Table 3

Table demonstrating haemodynamic measurements from the LiDCO and PulseCO monitors for up to eight hours. HR, MAP, CVP, LiDCO and PulseCO (median and IQR), bias and percentage error at measurements intervals.

Baseline

+1 hour

+ 2 hours

+4 hours

+8 hours


HR/min

104 (92–117)

110 (96–123)

111 (95–117)

107 (94–117)

107 (92–110)

MAP mmHg

81 (72–93)

84 (77–93)

86 (75–98)

81 (75–88)

81 (75–99)

CVP mmHg

16 (12–19)

16 (10–20)

13 (11–18)

16 (11–21)

15 (11–16)

LiDCO L/min

7.5 (5.1–9.0)

7.0 (5.0–10.6)

6.8 (5.6–10.1)

6.4 (5.3–9.8)

5.5 (4.9–6.7)

PulseCO L/min

#

7.5 (4.4–9.8)

8.6 (5.5–9.9)

7.2 (4.8–9.2)

6.1 (5.2–7.7)

Bias ± 2SD L/min

#

-0.3 ± 2.3

0.1 ± 1.9

-0.1 ± 2.2

0.2 ± 2.3

Percentage error

#

29%

22%

28%

36%


Cecconi et al. BMC Anesthesiology 2008 8:3   doi:10.1186/1471-2253-8-3

Open Data