| Design and implementation of GRIP: a computerized glucose control system at a surgical intensive care unit1Surgical Intensive Care Unit, Department of Surgery, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands 2Department of Cardiology, Thoraxcenter, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands
BMC Medical Informatics and Decision Making 2005, 5:38doi:10.1186/1472-6947-5-38 The electronic version of this article is the complete one and can be found online at: http://www.biomedcentral.com/1472-6947/5/38
©
2005 Vogelzang et al; licensee BioMed Central Ltd. AbstractBackgroundTight glucose control by intensive insulin therapy has become a key part of critical care and is an important field of study in acute coronary care. A balance has to be found between frequency of measurements and the risk of hypoglycemia. Current nurse-driven protocols are paper-based and, therefore, rely on simple rules. For safety and efficiency a computer decision support system that employs complex logic may be superior to paper protocols. MethodsWe designed and implemented GRIP, a stand-alone Java computer program. Our implementation of GRIP will be released as free software. Blood glucose values measured by a point-of-care analyzer were automatically retrieved from the central laboratory database. Additional clinical information was asked from the nurse and the program subsequently advised a new insulin pump rate and glucose sampling interval. ResultsImplementation of the computer program was uneventful and successful. GRIP treated 179 patients for a total of 957 patient-days. Severe hypoglycemia (< 2.2 mmol/L) only occurred once due to human error. With a median (IQR) of 4.9 (4.2 – 6.2) glucose measurements per day the median percentage of time in which glucose fell in the target range was 78%. Nurses rated the program as easy to work with and as an improvement over the preceding paper protocol. They reported no increase in time spent on glucose control. ConclusionA computer driven protocol is a safe and effective means of glucose control at a surgical ICU. Future improvements in the recommendation algorithm may further improve safety and efficiency. BackgroundCritically ill patients often suffer from 'stress hyperglycemia', a condition in which insulin resistance due to increased catecholamine levels causes high blood glucose values [1]. The association between stress hyperglycemia and adverse outcome has been observed in numerous patient categories, ranging from patients admitted to the general ward [2] to myocardial infarction [3] and stroke patients [4]. For decades, stress hyperglycemia was thought to be merely a marker of disease, and was tolerated as long as glucose levels were not excessively high (e.g., over 11.0 mmol/L). The publication of the Leuven intensive insulin therapy study in September 2001 caused a paradigm shift in critical care medicine [5]. This study showed that rapid lowering of blood glucose levels below 6.1 mmol/L and subsequent maintaining of normoglycemia reduce mortality and morbidity markedly. These results were confirmed by a before-after study performed by Krinsley, in which also a decrease in mortality was achieved with tight glucose control [6]. Especially for septic patients, guidelines now recommend using insulin to reduce high glucose levels [7]. However, infusing insulin in order to lower glucose levels bears the risk of inducing life-threatening hypoglycemia, especially in sedated patients admitted to an intensive care unit (ICU). In order to cut back this risk, glucose levels must be frequently measured. Each measurement calls for a decision on what action to take to keep glucose levels in the normal range. With recommended sampling frequencies ranging from every 1 – 2 hours to every 6 hours, implementation of tighter glucose control poses an important logistic challenge. Many investigators have proposed nurse-driven protocols for glucose control. After each glucose measurement, simple if-then rules or lookup tables yield an advice on how much insulin needs to be administered [8-10]. Even though glucose sampling frequency is high, reduction of hyperglycemia is often not satisfactory, and, more important, hypoglycemia is relatively common [9]. Glucose metabolism is also an important topic in acute coronary care. Several studies have evaluated glucose-related therapies as strategies to improve outcome in acute coronary syndromes, such as high dose glucose-insulin-potassium (GIK) infusion [11,12], or combined glucose-insulin infusion to reduce glucose levels [13,14]. Although clinical results have been mixed, with most recent results being negative, the efficacy of glucose-lowering interventions in acute coronary care is still unknown since none of the published trials achieved tight glycemic control [11,14]. Because coronary care units (CCU) are a less controlled environment with a lower personnel-to-patient ratio than ICUs, intensive insulin therapy is hard to achieve with paper protocols [14]. An optimized decision making algorithm might be able to make tighter control possible. We hypothesized that a computer program can employ the necessary complex logic to achieve the desired level of both safety and efficiency of glucose control without excessive glucose sampling frequencies. In the beginning of 2003 we initiated development of a computer controlled decision support system. MethodsDesign rationaleWe conceived the computer decision support system (CDSS) to be used primarily by nurses. Our first implementation site was our 12-bed surgical ICU at a tertiary teaching hospital, and we planned for future implementation at other ICUs and at the CCU. A thorough analysis of factors which lead to successful CDSS deployment and usage has previously shown integration of the system into routine clinical workflow to be of paramount importance [15]. Automatic provision of a concrete recommendation instead of merely an assessment of the situation, and provision of decision support at the time and place of decision making were other elements that were highly predictive of successful implementation. Figure 1 shows the different steps a nurse takes when performing glucose control. Repetitive glucose measurements are followed by appropriate action to maintain normoglycemia. At our surgical ICU, as well as at the CCU, blood glucose values were determined by a point-of-care blood gas analyzer, giving a very short delay between blood sampling and availability of the glucose value (~2 minutes). As most nurses wait at the point-of-care analyzer until the result of the analysis is known, a computer situated next to the analyzer was the ideal spot for the system to interact with the nurses and to give its recommendations. This ensured an easy integration in standard clinical workflow, and made sure that the recommendations were provided to the responsible nurse at the place and time when needed.
