Glucose prediction by analysis of exhaled metabolites – a systematic review

Abstract

Background

In critically ill patients, glucose control with insulin mandates time– and blood–consuming glucose monitoring. Blood glucose level fluctuations are accompanied by metabolomic changes that alter the composition of volatile organic compounds (VOC), which are detectable in exhaled breath. This review systematically summarizes the available data on the ability of changes in VOC composition to predict blood glucose levels and changes in blood glucose levels.

Methods

A systematic search was performed in PubMed. Studies were included when an association between blood glucose levels and VOCs in exhaled air was investigated, using a technique that allows for separation, quantification and identification of individual VOCs. Only studies on humans were included.

Results

Nine studies were included out of 1041 identified in the search. Authors of seven studies observed a significant correlation between blood glucose levels and selected VOCs in exhaled air. Authors of two studies did not observe a strong correlation. Blood glucose levels were associated with the following VOCs: ketone bodies (e.g., acetone), VOCs produced by gut flora (e.g., ethanol, methanol, and propane), exogenous compounds (e.g., ethyl benzene, o–xylene, and m/p–xylene) and markers of oxidative stress (e.g., methyl nitrate, 2–pentyl nitrate, and CO).

Conclusion

There is a relation between blood glucose levels and VOC composition in exhaled air. These results warrant clinical validation of exhaled breath analysis to monitor blood glucose levels.

Peer Review reports

Background

Many, if not all, critically ill patients are treated with insulin at some point during their stay in the intensive care unit (ICU) [1]. Intensive monitoring of the blood glucose level is a prerequisite for both efficient and safe insulin titration in these patients [2]. Current practice in the ICU holds that glucose levels are monitored manually through intermittent measurements of the blood glucose level in central laboratories or using laboratory–based blood gas analyzers and/or glucose strips at the bedside [3]. Intermittent manual glucose monitoring however, is expensive and time and blood consuming [4]. Moreover, intermittent glucose monitoring lacks the ability to detect temporal trends, potentially causing dangerous insulin titration errors in critically ill patients [5].

Glucose is a central molecule in metabolism [6, 7]. Indeed, metabolic pathways are activated to maintain normoglycemia when the concentration of glucose changes [6, 8]. Changes in the activity of these pathways could result in changes in production of volatile metabolites. These so–called volatile organic compounds (VOCs) can be detected in exhaled breath [9].

We hypothesize that there is an association between VOCs in exhaled breath and blood glucose levels. Previous excellent reviews focused on the correlation between glucose and exhaled breath condensate (thus soluble markers) [10] in diabetes [11, 12], but none compared all available literature or discussed the implications for the ICU population. The specific aim of this systematic review is to provide an overview of the available data on the association breath VOCs and blood glucose levels and to discuss techniques for VOC detection.

Methods

This systematic review was done according to standard methodology [13, 14]. Medline was searched through Pubmed using the following search terms: (“Blood Glucose”[Mesh] OR “Glucose”[MeSH Terms] OR glucose[tiab]) AND (“Exhalat*” [MeSH Terms] OR “Volatile Organic Compounds” [Mesh] OR exhal* [tiab] OR Volatile Organic Compound* [tiab] OR Volatile Compound* [tiab] OR electronic nose [tiab] OR breath [tiab]. The search was conducted on the 3rd of January 2014. No limits were used for year of publication and language. Only human studies were included, with no restriction on subject health, age, gender or study setting.

Two independent researchers (JHL, LDB) selected articles for full–text assessment when the title and abstract suggested investigating the use of exhaled breath to measure or estimate blood glucose levels. Articles were only included if an association between blood glucose levels and VOCs in exhaled air was investigated. Also, VOC compositions of exhaled air had to be measured by an analytical technique that allows for separation, quantification and identification of individual VOCs, including gas chromatography and mass spectrometry (GC–MS), ion mobility mass spectroscopy (IMS), ionmolecule reaction mass spectrometry (IMR–MS), proton transfer reaction (time of flight) mass spectrometry (PTR(−TOF)-MS) and/or selected ion flow tube mass spectrometry (SIFT–MS).

