Comparative assessment of commercial ELISA kits for detection of HIV in India

Abstract

Background

India harbors the 3rd highest HIV infected population globally. The magnitude of the HIV detection challenge is enormous. ELISA is the most commonly used screening technique for HIV. There is always an acute need for good quality ELISA kits. However, the quality evaluation data on Indian kits are very limited in comparison with internationally recognized kits. This study aimed to evaluate the performance and diagnostic usefulness of five commercially available ELISA kits which are frequently used in India.

Findings

The ELISA kits evaluated using an in-house well characterized 100 member sera panel revealed 100% sensitivity for all the batches. However, batch to batch variation in terms of specificity, positive predictive value (PPV) and efficiency, although not statistically significant (p > 0.05), was observed. For specificity, the 3rd generation kits (mean 99.6% to 99.3%) were comparatively better than the 4th generation assays (97.2% to 96.9%). But the 4th generation kits performed far better in the ability for early detection post HIV infection in the 25 member commercial seroconversion panel with a margin of at least 22 days and as high as 35 days than the 3rd generation assays.

Conclusions

The commercial ELISA kits with 100% sensitivity seem appropriate for HIV screening. The ability of early detection post HIV infection favors use of 4th generation kits for ensuring HIV free blood for transfusion. Lot to lot variations, especially kits having the specificity level ≤98.0%, indicate the need for a regular mechanism of kit evaluation for each batch for procuring kits appropriate for intended use.

Findings

Background

HIV is a major global public health issue [1]. For assuring a safe blood supply and preventing HIV infection, proper and accurate detection of HIV is essential [2]. In India, diagnosis of HIV infection is a major challenge [3, 4]. Several commercial assays are available for detection of HIV infection. ELISA is the most commonly used screening assay for HIV [2, 5]. A number of ELISA kits for HIV detection with different principles are available. Nowadays, in India 3rd generation ELISA are most commonly used. The 4th generation assays are based on combined detection of antigen and antibodies simultaneously and reduce the diagnostic window period further, compared to third generation ELISA which is based on anti-HIV antibody assay [68]. The improved sensitivity for ELISA is mostly accompanied by a decreased specificity. In an Indian perspective, limited articles on evaluation and performance of ELISA kits are available [9] though HIV testing is being done for a vast numbers of individuals as well as large number of specimens for ensuring HIV free safe blood for transfusion. Being the 2nd most populous country with the 3rd largest burden of HIV in the world [10], the magnitude of HIV testing challenge in India is enormous and the appropriate response to the challenge is to ensure the quality of the assay kits suitable for the intended use. This study aims to evaluate the quality of commonly available commercial ELISA kits for their ability to detect HIV suitable for appropriate use in India.

Materials and methods

The study was carried out at a National HIV Reference Laboratory designated for evaluation of diagnostic kits, including ELISA, in India. A well characterized, 100 members, in-house HIV serum panel was used to evaluate and compare the performance of the kits. The sera used for preparing the in-house panel were obtained anonymously from attendees of the Counseling and Testing Centre by taking informed consent as per the protocol approved by Institutional Ethical Committee of National Institute of Cholera and Enteric Diseases. Beside the negative and positive sera, this serum panel also contained low positive sera that have shown uniform results in all assays used for validation. The characterization of in-house panel was done using United States Food and Drug Administration (U.S. FDA) or Indian Central Drug Standard Control Organization (CDSCO) approved kits (Table 1). Samples non-reactive for all assays were defined as negative and reactive for all assays were defined as positive member in the panel. A commercial seroconversion panel (Lot# RP-018, Bio-Rad Laboratories, U.S.A) used to evaluate kits consists of a series of 25 specimens collected from an individual infected with HIV undergoing seroconversion.