The project was named GRIP (Glucose Regulation for Intensive care Patients). One of the key design principles was that the output should be regarded as recommendations, not as orders. The recommendations that GRIP gives could be overruled at any time. If overruled, GRIP continues to give meaningful advice the next time (as long as it knows that its recommendation was not followed, and what action was taken instead). In this way, GRIP remains usable in various unforeseen situations. Technical designA high-level overview of GRIP's design is shown in figure 2. To enhance safety and facilitate user acceptance, all information that was available in the central hospital database was queried from there and not from the nurse. To ensure correct glucose values, each glucose value that was retrieved from the laboratory system required a validation from the responsible nurse. In this way, measurement errors, patient swaps, and other mistakes were prevented from disturbing the decision making of GRIP. After each glucose measurement, a number of clinical variables were queried from the nurse (table 1). The communication with the hospital database was designed using standard communication protocols such as Health Level 7 (HL7). Data collected and computed by GRIP, as well as error messages, were stored in a relational database using standard query language (SQL) queries. Being in a database system, the data was easily accessible for special-purpose applications, such as a report generator for periodic auditing of quality of glucose control or other aspects of GRIP, or for a data extraction program for purposes of research. Most data in the database was stored as pieces of eXtensible Markup Language (XML), which facilitates extension of data storage in the future.
Table 1. Data input into GRIP Recommendation generation algorithmRecommendations generated by GRIP mainly consist of two parameters: the recommended insulin pump rate and the time at which the next blood sample should be taken. Also, in special cases (currently on occurrence of glucose values > 15 or < 4 mmol/L) the recommendation is accompanied by a request to consult the attending physician. On low glucose levels (< 3 mmol/L) the 'hypoglycemia advice' is given, which consists of an intravenous glucose dose, an insulin pump rate of zero, and prompt notification of the attending physician. Our current implementation of GRIP is based on the use of continuous insulin infusion and does not recommend boluses of insulin. It has been observed that due to a saturation effect, continuously infused insulin is more effective than boluses of insulin [16]. We will briefly describe the algorithm for pump rate. We hypothesized that by using solely the most recent glucose value, efficient glucose control is hard, if not impossible, to achieve, and chose to include rate of change of glucose as secondary parameter. The basis for our pump rate algorithm is the following formula: ΔI = (1 + 0.25· In this formula, ΔI is the proposed change in pump rate, The algorithm that calculates the desired time of the next glucose measurement is considerably more complex. Its main feature is a calculated measure conceived to quantify the risk of hypoglycemia. The following measures lead to a higher predicted risk of hypoglycemia, and consequently to a shorter interval to the next advised glucose sample: high insulin pump rate, high recommended pump increase, fast glucose decrease (as measured by the extrapolated glucose in 4 hours, derived from the current glucose value and the current User interfaceA graphical user interface was designed to both acquire clinical information from the nurse and communicate advice to the nurse. To make user input as simple as possible, all filled out information is automatically filled in with the previous values so that only changes require user action. Figure 3 shows the main screen of GRIP. This screen provides a quick overview of all patients. Figure 4 shows the screen providing details for one patient. The required action which a nurse should take was divided in three steps:
1. Take a blood sample and analyze it. Clicking the button in GRIP makes GRIP query the hospital information system to see if it can retrieve the new value. 2. Review the current information of a patient and update this information if changes have occurred. This step also includes validating each newly acquired glucose measurement. 3. Review the insulin pump advice given by GRIP and tell GRIP whether it was accepted and executed or, if not, what the new insulin pump rate is. These three steps normally follow each other in order, but each step is also immediately accessible without following the previous steps (e.g., a physician who changes an insulin pump could immediately tell this to GRIP by taking step 3, without the need to first take blood, etc.) Hardware and availabilityThe program was designed to run on a standard commodity personal computer, situated directly next to the point-of-care glucose analyzer for optimal integration in the clinical workflow. The program was developed in Java, a freely available multi-platform language. Because MySQL® (a freely available open-source database server) was chosen as the database system, the entire system can be run using only free software. Upon completion of a new feature-complete version scheduled for the beginning of 2006, GRIP will be released under the open source Gnu Public License [20], and will be downloadable from http://grip-glucose.sourceforge.net webcite. This means GRIP can be used by anyone free of charge. Evaluation of the systemGlucose control before implementation of GRIP was performed by a sliding-scale paper protocol. This protocol advised an insulin pump rate which was based on the last measured glucose level. Measurements were taken 4 times a day at a fixed schedule. When needed, the scheme was adjusted for individual patients on a daily basis. The medical staff of the ICU was informed and unanimously agreed to gradually implement GRIP. Since the move from conventional practice to GRIP constituted an organizational and logistic change that aimed to improve existing and accepted medical practice, approval by the medical ethical committee or patient consent was deemed unnecessary. To get a basic idea of control achieved by the system, we recorded all glucose measurements of all patients during the first 4 month period in which GRIP was in effect. We also retrieved age, reason of admission, history of diabetes, APACHE II score, and both ICU and hospital mortality for these patients. Safety of control was assessed by checking all patients for occurrences of severe hypoglycemia (< 2.2 mmol/L) or mild hypoglycemia (< 3.5 mmol/L). To assess the ability to reduce hyperglycemia, all patients staying longer than 24 hours were analyzed. As the target level was 6.5 mmol/L, we defined a target glucose range of 4–7.5 mmol/L. For each patient, we determined how much time the glucose levels were within this target range and we represented this value as the fraction of the entire length of stay. We determined the time from admission to the first glucose under 7.5 mmol/L, and calculated median and interquartile range for glucose levels at 6-hour intervals for the first two days. Finally the hyperglycemic index, a measure indicative of overall hyperglycemia that is not lowered by hypoglycemic episodes, was calculated for each patient [21]. Patients at the ICU were enterally fed as soon as possible. When enteral feeding was not possible for prolonged periods, total parenteral nutrition or concentrated glucose infusions were started at moderate doses (100–200 grams per day). User acceptance was assessed by a questionnaire. Nurses were asked to fill out the questionnaire 1 month before and 6 months after implementation of GRIP. The questionnaire held 6 months after implementation of GRIP contained a number of questions asking for a direct comparison between working with GRIP and working with the paper protocol. Statistical analysisFor the nurse questionnaires, a 7 point scale of agreement was used. Differences before and after usage of GRIP were tested with the Mann-Whitney U test. All data are presented as median (interquartile range). Statistical analysis was performed using R version 2.1.0 [http://www.r-project.org webcite]. ResultsImplementation detailsImplementation of GRIP began in the beginning of 2003. In the fall of 2003 an initial version of GRIP was tested for three weeks at the surgical ICU without nurses or doctors following its recommendations. Nurse feedback indicated that this test version's user interface was insufficiently intuitive. In November 2004 the final version of GRIP, including the improved user interface as described in the methods, was deployed. First it was monitored for 6 weeks, while the regular paper protocol was still followed. In the next two weeks we started following GRIP's recommendations, gradually increasing its usage bed-by-bed. From January 