Data from included studies were extracted and methodological quality was assessed independently by two researchers (JHL, LDB) using the QUADAS–2 tool for quality assessment [15]. The tool was adapted to be more relevant to the included literature. Disagreement between the two reviewers on inclusion of studies was resolved by consensus. The adjusted version of QUADAS–2 is presented in Additional file 1. Extracted data included: 1) characteristics of the study (design, year of publication and country of study conduction); 2) characteristics of the study population (including age, sex distribution and health status); 3) characteristics of the index test (including technique and included VOCs); 4) characteristics of the reference standard (blood glucose); 5) characteristics of the outcome (including main results and correlation coefficient between exhaled breath and glucose); 6) statistical validation technique used.

Results

Search results

The literature search in Pubmed yielded 1041 titles (Figure 1). After reading titles and abstracts, 1012 articles were excluded because the topic was outside of the scope of this review and 29 articles were retained for full–text assessment. After the exclusion of 20 papers (5 reviews/non-original studies, 13 on unrelated topics, 2 index test not compliant with inclusion criteria), 9 articles were included in the analysis. Characteristics of selected articles are presented in Table 1. Five studies included healthy non-diabetic subjects, two studies included Type 1 Diabetes Mellitus (T1DM) subjects, one study included Type 2 Diabetes Mellitus (T2DM) subjects and one study included both healthy and T1DM subjects.

Figure 1
figure 1

Flow diagram.

Table 1 Characteristics of included studies

Results of the quality assessment using the QUADAS–2-tool are presented in Table 2. The risk of bias was considered high for all studies; none of the studies used a random sample of patients, all using a pre-specified target group such as T1DM or T2DM patients. The use of blood gas measurements or central lab measurements was considered to be the correct reference standard [16]. The adequate reference standard was used in four studies. Four studies used finger prick measurement, which increases the possibility of incorrect insulin titration in clinical practice [16]. Comparing these measurements to exhaled breath could lead to biased results. However, none of these studies were excluded from our review.

Table 2 Results of QUADAS-2 tool

Point correlation

Authors of seven out of nine studies found a strong correlation between one or more metabolites in exhaled breath and blood glucose levels, with a mean linear regression coefficient of 0.82 [range: 0.08-0.98] (Table 1). However, none of the included studies validated their results internally (e.g. cross-validation) or externally (e.g. in an separate validation cohort). A total of ten metabolites have been reported to correlate with blood glucose levels, including exhaled acetone, VOCs produced by gut flora (ethanol, methanol, and propane), exogenous compounds (Ethyl benzene, o-xylene, and m/p-xylene) and VOCs that reflect oxidative status (methyl nitrate, 2-pentyl nitrate, and carbon monoxide (CO)) (Table 3). Authors of two studies did not observe a strong correlation. The first one of those did not find a significant correlation between a single measurement of breath acetone and blood glucose in T2DM subjects. Authors of the second study were unable to demonstrate a strong correlation between glucose levels and exhaled CO in healthy subjects. Researchers in one of the studies that did show a strong correlation between breath metabolites and glucose levels, only observed this after overnight fast, showing a weak correlation after consuming a meal [17].

Table 3 VOCs found to correlate with glucose levels

Temporal association

Researchers in seven out of nine studies performed multiple measurements with an interval ranging from 2.5 to 40 minutes. Two studies had a cross-sectional design and only performed a single measurement, or two unpaired measurements. None of the authors of the included studies reported on the possibility of predicting glucose trend.

