Table 1 |
||
|
Desflurane consumption (cumulative dose, mL liquid) was higher at all times (P < 0.05) during automated closed-circuit anesthesia administration (Zeus®) than with a conventional machine used with the predefined low flow sequence (ADU®). |
||
|
Time (min) |
ADU group |
Zeus group |
|
|
||
|
0 |
0.0 ± 0.0 |
0.0 ± 0.0 |
|
5 |
6.0 ± 0.0 |
11.7 ± 4.0 |
|
10 |
7.1 ± 0.0 |
12.9 ± 4.1 |
|
15 |
8.2 ± 0.0 |
13.9 ± 4.2 |
|
20 |
9.1 ± 0.2 |
14.8 ± 4.5 |
|
25 |
10.0 ± 0.3 |
15.4 ± 4.3 |
|
30 |
11.0 ± 0.3 |
15.9 ± 4.5 |
|
35 |
11.9 ± 0.3 |
16.6 ± 4.6 |
|
40 |
12.8 ± 0.4 |
17.2 ± 4.8 |
|
|
||
|
De Cooman et al. BMC Anesthesiology 2008 8:4 doi:10.1186/1471-2253-8-4 |
||