2005 on, insulin therapy of patients on all beds was determined by GRIP. During the first weeks of deployment a number of small changes to solve trivial problems were needed. Thereafter, two major problems have occurred. First, the computer running GRIP experienced a sudden complete hard-disk crash in April 2005, losing all data. Using our automatic daily backup, the system was up and running again within 6 hours, with less than 24 hours data loss, which was easily reentered into GRIP from the charts. The second problem was the transition to daylight savings time. This was handled correctly by GRIP, but unfortunately the central hospital system only picked up the correct time later during the next day. Before the central system was adjusted to the proper time, all measurements appeared to GRIP one hour later than they actually were. Apart from inconvenience to the nurses, these problems caused no misleading recommendations or dangerous situations. Currently, the system has been running without interruption for over 3 months without any change, maintenance, crash or other problem. A number of other problems have occurred, all external to GRIP, which would have affected a paper protocol equally as much as it affected GRIP. These included measurements that failed to be included in the central hospital database and malfunction of the point-of-care analyzer. After these problems were resolved and GRIP started receiving correct data again, its recommendations could immediately be followed, because GRIP ignored the erroneous data. Training of nursesBasic data entry in GRIP proved to be very straightforward. Only data entry of events that happened in the past was not completely obvious. During the first 2 month observation period, residents were asked to perform data entry and as no recommendations were used, they could experiment and save any questions until the lead programmer (MV) was present at the ICU. When real use of GRIP by nurses was initiated, almost all residents knew how to work with GRIP and could provide explanation to nurses. The staff intensivist (MN) involved in the design of GRIP also provided explanation when asked. No special meetings for training had to be scheduled. Safety and efficiencyDuring the 4-month period from January 1st to May 1st, 2005, 179 patients were treated at our ICU, for a total of 957 patient-days. Patient characteristics are shown in table 2. Severe hypoglycemia (< 2.2 mmol/L) occurred once in one patient (0.6 %). Analysis revealed that this was caused by a human error involving accidentally increasing the rate of the insulin pump. Mild hypoglycemia (< 3.5 mmol/L) occurred in 20 patients (11.2 %). 109 out of 179 patients stayed longer than one day at the ICU. Data regarding hyperglycemia in these patients is presented in table 3. Glucose levels met the target range for more than three-quarters of the time in 66 patients (61 %). The glucose sampling frequency was less than 6 times a day in 76 patients (70 %). The median hyperglycemic index indicates that the median patient had a time-weighted glucose level of 6.9 mmol/L. The maximum insulin dose of 10 units/hour was needed in 5 patients (5 %). At the time of maximum insulin infusion, the glucose values of these patients were 8.3, 8.8, 9.3, 14.9 and 17.6 mmol/L. Figure 5 shows median and interquartile range of glucose levels during the first 48 hours of ICU stay. Table 2. Patient characteristics Table 3. Glycemic control
User acceptanceQuestionnaires were filled out by 32 nurses before implementation of GRIP and by 22 nurses after implementation. After deployment of GRIP, nurses judged glucose control to be running more smoothly than before deployment of GRIP (p < 0.001, figure 6). There was a trend towards nurses finding that glucose values are less often too high or too low. GRIP was judged as being simple to work with, and chosen almost unanimously as an improvement over the old paper protocol. Estimated time spent on glucose control per working shift was 10 (6 to 12) minutes before GRIP and 10 (10 to 15) minutes after GRIP (p = 0.13). Analysis of paired observations before and after implementation of GRIP revealed similar results. From the free text section of the questionnaire it was clear that the nurses found that GRIP was able to adequately manage glucose control in more patients than the paper protocol. Nurses therefore did not need to call the attending physician as often as before and thus were able to devote more time to other tasks.