Table 1 Details of kits used for characterization of in-house panel sera

Five batches each of 5 commonly available commercial ELISA kits, including 3rd and 4th generations, for HIV detection were evaluated (Table 2). All the kits were tested and results were validated strictly adhering to manufacturers’ instruction. The evaluation process maintained an unbiased method following a double blind procedure by using different personnel for pre-analytical and analytical testing sections. The final analysis of results with interpretation was done by the laboratory in-charge. The status of samples was unknown to the persons involved in pre-analytical and analytical procedures. The performance of kits was evaluated and compared in terms of sensitivity ([TP/(TP + FN)]×100), specificity ([TN/(TN + FP)]×100), positive predictive value (PPV = [TP/(TP + FP)]×100), negative predictive value (NPV = [TN/(TN + FN)]×100) and efficiency ([(TP + TN)/(TP + FN + TN + FP)]×100), where TP = number of true positives, TN = number of true negatives, FP = number of false positives and FN = number of false negatives [11]. Confidence Interval (CI) was used to address precision of the proportion estimates and the degree of confidence was set to 95% [12]. Chi-square analysis was performed to assess the variation for specificity, PPV and efficiency among different kits as well as in different batches of same kit.

Table 2 Performance of HIV ELISA kits with In-house Sera Panel

Results

None of the ELISA kits evaluated was able to identify all the panel members correctly by all the batches (Table 2). But all the kits were found to be 100% sensitive in all the batches. Variation in batches of all the kits was evident in terms of specificity, PPV and efficiency (Table 3). ERBA LISA HIV 1 + 2 provided correct results in 4 batches by identifying all panel samples correctly with 99.8% efficiency. Microlisa HIV and Enzaids HIV 1 + 2 both performed equally by correct result only in 3 batches with 99.6% efficiency. The 4th generation kits, Genscreen and Vironostika, showed the false positivity rates higher than the 3rd generations, but the variations were not statistically significant in terms of specificity (χ 2 = 0.0683, df = 4, p > 0.05), PPV (χ 2 = 0.1253, df = 4, p > 0.05) and efficiency (χ 2 = 0.0230, df = 4, p > 0.05) of all batches of all the ELISA kits. The performance of the kits evaluated using seroconversion panel revealed that all the 3rd generation kits showed equal sensitivity by detecting HIV positivity. In contrast, the 4th generation kits, Genscreen (detected panel member 5, day 16) and Vironostika (detected member 7, day 29) were significantly more sensitive and were able to detect HIV positivity 35 and 22 days earlier respectively than the 3rd generation ELISA kits (Table 4).

Table 3 Performance characteristic of HIV ELISA kits used for comparative evaluation
Table 4 Performance of HIV kits with Seroconversion Panel Sera (Lot# RP-018)

Discussion

ELISA is the type of test most commonly used for detection of HIV particularly for large numbers of specimens [2]. But the discordance between the results of different ELISA kits as well as in different lots of the same kit (particularly false positive rates), as evident in this study, highlights an important problem of potentially causing stress to falsely-positive individuals and may also lead to additional expenses [13, 14]. Hence, evaluation of diagnostic ELISA kits gains importance for ensuring the availability of suitable kits with better performance in terms of recommended sensitivity and specificity [11], as in case of blood bank testing where a high degree of sensitivity is also recommended for choosing the testing kit [11]. Better performance, comparatively, was observed for ERBALISA kits with 100% efficiency in 4 out of 5 batches (mean efficiency 99.8%). Microlisa and Enzaids each showed mean efficiency of 99.6%. The performance of Genscreen and Vironostika was compromised in terms of specificity, PPV and efficiency as these kits give few false positive results. The PPV as estimated based on the composition of panel sera will change according to the prevalence in the targeted population to be tested [9]. Thus, in an Indian scenario with 0.27% HIV prevalence [15], the estimated PPV of the HIV ELISA kits would be 40.30% for both J.Mitra & Co. Pvt. Ltd. and Span Diagnostics Ltd., 57.40% for Transasia Bio-Medicals Ltd. and 14.40% for Bio-Rad Laboratories. The performance of all five kits in terms of NPV favors their use as a primary screening assay for HIV infection. Unique combination of simultaneous Ag-Ab assay gives better performance in a seroconversion panel as well as reducing the testing window period by 3.82 days on average [6]. In this study, Genscreen and Vironostika, both 4th generation assays, outperformed the 3rd generation ELISA by reducing the window by 35 and 22 days respectively in the seroconversion panel. Genscreen showed better performance and identified early seroconversion than Vironostika in another study [16]. The 3rd generation kits demonstrated efficiency ranging from 98.6 to 99.8%. Though the efficiency of the 4th generation assays is lower, the high sensitivity demonstrated by the kits may favor them for HIV screening purposes, because of their early detection by about 3 weeks post infection. Although lot to lot variation is evident in the study it is not statistically significant (p > 0.05), but kits with specificity level <98% are not recommended for diagnosis of HIV infection in India according to the national guideline [11].