Discussion

This systematic review identified nine studies that investigated the ability of exhaled breath to measure or estimate blood glucose levels. A significant correlation between VOCs in exhaled breath and blood glucose levels was found in seven studies. These results indicate that there is an association between the two, although not all studies are consistent. Researchers in one of these seven studies only found a strong correlation after overnight fasting of the subjects and were unable to replicate the results after a meal [17]. Authors of two negative studies did not find a strong correlation, possibly due to a different study design [18] and the VOC (e.g. CO) that was studied [19]. Authors of the study that included subjects with T2DM did not show a significant correlation between exhaled VOCs and blood glucose levels. This study also had a different (cross-sectional) design. The analytical technique used for VOC detection did not modify the reported correlation. None of the studies monitored breath continuously. Also the glucose trend, thus the temporal association between glucose and exhaled VOCs, was not explicitly investigated. However, the data from three longitudinal studies [2022] suggest that trends in glucose levels could possibly be monitored using exhaled breath when measurements are taken more frequently.

Index tests: exhaled breath analysis

A significant correlation between metabolites in exhaled breath and blood glucose levels was found using GC-MS, SIFT-MS, PTR-TOF-MS, a nano-sensing film-based sensor, and an electrochemical analyzer as analytical method. GC-MS is considered to be the gold standard for VOC detection and has shown to have a high sensitivity to identify single VOCs [23]. Therefore, GC-MS is suitable to accurately quantify a number of different VOCs in a cross-sectional study. However, the time-consuming nature of the technique limits use of the device for real-time and continuous measurements, which hampers clinical application. Other analytical techniques such as SIFT-MS [22, 23] and PTR-TOF-MS [2325] can also identify single VOCs and can be used for real-time continuous measurements. Disadvantages include possible selection bias [26] and the limitation to the concentration range that can be detected [25].

The electrochemical analyzers used in selected studies are two different Smokerlyzer Micro (Bedfont, UK) devices. These devices measure the amount of CO in exhaled breath. However, there is a cross sensitivity to hydrogen [19]. While a correlation between exhaled CO and glucose levels was found by the researchers of one study [27], researchers of another study [19] could not reproduce these findings. Contrasting results may be due to the high cross sensitivity to hydrogen in the electrochemical analyzer used previously [27], which was less apparent using a newer device [19]. This exemplifies the importance of an adequate analytical technique that suits the aim of the study.

An important limitation of the techniques used in included studies is that none of them was used to continuously monitor exhaled breath. Continuous analysis of the exhaled breath was previously described by means of IMR-MS [28], PTR-MS [24], and electronic nose [29, 30]. After a training phase, electronic noses learn to recognize specific disease states and can therefore be used for classification. The devices cannot identify and quantify single VOCs, but they do give a rapid, bedside diagnosis, which, from a clinical perspective, renders this device attractive. The electronic nose has been used to discriminate between patients with and without diabetes [31]. One could postulate that the ability to diagnose diabetes is partly due to the metabolomic alterations because of higher blood glucose levels in diabetic patients. Therefore, electronic nose analysis may complement mass-spectrometry based techniques for the monitoring of blood glucose levels in clinical practice, providing signals based on probabilistic training and validation. Alternatively to semi-selective recognition, nanosensors also rely on specific recognition of certain VOCs [30]. In one study, an acetone-selective nanomaterial-based sensor was used alongside PTR-TOF-MS and showed a strong correlation between acetone, glucose and the sensor [17]. Small size of devices using nanomaterial-based sensors as compared to spectrometry-based methods facilitates clinical application.

VOCs associated with blood glucose levels

Mechanisms related to the association between VOCs and glucose levels can be found in Table 3. Acetone appeared to be associated with blood glucose levels [17, 2022, 32]. As a result of increased synthesis of acetone and degradation of ketone bodies, acetone is expected to be higher in diabetics [33]. On the other hand, healthy humans only have elevated levels of ketone bodies when fasting or exercising [11]. Therefore, it is more likely to find a correlation between exhaled acetone and glucose levels after fasting compared to finding a correlation after lunch [17]. Acetone possibly is a good marker for glucose levels in ICU-patients. However, the large variation in breath acetone levels between subjects [18, 3436] may result in low accuracy when using acetone cross-sectionally.

