Resolution: standard / high Figure 2.
Mean reduction (log10colony-forming units [CFU]/cm2) in recovery of multiple strains of Clostridium difficile, methicillin-resistant Staphylococcus aureus (MRSA), and vancomycin-resistant Enterococcus (VRE) from laboratory bench top surfaces after the use of the Tru-D device. For each pathogen, the inoculum applied to the bench top was adjusted such that
103 to 105 CFU were recovered from the positive control specimens. The Tru-D device was operated
at a reflected dose of 22,000 μWs/cm2 for ~45 minutes.
Nerandzic et al. BMC Infectious Diseases 2010 10:197 doi:10.1186/1471-2334-10-197 |