Pharmacokinetics of recombinant human growth hormone administered by cool.click™ 2, a new needle-free device, compared with subcutaneous administration using a conventional syringe and needle

Chris Brearley1, Anthony Priestley2, James Leighton-Scott2 and Michel Christen3*

Author Affiliations

1 Clinical Research, Serono International SA, 1211 Geneva, Swizerland

2 LCG Bioscience, Bourn Hall, Bourn, Cambridge CB3 7TR, UK

3 Patients Care Technologies Center of Expertise, Merck Serono International SA, 1211 Geneva, Switzerland

For all author emails, please log on.

BMC Clinical Pharmacology 2007, 7:10  doi:10.1186/1472-6904-7-10

Published: 8 October 2007

Abstract

Background

Growth hormone (GH) is used to treat growth hormone deficiency (GHD, adult and paediatric), short bowel syndrome in patients on a specialized diet, HIV-associated wasting and, in children, growth failure due to a number of disorders including Turner's syndrome and chronic renal failure, and in children born small for gestational age. Different brands and generic forms of recombinant human growth hormone (r-hGH) are approved for varying indications in different countries. New ways of administering GH are required because the use of a needle and syringe or a device where a patient still has to insert the needle manually into the skin on a daily basis can lead to low adherence and sub-optimal treatment outcomes. The objective of this study was to assess the relative bioavailability of r-hGH (Saizen®, Merck Serono) administered by a new needle-free device, cool.click™ 2, and a standard needle and syringe.

Methods

The study was performed with 38 healthy volunteers who underwent pituitary somatotrope cell down-regulation using somatostatin, according to a randomized, two-period, two-sequence crossover design. Following subcutaneous administration of r-hGH using cool.click™ 2 or needle and syringe, pharmacokinetic parameters were analysed by non-compartmental methods. Bioequivalence was assessed based on log-transformed AUC and Cmax values.

Results

The 90% confidence intervals for test/reference mean ratio of the plasma pharmacokinetic variables Cmax and AUC0-inf were 103.7–118.3 and 97.1–110.0, respectively, which is within the accepted bioequivalence range of 80–125%. r-hGH administered by cool.click™ 2 is, therefore, bioequivalent to administration by needle and syringe with respect to the rate and extent of GH exposure. Treatment using cool.click™ 2 was found to be well tolerated. With cool.click™ 2 the tmax was less (3.0 hours) than for needle and syringe delivery (4.5 hours), p = 0.002 (Friedman test), although this is unlikely to have any clinical implications.

Conclusion

These results demonstrate that cool.click™ 2 delivers subcutaneous r-hGH exposure that is bioequivalent to the conventional mode of injection. The new device has the additional advantage of being needle-free, and should help to increase patient adherence and achieve good therapeutic outcomes from r-hGH treatment.