VOCs such as ethanol [17, 20, 22, 32], propane [22] and methanol [17, 22] are likely to reflect gut flora activity, since the metabolism of gut bacteria is responsive to glycemic fluctuations [22, 32]. However, we cannot exclude that other biochemical pathways also contribute to the production of these compounds. In critically ill patients on the ICU, the quantity and composition of the gut microbiome are changing over time and therefore the amount of VOCs they produce may not be stable [37] Therefore, these markers are less likely to predict glucose levels in ICU-patients.

Ethyl benzene [20, 22], o-xylene [20] and m/p-xylene [20] are gasses that are inhaled, partially metabolized by the liver and subsequently exhaled at lower concentrations [20]. Rapid-onset hyperglycemia likely suppresses hepatic metabolism, thus causing peak concentrations of these compounds in exhaled air. Recent evidence suggests that cyclic hydrocarbons such as ethyl benzene and xylene are emitted by the ventilator and tubing [38]. Given that exhaled air is readily accessible for measurements in mechanically ventilated ICU patients, use of exhaled air for the prediction of glucose levels is therefore plausible.

An isomer of methyl nitrate [22, 39] is formed when methanol reacts with nitric oxide, which in turn reacts with superoxide ion (O2 -), a by-product of oxidative reactions [39]. Furthermore, 2-pentyl nitrate [22] is generated through pathways involving organic peroxy radical (RO2) and NO or NO2. This could be modulated by acute changes in systematic oxidative status [22]. Changes in CO [27] in exhaled breath are possibly related to oxidative stress. When glucose levels rise, particularly in diabetic patients, this can lead to oxidative stress. As a protective response, heme oxygenase is activated, leading to the positive modulation of CO on insulin secretion [27]. For critically ill patients on the ICU however, markers of oxidative stress will be non-specific for high blood glucose, as they increase with any form of oxidative stress such as sepsis, high inspired-oxygen fraction and acute respiratory distress syndrome [40].

Study design

The observed correlation between blood glucose levels and exhaled VOCs may be due to the inclusion of T2DM patients and/or a cross-sectional study design. First, T2DM influences the responsiveness of the body to changes in blood glucose levels [41]. This is typically characterized by insulin resistance but may also influence the formation of ketone bodies and the induction of liver enzymes. Second, breath acetone levels tend to differ between T1DM, T2DM, and healthy subjects [18, 21, 35]. Therefore, a decrease in blood glucose levels may not induce the same rise in breath acetone levels with different baseline values and in the context of different co-morbidities. Finally, in line with the previous point, correction for baseline differences between subjects cannot be accomplished with a cross-sectional study design. This is further acknowledged by the fact that the predictive algorithm requires calibration for every subject in several studies [20, 22]. Since the relation between exhaled breath metabolites and blood glucose levels shows high inter-person variation, a cross-sectional design may not be ideal for predicting glucose levels using breath metabolites. The possibility of using a single breath maneuver to estimate blood glucose levels thus seems implausible. Future studies may therefore focus on longitudinal measurements in the same subject.

Five included studies used a clamp study design and 2 studies used an oral glucose tolerance test (OGTT). Clamp studies and OGTT result in a more or less predictable course of blood glucose levels. Although a clamp design is ideal for research purposes and enables comparability between studies, clinical practice is often very different and less predictable. The transition of the results of these studies to the clinical setting will be a major challenge for the field of blood glucose estimation by exhaled breath analysis.

Strengths and limitations

We used a standardized systematic review approach, combining all evidence available. All VOCs that are linked to changes in glucose levels are discussed and their most likely biochemical pathways are described. In addition, we carefully assessed the quality of the included studies.

This systematic review also has an important limitation. The included studies were highly heterogeneous with respect to patient selection, exhaled breath sampling and analysis and blood glucose measurement, limiting the comparability of the studies. Therefore, we decided to describe the results separately. Most of the included studies had a relatively high risk of bias and we found that included studies did not validate their results. Possibly, this is inevitable in the early stages of biomedical research but it hinders strong conclusions. Furthermore, models can possibly be overfit, yielding overoptimistic results. Our search only identified one negative study. Negative studies are often not published leading to publication bias.

