               </au>
               <au>
                  <snm>Egger</snm>
                  <fnm>M</fnm>
               </au>
               <au>
                  <snm>Hug</snm>
                  <fnm>K</fnm>
               </au>
               <au>
                  <snm>Huwiler-M&#252;ntener</snm>
                  <fnm>K</fnm>
               </au>
               <au>
                  <snm>R&#246;&#246;sli</snm>
                  <fnm>M</fnm>
               </au>
            </aug>
            <source>Environmental Health Perspectives</source>
            <pubdate>2007</pubdate>
            <volume>115</volume>
            <issue>1</issue>
            <fpage>1</fpage>
            <lpage>4</lpage>
            <xrefbib>
               <pubidlist>
                  <pubid idtype="pmcid">1797826</pubid>
                  <pubid idtype="pmpid">17366811</pubid>
               </pubidlist>
            </xrefbib>
         </bibl>
         <bibl id="B5">
            <title>
               <p>Base Stations and Wireless Networks&#8211;Radiofrequency (RF) Exposures and Health Consequences</p>
            </title>
            <aug>
               <au>
                  <snm>Valberg</snm>
                  <fnm>PA</fnm>
               </au>
               <au>
                  <snm>van Deventer</snm>
                  <fnm>TE</fnm>
               </au>
               <au>
                  <snm>Repacholi</snm>
                  <fnm>MH</fnm>
               </au>
               <au>
                  <snm>Workgroup</snm>
                  <fnm>Report</fnm>
               </au>
            </aug>
            <source>Environmental Health Perspectives</source>
            <pubdate>2007</pubdate>
            <volume>115</volume>
            <issue>3</issue>
            <fpage>416</fpage>
            <lpage>424</lpage>
            <xrefbib>
               <pubidlist>
                  <pubid idtype="pmcid">1849947</pubid>
                  <pubid idtype="pmpid">17431492</pubid>
               </pubidlist>
            </xrefbib>
         </bibl>
         <bibl id="B6">
            <title>
               <p>The INTERPHONE study: design, epidemiological methods, and description of the study population</p>
            </title>
            <aug>
               <au>
                  <snm>Cardis</snm>
                  <fnm>E</fnm>
               </au>
               <au>
                  <snm>Richardson</snm>
                  <fnm>L</fnm>
               </au>
               <au>
                  <snm>Deltour</snm>
                  <fnm>I</fnm>
               </au>
               <au>
                  <snm>Armstrong</snm>
                  <fnm>B</fnm>
               </au>
               <au>
                  <snm>Feychting</snm>
                  <fnm>M</fnm>
               </au>
               <au>
                  <snm>Johansen</snm>
                  <fnm>C</fnm>
               </au>
               <au>
                  <snm>Kilkenny</snm>
                  <fnm>M</fnm>
               </au>
               <au>
                  <snm>McKinney</snm>
                  <fnm>P</fnm>
               </au>
               <au>
                  <snm>Modan</snm>
                  <fnm>B</fnm>
               </au>
               <au>
                  <snm>Sadetzki</snm>
                  <fnm>S</fnm>
               </au>
               <au>
                  <snm>Sch&#252;z</snm>
                  <fnm>J</fnm>
               </au>
               <au>
                  <snm>Swerdlow</snm>
                  <fnm>A</fnm>
               </au>
               <au>
                  <snm>Vrijheid</snm>
                  <fnm>M</fnm>
               </au>
               <au>
                  <snm>Auvinen</snm>
                  <fnm>A</fnm>
               </au>
               <au>
                  <snm>Berg</snm>
                  <fnm>G</fnm>
               </au>
               <au>
                  <snm>Blettner</snm>
                  <fnm>M</fnm>
               </au>
               <au>
                  <snm>Bowman</snm>
                  <fnm>J</fnm>
               </au>
               <au>
                  <snm>Brown</snm>
                  <fnm>J</fnm>
               </au>
               <au>
                  <snm>Chetrit</snm>
                  <fnm>A</fnm>
               </au>
               <au>
                  <snm>Christensen</snm>
                  <fnm>HC</fnm>
               </au>
               <au>
                  <snm>Cook</snm>
                  <fnm>A</fnm>
               </au>
               <au>
                  <snm>Hepworth</snm>
                  <fnm>S</fnm>
               </au>
               <au>
                  <snm>Giles</snm>
                  <fnm>G</fnm>
               </au>
               <au>
                  <snm>Hours</snm>