The panel size is small with 100 members and one seroconversion panel only. Kit evaluation with small panel size may be valuable where studies are limited [17]. Use of a range of seroconversion panels is essential to test the biological differences in the timing of appearance of different antibodies to specific antigens in the host response to the various HIV antigens in a range of individuals. A more robust assessment requires the testing of, for example, 10 different seroconversion panels [18, 19].

Conclusion

With 100% sensitivity, both the 3rd and 4th generation commercial ELISA kits seem are appropriate as screening assay for detection of HIV infection. Earlier detection post HIV infection favors the use of 4th generation kits for ensuring HIV free blood for transfusion. The lot to lot variation in terms of specificity warrants batch pre-acceptance testing of all new lots or batches of commercially available ELISA kits in India to ensure that new batches perform as well as previous ones.

Abbreviations

AIDS:

Acquired Immuno Deficiency Syndrome

CDSCO:

Central Drug Standard Control Organization

ELISA:

Enzyme Linked Immunosorbent Assay

HIV:

Human Immunodeficiency Virus

NPV:

Negative Predictive Value

PPV:

Positive Predictive Value

USFDA:

United States Food and Drug Administration.

References

  1. Joint United Nations Programme on HIV/AIDS (UNAIDS): UNAIDS World AIDS Day Report. 2011, Switzerland: UNAIDS

    Google Scholar 

  2. Beelaert G, Vercauteren G, Fransen K, Mangelschots M, Rooy MD, Garcia-Ribas S, Groen G: Comparative evaluation of eight commercial enzyme linked immunosorbent assays and 14 simple assays for detection of antibodies to HIV. J Virol Methods. 2002, 105: 197-206.

    Article  PubMed  CAS  Google Scholar 

  3. Ribeiro RR, Da Silva LF, Cunha CB, Cabral VP, Dietze R: Performance characteristics of a raid new immunochromatographic test for detection of antibodies to human immunodeficiency virus. Clin Diag Lab Immunol. 2003, 10: 303-307.

    Google Scholar 

  4. Iqbal HS, Solomon S, Murugavel KG, Solomon SS, Balakrishnan P: Evaluation of two indigenous rapid and two ELISA assays for the diagnosis of HIV infection India. Indian J Medical Microbiol. 2012, 30 (4): 397-402.

    Article  CAS  Google Scholar 

  5. Crofts N, Maskill W, Gust ID: Evaluation of enzyme-linked immunosorbent assays: a method of data analysis. J Virol Methods. 1988, 22: 51-59.

    Article  PubMed  CAS  Google Scholar 

  6. Weber B, Berger A, Rabenau H, Doerr HW: Evaluation of a new combined antigen and antibody human immunodeficiency virus screening assay, VIDAS HIV DUO Ultra. J Clin Microbiol. 2002, 40: 1420-1426.

    Article  PubMed  CAS  PubMed Central  Google Scholar 

  7. Ly TD, Laperche S, Couroucé AM: Early detection of human immunodeficiency virus infection using third- and fourth-generation screening assays. Eur J Clin Microbiol Infect Dis. 2001, 20: 104-110.

    PubMed  CAS  Google Scholar 

  8. Saville RD, Constantine NT, Cleghorn FR, Jack N, Bartholomew C, Edwards J, Gomez P, Blattner WA: Fourth-generation enzyme- linked immunosorbent assay for the simultaneous detection of human immunodeficiency virus antigen and antibody. J Clin Microbiol. 2001, 39: 2518-2524.