None of the studies investigated ICU-patients, while glucose fluctuations are large and frequent in this population [42]. Therefore, we cannot draw firm conclusions on the use of these methods in ICU-patients. We did try to identify potential pitfalls for the implementation of these methods in ICU patients by reviewing the biochemical pathways for the formation of VOCs.

Finally, the use of exhaled breath to monitor glucose trends was not discussed in any of the articles. Monitoring glucose trends (in ICU patients) however, has several potential advantages over using single values. First, trend has a better predictive value compared to single glucose levels; recent trend can be used to predict future levels. In ICU patients, this can lead to improved insulin titration. Second, because outliers can be filtered out, trend is less susceptible to random noise. Third, possible bias (constantly predicting values too high/low) will be constant throughout the trend, having a smaller effect. Potential disadvantages of using glucose trend are possible lag in the signal, and the potential of amplification of errors.

Conclusion

In conclusion, a significant association between VOCs in exhaled breath and blood glucose levels was found in the majority of studies included in this systematic review. Acetone, carbon monoxide, ethanol, ethyl benzene, M/P-xylene, methanol, O-xylene, and propane were correlated with blood glucose levels. Several potential effect modifiers were identified for ICU-patients. The included studies were performed under highly controlled circumstances, which limit generalizability. Our results warrant clinical validation of exhaled breath analysis for the monitoring of blood glucose levels in critically ill ICU-patients.

Abbreviations

CO:

Carbon monoxide

GC–MS:

Gas chromatography and mass spectrometry

ICU:

Intensive care unit

IMS:

Ion–mobility spectroscopy

OGTT:

Oral glucose tolerance test

PTR-(TOF)-MS:

Proton transfer reaction (time of flight) mass spectrometry

SIFT-MS:

Selected ion flow tube mass spectrometry

T1DM:

Type 1 diabetes mellitus

T2DM:

Type 2 diabetes mellitus

VOC:

Volatile organic compound.

References

  1. Schultz MJ, Spronk PE, van Braam Houckgeest F: Glucontrol, no control, or out of control?. Intensive Care Med. 2010, 36: 173-174. author reply 175–6

    Article  PubMed  Google Scholar 

  2. Schultz MJ, Harmsen RE, Spronk PE: Clinical review: strict or loose glycemic control in critically ill patients–implementing best available evidence from randomized controlled trials. Crit Care. 2010, 14: 223-

    Article  PubMed  PubMed Central  Google Scholar 

  3. Finfer S, Wernerman J, Preiser J-C, Cass T, Desaive T, Hovorka R, Joseph JI, Kosiborod M, Krinsley J, Mackenzie I, Mesotten D, Schultz MJ, Scott MG, Slingerland R, Van den Berghe G, Van Herpe T: Clinical review: consensus recommendations on measurement of blood glucose and reporting glycemic control in critically ill adults. Crit Care. 2013, 17: 229-

    Article  PubMed  PubMed Central  Google Scholar 

  4. Aragon D: Evaluation of nursing work effort and perceptions about blood glucose testing in tight glycemic control. Am J Crit Care. 2006, 15: 370-377.

    PubMed  Google Scholar 

  5. Karon BS, Boyd JC, Klee GG: Glucose meter performance criteria for tight glycemic control estimated by simulation modeling. Clin Chem. 2010, 56: 1091-1097.