                  <fnm>M</fnm>
               </au>
               <au>
                  <snm>Iavarone</snm>
                  <fnm>I</fnm>
               </au>
               <au>
                  <snm>Jarus-Hakak</snm>
                  <fnm>A</fnm>
               </au>
               <au>
                  <snm>Klaeboe</snm>
                  <fnm>L</fnm>
               </au>
               <au>
                  <snm>Krewski</snm>
                  <fnm>D</fnm>
               </au>
               <au>
                  <snm>Lagorio</snm>
                  <fnm>S</fnm>
               </au>
               <au>
                  <snm>L&#246;nn</snm>
                  <fnm>S</fnm>
               </au>
               <au>
                  <snm>Mann</snm>
                  <fnm>S</fnm>
               </au>
               <au>
                  <snm>McBride</snm>
                  <fnm>M</fnm>
               </au>
               <au>
                  <snm>Muir</snm>
                  <fnm>K</fnm>
               </au>
               <au>
                  <snm>Nadon</snm>
                  <fnm>L</fnm>
               </au>
               <au>
                  <snm>Parent</snm>
                  <fnm>ME</fnm>
               </au>
               <au>
                  <snm>Pearce</snm>
                  <fnm>N</fnm>
               </au>
               <au>
                  <snm>Salminen</snm>
                  <fnm>T</fnm>
               </au>
               <au>
                  <snm>Schoemaker</snm>
                  <fnm>M</fnm>
               </au>
               <au>
                  <snm>Schlehofer</snm>
                  <fnm>B</fnm>
               </au>
               <au>
                  <snm>Siemiatycki</snm>
                  <fnm>J</fnm>
               </au>
               <au>
                  <snm>Taki</snm>
                  <fnm>M</fnm>
               </au>
               <au>
                  <snm>Takebayashi</snm>
                  <fnm>T</fnm>
               </au>
               <au>
                  <snm>Tynes</snm>
                  <fnm>T</fnm>
               </au>
               <au>
                  <snm>van Tongeren</snm>
                  <fnm>M</fnm>
               </au>
               <au>
                  <snm>Vecchia</snm>
                  <fnm>P</fnm>
               </au>
               <au>
                  <snm>Wiart</snm>
                  <fnm>J</fnm>
               </au>
               <au>
                  <snm>Woodward</snm>
                  <fnm>A</fnm>
               </au>
               <au>
                  <snm>Yamaguchi</snm>
                  <fnm>N</fnm>
               </au>
            </aug>
            <source>Eur J Epidemiol</source>
            <pubdate>2007</pubdate>
            <volume>22</volume>
            <issue>9</issue>
            <fpage>647</fpage>
            <lpage>664</lpage>
            <xrefbib>
               <pubidlist>
                  <pubid idtype="doi">10.1007/s10654-007-9152-z</pubid>
                  <pubid idtype="pmpid" link="fulltext">17636416</pubid>
               </pubidlist>
            </xrefbib>
         </bibl>
         <bibl id="B7">
            <title>
               <p>Preliminary study of cause-specific mortality of a population exposed to 50 Hz magnetic fields, in a district of Rome municipality</p>
            </title>
            <aug>
               <au>
                  <snm>Fazzo</snm>
                  <fnm>L</fnm>
               </au>
               <au>
                  <snm>Grignoli</snm>
                  <fnm>M</fnm>
               </au>
               <au>
                  <snm>Iavarone</snm>
                  <fnm>I</fnm>
               </au>
               <au>
                  <snm>Polichetti</snm>
                  <fnm>A</fnm>
               </au>
               <au>
                  <snm>De Santis</snm>
                  <fnm>M</fnm>
               </au>
               <au>
                  <snm>Fano</snm>
                  <fnm>V</fnm>
               </au>
               <au>
                  <snm>Forastiere</snm>
                  <fnm>F</fnm>
               </au>
               <au>
                  <snm>Palange</snm>
                  <fnm>S</fnm>
               </au>
               <au>
                  <snm>Pasetto</snm>
                  <fnm>R</fnm>
               </au>
               <au>
                  <snm>Vanacore</snm>
                  <fnm>N</fnm>
               </au>
               <au>
                  <snm>Comba</snm>
                  <fnm>P</fnm>
               </au>
            </aug>
            <source>Epidemiol Prev</source>
            <pubdate>2005</pubdate>
            <volume>29</volume>
            <issue>5&#8211;6</issue>
            <fpage>243</fpage>