DiscussionIn this report we present GRIP, a computer decision support system for glucose control by intensive insulin therapy. This system was successfully implemented at a surgical ICU, and was found to provide safe and efficient glycemic control. User acceptance, an important hurdle for successful implementation of a clinical decision support system, was excellent with minimal training. Nurses clearly rated GRIP as an improvement over the conventional sliding scale protocol. GRIP will be released under an open source license and will thus be free to use and improve by anyone in the future. We think especially GRIP's independence of a patient data management system (PDMS) is a strong point that makes it widely usable. At ICUs where a PDMS is already operational, GRIP might introduce double data entry. However, this can be easily resolved by either letting GRIP query the PDMS for information or by converting GRIP's advice module into a plug-in of the PDMS. The efficiency and safety of control achieved by GRIP's current relatively simple algorithm was satisfactory. In the Leuven study, a mean glucose level of 5.7 mmol/L was achieved in a cohort of patients with a median APACHE II score of 9 [5]. Krinsley achieved a mean of 7.3 mmol/L in a sicker cohort of patients (median APACHE of 15) [6]. Our group had APACHE scores similar to Krinsley's, and a mean glucose level that was 0.4 mmol/L lower. With regard to hypoglycemia, in the Leuven study 39 out of 765 patients (5.1 %) had one or more values lower than 2.2 mmol/L, compared to 1 out of 179 (0.6 %) in our group. Krinsley describes the number of hypoglycemic episodes as the proportion of total number of glucose measurements, instead of as the proportion of patients at risk. In Krinsley's study, 0.34 % of measurements were lower than 2.2 mmol/L, compared to 0.02 % in our group. During the design of the current recommendation algorithm a number of arbitrary decisions were taken. Based on experience, we chose to include glucose value and glucose difference as the two most important determinants of advised pump rate, and we chose to limit pump rate to 10 units per hour. Although evidence for the rationality of the latter decision exists [16,19], most decisions were taken purely based on prior experience in controlling glucose levels. Our results show that this algorithm gives satisfactory results, but we are aware that GRIP has the potential for substantial improvements in the future, as our choices are unlikely to be the most optimal. For example, the current algorithm is very cautious and only takes small steps when increasing the insulin pump rate, sacrificing rapid control in favor of safety. Analysis of the data that have been collected thus far may reveal situations in which the algorithm can safely take larger steps to achieve more timely glucose control. It will be interesting to analyze differences between groups of patients with respect to factors such as reason of admission, presence of diabetes, and concomitant drug use. This will allow glucose control to transition from the current general "one size fits all" approach, which uses the same advice for every patient, to a more tailor-made approach, which makes use of as much information as possible to make its advice fit an individual patient's needs. As the advice generation module can be changed without affecting the user interface, algorithm improvements can be implemented without additional training of the nurses. Prior studies have evaluated different types of computer systems for glucose control in critically ill patients. A number of studies have evaluated feasibility of continuous glucose sensors, which in theory provide ultimate efficiency and safety of control. However, apart from the obvious extra costs, continuous sensors have been found to lack reliability, requiring frequent replacements of the sensor, and report inaccurate data in the hypoglycemic range [22,23]. We believe that for the foreseeable future, the vast majority of glucose control schemes will still employ sequential discontinuous measurements. Chase et al. have worked on developing models for accurate insulin dose prediction in critically ill patients, but with 2 measurements per hour, measurement frequency was much higher than GRIP's [19,24]. Rood and colleagues have previously investigated computerizing guidelines for glucose control and found that glucose measurements were taken at the prescribed time more often, and glucose control improved in comparison with the preceding paper protocol [25]. Unfortunately, a PDMS plug-in with an undisclosed 4-page flowchart was used, limiting widespread usability, and in our opinion a formula-based approach as taken by GRIP will be easier to optimize and customize to different patient groups than a flowchart-based set of rules. Naturally, in case the latter approach may prove to yield better control, GRIP's advice generation module can be easily altered to follow that approach. We are currently in the process of starting implementation of GRIP at other ICUs. Furthermore, we are implementing GRIP in the coronary care unit. We consider a personalized algorithm to be the key to more efficient and safe glucose control in acute coronary care. ConclusionGRIP, a computer decision support system for glucose control by intensive insulin therapy, exhibited efficient glucose control without inducing severe hypoglycemia during a 4 month period. Acceptance by nurses was excellent, with minimal training needed. GRIP will be released as free/open source software. Competing interestsThe author(s) declare that they have no competing interests. Authors' contributionsMV is the main designer and developer of the GRIP computer program, and was responsible for instructing nurses, taking questionnaires, and data analysis. FZ helped extending the program to the cardiologic domain. MN co-designed the GRIP program and the advice algorithm and is primary investigator of the project. All authors were involved in drafting the manuscript and agree to its publication. AcknowledgementsThe authors want to thank the residents and nurses of the surgical ICU for their excellent cooperation and useful feedback during the development of GRIP. We are grateful to the local ICT department for their help establishing the link between GRIP and the central hospital database. References
Pre-publication historyThe pre-publication history for this paper can be accessed here: http://www.biomedcentral.com/1472-6947/5/38/prepub Have something to say? Post a comment on this article! |