    Article  CAS  PubMed  Google Scholar 

  6. Shaham O, Wei R, Wang TJ, Ricciardi C, Lewis GD, Vasan RS, Carr SA, Thadhani R, Gerszten RE, Mootha VK: Metabolic profiling of the human response to a glucose challenge reveals distinct axes of insulin sensitivity. Mol Syst Biol. 2008, 4: 214-

    Article  PubMed  PubMed Central  Google Scholar 

  7. Duarte NC, Becker SA, Jamshidi N, Thiele I, Mo ML, Vo TD, Srivas R, Palsson BØ: Global reconstruction of the human metabolic network based on genomic and bibliomic data. Proc Natl Acad Sci U S A. 2007, 104: 1777-1782.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  8. Wopereis S, Rubingh CM, van Erk MJ, Verheij ER, van Vliet T, Cnubben NHP, Smilde AK, van der Greef J, van Ommen B, Hendriks HFJ: Metabolic profiling of the response to an oral glucose tolerance test detects subtle metabolic changes. PLoS One. 2009, 4: e4525-

    Article  PubMed  PubMed Central  Google Scholar 

  9. Bijland LR, Bomers MK, Smulders YM: Smelling the diagnosis: a review on the use of scent in diagnosing disease. Neth J Med. 2013, 71: 300-307.

    CAS  PubMed  Google Scholar 

  10. Roberts K, Jaffe A, Verge C: Noninvasive monitoring of glucose levels: is exhaled breath the answer?. J Diabetes Sci Technol. 2012, 6: 659-664.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Turner C: Potential of breath and skin analysis for monitoring blood glucose concentration in diabetes. Expert Rev Mol Diagn. 2011, 11: 497-503.

    Article  CAS  PubMed  Google Scholar 

  12. Minh TDC, Blake DR, Galassetti PR: The clinical potential of exhaled breath analysis for diabetes mellitus. Diabetes Res Clin Pract. 2012, 97: 195-205.

    Article  PubMed Central  Google Scholar 

  13. Khan KS, Kunz R, Kleijnen J, Antes G: Five steps to conducting a systematic review. J R Soc Med. 2003, 96: 118-121.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Liberati A, Altman DG, Tetzlaff J, Mulrow C, Gøtzsche PC, Ioannidis JPA, Clarke M, Devereaux PJ, Kleijnen J, Moher D: The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate healthcare interventions: explanation and elaboration. BMJ. 2009, 339: b2700-

    Article  PubMed  PubMed Central  Google Scholar 

  15. Whiting PF, Rutjes AWS, Westwood ME, Mallett S, Deeks JJ, Reitsma JB, Leeflang MMG, Sterne JAC, Bossuyt PMM, Group Q-2: QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med. 2011, 155: 529-536.

    Article  PubMed  Google Scholar 

  16. Kanji S, Buffie J, Hutton B, Bunting PS, Singh A, McDonald K, Fergusson D, McIntyre LA, Hébert PC: Reliability of point-of-care testing for glucose measurement in critically ill adults. Crit Care Med. 2005, 33: 2778-2785.

    Article  PubMed  Google Scholar 

  17. Righettoni M, Schmid A, Amann A, Pratsinis SE: Correlations between blood glucose and breath components from portable gas sensors and PTR-TOF-MS. J Breath Res. 2013, 7: 037110-

    Article  CAS  PubMed  Google Scholar 

  18. Storer M, Dummer J, Lunt H, Scotter J, McCartin F, Cook J, Swanney M, Kendall D, Logan F, Epton M: Measurement of breath acetone concentrations by selected ion flow tube mass spectrometry in type 2 diabetes. J Breath Res. 2011, 5: 46011-

    Article  Google Scholar 

  19. Fritsch T, van Herpen M, von Basum G, Hering P, Murtz M: Is exhaled carbon monoxide level associated with blood glucose level? A comparison of two breath analyzing methods. J Biomed Opt. 2008, 13: 34012-

    Article  Google Scholar 

  20. Lee J, Ngo J, Blake D, Meinardi S, Pontello AM, Newcomb R, Galassetti PR: Improved predictive models for plasma glucose estimation from multi-linear regression analysis of exhaled volatile organic compounds. J Appl Physiol. 2009, 107: 155-160.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  21. Turner C, Walton C, Hoashi S, Evans M: Breath acetone concentration decreases with blood glucose concentration in type I diabetes mellitus patients during hypoglycaemic clamps. J Breath Res. 2009, 3: 46004-