            <lpage>252</lpage>
            <xrefbib>
               <pubid idtype="pmpid">16669160</pubid>
            </xrefbib>
         </bibl>
         <bibl id="B8">
            <title>
               <p>Exposure limits, attention levels and quality targets for protection of people exposure to electric, magnetic and electromagnetic fields produced with frequency between 100 kHz and 300 GHz</p>
            </title>
            <aug>
               <au>
                  <cnm>Italian DPCM del 8.7.2003</cnm>
               </au>
            </aug>
            <source>G.U. 28/08/n. 199</source>
            <pubdate>2003</pubdate>
         </bibl>
         <bibl id="B9">
            <title>
               <p>Exposure limits, attention levels and quality targets for protection of people exposure to electric and magnetic fields produced by power lines at the frequency of 50 Hz</p>
            </title>
            <aug>
               <au>
                  <cnm>Italian DPCM 8.7.2003</cnm>
               </au>
            </aug>
            <source>G.U. 29/8/n. 200</source>
            <pubdate>2003</pubdate>
         </bibl>
         <bibl id="B10">
            <title>
               <p>Guidelines for Limiting Exposure to Time-Varying Electric, Magnetic, and Electromagnetic Fields (up to 300 GHz)</p>
            </title>
            <source>ICNIRP Health Physics</source>
            <pubdate>1998</pubdate>
            <volume>74</volume>
            <fpage>494</fpage>
            <lpage>522</lpage>
            <xrefbib>
               <pubidlist>
                  <pubid idtype="pmpid">9525427</pubid>
               </pubidlist>
            </xrefbib>
         </bibl>
         <bibl id="B11">
            <title>
               <p>On the limitation of exposure of the general public to electromagnetic fields</p>
            </title>
            <aug>
               <au>
                  <cnm>Council Recommendation of 12 July 1999</cnm>
               </au>
            </aug>
            <source>(0 Hz to 300 GHz) 1999/519/CE/30</source>
            <pubdate>1999</pubdate>
         </bibl>
         <bibl id="B12">
            <title>
               <p>Directive 2004/40/EC of the European Parliament and of the Council of 29 April 2004 on the minimum health and safety requirements regarding the exposure of workers to the risks arising from physical agents (electromagnetic fields)</p>
            </title>
            <source>Eighteenth individual Directive within the meaning of Article 16(1) of Directive 89/391/EEC</source>
         </bibl>
         <bibl id="B13">
            <title>
               <p>Attuazione della direttiva 2004/40/CE sulle prescrizioni minime di sicurezza e di salute relative all'esposizione dei lavoratori ai rischi derivanti dagli agenti fisici (campi elettromagnetici)</p>
            </title>
            <aug>
               <au>
                  <cnm>Italian D. Lgs. 19 novembre n. 257</cnm>
               </au>
            </aug>
            <source>G.U. 11/01/2008, n. 9</source>
            <pubdate>2007</pubdate>
         </bibl>
         <bibl id="B14">
            <title>
               <p>Exposure of the critically ill patient to extremely low-frequency electromagnetic fields in the intensive care environment</p>
            </title>
            <aug>
               <au>
                  <snm>Petrucci</snm>
                  <fnm>N</fnm>
               </au>
            </aug>
            <source>Intensive Care Med</source>
            <pubdate>1999</pubdate>
            <volume>25</volume>
            <fpage>847</fpage>
            <lpage>851</lpage>
            <xrefbib>
               <pubidlist>
                  <pubid idtype="doi">10.1007/s001340050963</pubid>
                  <pubid idtype="pmpid" link="fulltext">10447544</pubid>
               </pubidlist>
            </xrefbib>
         </bibl>
         <bibl id="B15">
            <title>
               <p>Possible electromagnetic interference with electronic medical equipment by radio waves coming from outside the hospital</p>
            </title>
            <aug>
               <au>
                  <snm>Hanada</snm>
                  <fnm>E</fnm>
               </au>
               <au>
                  <snm>Kodama</snm>
                  <fnm>K</fnm>
               </au>
               <au>
                  <snm>Takano</snm>
                  <fnm>K</fnm>
               </au>
               <au>
                  <snm>Watanabe</snm>
                  <fnm>Y</fnm>
               </au>
               <au>
                  <snm>Nose</snm>
                  <fnm>Y</fnm>
               </au>
            </aug>
            <source>J Med Syst</source>
            <pubdate>2001</pubdate>
            <volume>25</volume>
            <issue>4</issue>