    Article  PubMed  CAS  PubMed Central  Google Scholar 

  9. Maity S, Nandi S, Biswas S, Sadhukhan S, Saha MK: Performance and diagnostic Usefulness of commercially available enzyme linked immunosorbent assay and rapid kits for detection of HIV, HBV and HCV in India. Virol J. 2012, 9: 290-

    Article  PubMed  PubMed Central  Google Scholar 

  10. Alvarez-Uria G: Factors associated with delayed entry into HIV medical care after HIV diagnosis in a resource-limited setting: Data from a cohort study in India. Peer J. 2013, 1: e90-e90 http://dx.doi.org/10.7717/peerj.90

    Article  PubMed  PubMed Central  Google Scholar 

  11. National AIDS Control Organization (Ministry of Health and Family Welfare Govt of India): Manual on Quality Standard for HIV Testing Laboratories. 2007, New Delhi: NACO

    Google Scholar 

  12. Iqbal HS, Solomon S, Murugavel KG, Solomon SS, Balakrishnan P: Evaluation and diagnostic usefulness of domeatic and imported enzyme-linked immunosorbent assays for detection of human immunodeficiency virus type 1 antibody in India. Clin Diagn Lab Immunol. 2005, 12: 1425-1428.

    PubMed  CAS  PubMed Central  Google Scholar 

  13. Matter L, Germann D: Detection of human immunodeficiency virus (HIV) type 1 antibodies by new automated microparticle enzyme immunoassay for HIV type 1 and 2. J Clin Microbiol. 1995, 33: 2338-2341.

    PubMed  CAS  PubMed Central  Google Scholar 

  14. Pyra H, Boni J, Schupbach J: Ultrasensitive retrovirus detection by a reverse transcriptase assay based on product enhancement. Proc Natl Acad Sci U S A. 1994, 91: 1544-1548.

    Article  PubMed  CAS  PubMed Central  Google Scholar 

  15. National AIDS Control Organization (Ministry of Health and Family Welfare, Government of India): Technical Report India HIV Estimates. 2012, New Delhi: UNAIDS and NACO

    Google Scholar 

  16. Ly TD, Martin L, Daghfal D, Sandridge A, West D, Bristow R, Laurence C, Qiu X, Lou SC, Hunt JC, Schochetman G, Devare SG: Seven Human Immunodeficiency Virus (HIV) antigen-antibody combination assays: evaluation of HIV seroconversion sensitivity and subtype detection. J Clin Microbiol. 2001, 39 (9): 3122-

    Article  PubMed  CAS  PubMed Central  Google Scholar 

  17. Torane VP, Shastri JS: Comparison of ELISA and rapid screening tests for the diagnosis of HIV, hepatitis B and hepatitis C among healthy blood donors in a tertiary care hospital in Mumbai. Indian J Med Microbiol. 2008, 26: 284-285.

    Article  PubMed  CAS  Google Scholar 

  18. Perry KR, Ramskill S, Eglin RP, Barbara JA, Parry JV: Improvement in the performance of HIV screening kits. Transfus Med. 2008, 18 (4): 228-240.

    Article  PubMed  CAS  Google Scholar 

  19. Barbara J, Ramskill S, Perry K, Parry J, Nightingale M: The National Blood Service (England) approach to evaluation of kits for detecting infectious agents. Transfus Med Rev. 2007, 21 (2): 147-158.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

We acknowledge NACO, New Delhi and West Bengal State AIDS Prevention and Control Society for partially supporting the study.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Malay Kumar Saha.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

All the authors conceived the study, SN, SM and SCB carried out the laboratory work, MKS analyzed and interpreted the data, SM performed the statistical analysis and all the authors contributed in drafting, and critically reviewed and approved the final manuscript.

Rights and permissions

Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Nandi, S., Maity, S., Bhunia, S.C. et al. Comparative assessment of commercial ELISA kits for detection of HIV in India. BMC Res Notes 7, 436 (2014). https://doi.org/10.1186/1756-0500-7-436

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1756-0500-7-436

Keywords