    Article  Google Scholar 

  22. Minh TDC, Oliver SR, Ngo J, Flores R, Midyett J, Meinardi S, Carlson MK, Rowland FS, Blake DR, Galassetti PR: Noninvasive measurement of plasma glucose from exhaled breath in healthy and type 1 diabetic subjects. Am J Physiol Endocrinol Metab. 2011, 300: E1166-E1175.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  23. Boots AW, van Berkel JJ, Dallinga JW, Smolinska A, Wouters EF, Van Schooten FJ: The versatile use of exhaled volatile organic compounds in human health and disease. J Breath Res. 2012, 6: 27108-

    Article  Google Scholar 

  24. White IR, Willis KA, Whyte C, Cordell R, Blake RS, Wardlaw AJ, Rao S, Grigg J, Ellis AM, Monks PS: Real-time multi-marker measurement of organic compounds in human breath: towards fingerprinting breath. J Breath Res. 2013, 7: 17112-

    Article  Google Scholar 

  25. Jordan A, Haidacher S, Hanel G, Hartungen E, Märk L, Seehauser H, Schottkowsky R, Sulzer P, Märk TD: A high resolution and high sensitivity proton-transfer-reaction time-of-flight mass spectrometer (PTR-TOF-MS). Int J Mass Spectrom. 2009, 286: 122-128.

    Article  CAS  Google Scholar 

  26. Spanĕl P, Smith D: Selected ion flow tube mass spectrometry for on-line trace gas analysis in biology and medicine. Eur J Mass Spectrom (Chichester, Eng). 2007, 13: 77-82.

    Article  Google Scholar 

  27. Paredi P, Biernacki W, Invernizzi G, Kharitonov SA, Barnes PJ: Exhaled carbon monoxide levels elevated in diabetes and correlated with glucose concentration in BloodA new test for monitoring the disease?. CHEST J. 1999, 116: 1007-1011.

    Article  CAS  Google Scholar 

  28. Dolch ME, Frey L, Hornuss C, Schmoelz M, Praun S, Villinger J, Schelling G: Molecular breath-gas analysis by online mass spectrometry in mechanically ventilated patients: a new software-based method of CO(2)-controlled alveolar gas monitoring. J Breath Res. 2008, 2: 37010-

    Article  CAS  Google Scholar 

  29. Bos LDJ, van Walree IC, Kolk AHJ, Janssen H-GH-G, Sterk PJ, Schultz MJ: Alterations of exhaled breath metabolite-mixtures in two rat models of lipopolysaccharide-induced lung injury. J Appl Physiol. 2013, 115: jap.00685.2013-

    Article  Google Scholar 

  30. Konvalina G, Haick H: Sensors for breath testing: from nanomaterials to comprehensive disease detection. Acc Chem Res. 2013, 47: 66-76.

    Article  PubMed  Google Scholar 

  31. Dalton P, Gelperin A, Preti G: Volatile metabolic monitoring of glycemic status in diabetes using electronic olfaction. Diabetes Technol Ther. 2004, 6: 534-544.

    Article  CAS  PubMed  Google Scholar 

  32. Galassetti PR, Novak B, Nemet D, Rose-Gottron C, Cooper DM, Meinardi S, Newcomb R, Zaldivar F, Blake DR: Breath ethanol and acetone as indicators of serum glucose levels: an initial report. Diabetes Technol Ther. 2005, 7: 115-123.

    Article  CAS  PubMed  Google Scholar 

  33. Pavlou AK, Turner AP: Sniffing out the truth: clinical diagnosis using the electronic nose. Clin Chem Lab Med. 2000, 38: 99-112.

    Article  CAS  PubMed  Google Scholar 

  34. Wang C, Mbi A, Shepherd M: A Study on breath acetone in diabetic patients using a cavity ringdown breath analyzer: exploring correlations of breath acetone with blood glucose and glycohemoglobin A1C. Sensors Journal, IEEE. 2010, 10: 54-63.