            <fpage>257</fpage>
            <lpage>267</lpage>
            <xrefbib>
               <pubidlist>
                  <pubid idtype="doi">10.1023/A:1010727220929</pubid>
                  <pubid idtype="pmpid">11463202</pubid>
               </pubidlist>
            </xrefbib>
         </bibl>
         <bibl id="B16">
            <title>
               <p>The electromagnetic field background in the urban area of Rome (Italy)</p>
            </title>
            <aug>
               <au>
                  <snm>Giliberti</snm>
                  <fnm>C</fnm>
               </au>
               <au>
                  <snm>Bedini</snm>
                  <fnm>A</fnm>
               </au>
               <au>
                  <snm>Palomba</snm>
                  <fnm>R</fnm>
               </au>
               <au>
                  <snm>D'Emilia</snm>
                  <fnm>E</fnm>
               </au>
               <au>
                  <snm>Magli</snm>
                  <fnm>CV</fnm>
               </au>
               <au>
                  <snm>Giuliani</snm>
                  <fnm>L</fnm>
               </au>
            </aug>
            <source>The situation until 2001 Proceedings of the IEEE 6th International Symposium on electromagnetic compatibility and electromagnetic ecology: 21&#8211;24 June 2005; Saint Petersburg</source>
            <fpage>208</fpage>
            <lpage>212</lpage>
         </bibl>
         <bibl id="B17">
            <title>
               <p>Cellular phone interference with the operation of mechanical ventilators</p>
            </title>
            <aug>
               <au>
                  <snm>Shaw</snm>
                  <fnm>CI</fnm>
               </au>
               <au>
                  <snm>Kacmarek</snm>
                  <fnm>RM</fnm>
               </au>
               <au>
                  <snm>Hampton</snm>
                  <fnm>RL</fnm>
               </au>
               <au>
                  <snm>Riggi</snm>
                  <fnm>V</fnm>
               </au>
               <au>
                  <snm>El Masry</snm>
                  <fnm>A</fnm>
               </au>
               <au>
                  <snm>Cooper</snm>
                  <fnm>JB</fnm>
               </au>
               <au>
                  <snm>Hurford</snm>
                  <fnm>WE</fnm>
               </au>
            </aug>
            <source>Crit Care Med</source>
            <pubdate>2004</pubdate>
            <volume>32</volume>
            <issue>4</issue>
            <fpage>928</fpage>
            <lpage>931</lpage>
            <xrefbib>
               <pubidlist>
                  <pubid idtype="doi">10.1097/01.CCM.0000120061.01431.DB</pubid>
                  <pubid idtype="pmpid" link="fulltext">15071380</pubid>
               </pubidlist>
            </xrefbib>
         </bibl>
         <bibl id="B18">
            <title>
               <p>Mobile phones in the hospital: improved mobile communication and mitigation of EMI concerns can lead to an overall benefit to healthcare</p>
            </title>
            <aug>
               <au>
                  <snm>Morrissey</snm>
                  <fnm>JJ</fnm>
               </au>
            </aug>
            <source>Health Phys</source>
            <pubdate>2004</pubdate>
            <volume>87</volume>
            <issue>1</issue>
            <fpage>82</fpage>
            <lpage>88</lpage>
            <xrefbib>
               <pubidlist>
                  <pubid idtype="doi">10.1097/00004032-200407000-00011</pubid>
                  <pubid idtype="pmpid" link="fulltext">15194927</pubid>
               </pubidlist>
            </xrefbib>
         </bibl>
         <bibl id="B19">
            <title>
               <p>Electromagnetic interference of GSM cellular phones and UHF radios with intensive-care and operating-room ventilators</p>
            </title>
            <aug>
               <au>
                  <snm>Barbaro</snm>
                  <fnm>V</fnm>
               </au>
               <au>
                  <snm>Bartolini</snm>
                  <fnm>P</fnm>
               </au>
               <au>
                  <snm>Benassi</snm>
                  <fnm>M</fnm>
               </au>
               <au>
                  <snm>Di Nallo</snm>
                  <fnm>AM</fnm>
               </au>
               <au>
                  <snm>Reali</snm>
                  <fnm>L</fnm>
               </au>
               <au>
                  <snm>Valsecchi</snm>
                  <fnm>M</fnm>
               </au>
            </aug>
            <source>Biomedical instrumentation and technology</source>
            <pubdate>2000</pubdate>
            <volume>34</volume>
            <fpage>361</fpage>
            <lpage>369</lpage>