    Article  Google Scholar 

  35. Deng C, Zhang J, Yu X, Zhang W, Zhang X: Determination of acetone in human breath by gas chromatography–mass spectrometry and solid-phase microextraction with on-fiber derivatization. J Chromatogr B Analyt Technol Biomed Life Sci. 2004, 810: 269-275.

    Article  CAS  PubMed  Google Scholar 

  36. Spanĕl P: Acetone, ammonia and hydrogen cyanide in exhaled breath of several volunteers aged 4–83 years. J Breath Res. 2007, 1: 11001-

    Article  Google Scholar 

  37. Hillman KM, Riordan T, O’farrell SM, Tabaqchali S: Colonization of the gastric contents in critically patients. Crit Care Med. 1982, 10: 444-

    Article  CAS  PubMed  Google Scholar 

  38. Bos LDJ, Wang Y, Weda H, Nijsen TME, Janssen APGE, Knobel HH, Vink TJ, Schultz MJ, Sterk PJ: A simple breath sampling method in intubated and mechanically ventilated critically ill patients. Respir Physiol Neurobiol. 2013, 191C: 67-74.

    Google Scholar 

  39. Novak BJ, Blake DR, Meinardi S, Rowland FS, Pontello A, Cooper DM, Galassetti PR: Exhaled methyl nitrate as a noninvasive marker of hyperglycemia in type 1 diabetes. Proc Natl Acad Sci U S A. 2007, 104: 15613-15618.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  40. Crader MK, Repine DJJ: Breath Biomarkers and the Acute Respiratory Distress Syndrome. J Pulm Respir Med. 2012, 2: 1-9.

    Article  Google Scholar 

  41. Grapov D, Adams SH, Pedersen TL, Garvey WT, Newman JW: Type 2 diabetes associated changes in the plasma non-esterified fatty acids: oxylipins and endocannabinoids. PLoS One. 2012, 7: e48852-

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  42. McCowen KC, Malhotra A, Bistrian BR: Stress-induced hyperglycemia. Crit Care Clin. 2001, 17: 107-124.

    Article  CAS  PubMed  Google Scholar 

Pre-publication history

Download references

Acknowledgements

LDJB is supported by a research grant (PhD Scholarship) of the Academic Medical Center (https://www.amc.nl/web/Onderwijs/PhD/AMC-Scholarships/AMC-PhD-Scholarships-2014.htm -winners.htm), by an unrestricted research grant from Philips Research and by the ESICM Young Investigator Award (http://www.esicm.org/research/awards). None of these funding bodies had any role in design, in the collection, analysis, and interpretation of data; in the writing of the manuscript; and in the decision to submit the manuscript for publication.

Financial support

All funding came from departmental sources.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jan Hendrik Leopold.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

JHL designed the review, extracted the data, summarized the findings, and composed the manuscript. RTMvH participated in the study design and contributed to the manuscript. PJS participated in the study design and contributed to the manuscript. AAH participated in the study design and contributed to the manuscript. MJS participated in the study design, contributed to the manuscript and coordinated the team efforts. LDB participated in the study design, extracted the data, summarized the findings and contributed to the manuscript. All authors read and approved the final manuscript.

Electronic supplementary material

12871_2014_236_MOESM1_ESM.pdf

Additional file 1: QUADAS-2 - Adapted for systematic review on Glucose Prediction by Analysis of Exhaled Metabolites.(PDF 285 KB)

Authors’ original submitted files for images

Below are the links to the authors’ original submitted files for images.

Authors’ original file for figure 1

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Leopold, J.H., van Hooijdonk, R.T., Sterk, P.J. et al. Glucose prediction by analysis of exhaled metabolites – a systematic review. BMC Anesthesiol 14, 46 (2014). https://doi.org/10.1186/1471-2253-14-46

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1471-2253-14-46

Keywords