Intraaortic Balloon Pump Counterpulsation and Cerebral Autoregulation: an observational study

Abstract

Background

The use of Intra-aortic counterpulsation is a well established supportive therapy for patients in cardiac failure or after cardiac surgery. Blood pressure variations induced by counterpulsation are transmitted to the cerebral arteries, challenging cerebral autoregulatory mechanisms in order to maintain a stable cerebral blood flow. This study aims to assess the effects on cerebral autoregulation and variability of cerebral blood flow due to intra-aortic balloon pump and inflation ratio weaning.

Methods

Cerebral blood flow was measured using transcranial Doppler, in a convenience sample of twenty patients requiring balloon counterpulsation for refractory cardiogenic shock (N = 7) or a single inotrope to maintain mean arterial pressure following an elective placement of an intra-aortic balloon pump for cardiac surgery (N = 13). Simultaneous blood pressure at the aortic root was recorded via the intra-aortic balloon pump. Cerebral blood flow velocities were recorded for six minute intervals at a 1:1 balloon inflation-ratio (augmentation of all cardiac beats) and during progressive reductions of the inflation-ratio to 1:3 (augmentation of one every third cardiac beat). Real time comparisons of peak cerebral blood flow velocities with systolic blood pressure were performed using cross-correlation analysis. The primary endpoint was assessment of cerebral autoregulation using the time delay between the peak signals for cerebral blood flow velocity and systolic blood pressure, according to established criteria. The variability of cerebral blood flow was also assessed using non-linear statistics.

Results

During the 1:1 inflation-ratio, the mean time delay between aortic blood pressure and cerebral blood flow was -0.016 seconds (95% CI: -0.023,-0.011); during 1:3 inflation-ratio mean time delay was significantly longer at -0.010 seconds (95% CI: -0.016, -0.004, P < 0.0001). Finally, upon return to a 1:1 inflation-ratio, time delays recovered to those measured at baseline. During inflation-ratio reduction, cerebral blood flow irregularities reduced over time, whilst cerebral blood flow variability at end-diastole decreased in patients with cardiogenic shock.

Conclusions

Weaning counterpulsation from 1:1 to 1:3 inflation ratio leads to a progressive reduction in time delays between systolic blood pressure and peak cerebral blood flow velocities suggesting that although preserved, there is a significant delay in the establishment of cerebral autoregulatory mechanisms. In addition, cerebral blood flow irregularities (i.e. surrogate of flow adaptability) decrease and a loss of cerebral blood flow chaotic pattern occurs during the end-diastolic phase of each beat in patients with cardiogenic shock.

Peer Review reports

Background

Aortic counterpulsation reduces myocardial ischemia by decreasing left ventricular afterload and increasing coronary blood flow. Pulse pressure variation is modified by the increase of diastolic-pressure time index and tension time index ratio (DPTI/TTI ratio). These principles are the physiological basis upon which the intra-aortic balloon pump (IABP) is used in states of cardiogenic shock and placed peri-cardiac surgery [1, 2]. Although some studies have shown that IABP also increases cerebral blood flow (CBF) in patients with subarachnoid hemorrhage and cerebral vasospasm [3], other studies have reported reductions in CBF, specifically inducing a phenomenon of transient reversed diastolic flow with secondary reversed diastolic cerebral blood flow velocity (CBFV) [4]. Cheung and colleagues demonstrated that IABP modifies the CBFV waveform, mirroring that of BP without affecting the mean CBFV [5]. Acute reductions in CBFV at end-diastole can be compensated by an increase in mean CBFV [5]. However, Cheung and colleagues assessed CBF in clinically stable patients requiring prophylactic insertion of IABP, attributing the stability of the mean CBFV to a presumed intact cerebral autoregulation, but their study did not assess cerebral autoregulation.

Cerebral autoregulation is the mechanism by which constant CBF is maintained despite changes in cerebral perfusion pressure (CPP). Cerebral autoregulation mediates states of hyperemia or ischemia to avoid vasogenic edema or infarction respectively [6]. Impaired autoregulation has been regarded as a risk factor associated with adverse neurological outcome after cardiac surgery [7, 8], head injury [9–11] and subarachnoid hemorrhage [12]. Cerebral autoregulation responds to spontaneous and induced changes in arterial blood pressure (BP) such as those occurring with IABP [13–15]. Neurological complications such as dense hemiplegia and long-term cognitive impairment are the most devastating adverse events after cardiac surgery [16], being these, mediated by an impaired cerebral autoregulation particularly in the elderly [7, 8] and during reperfusion processes [17]. During periods of global hypoperfusion, certain regions within the brain are more susceptible to developing an ischemic penumbra [18–20]. It is in these areas where cerebral autoregulation may be most impaired, aggravating the primary tissue ischemia [21]. Cerebral autoregulation has been extensively studied using transcranial Doppler (TCD) which measures CBFV as a surrogate of CBF [22–25]. From all these methods, cross-correlation function (CCF) incorporates powerful computerized systems that enable the analysis of the phase shifts between BP and CBFV waveforms with the advantage of being useful for continuous testing of cerebral autoregulation at the bedside [26, 27]. Using cross-correlation function, cerebral autoregulation is defined as preserved when there is a time delay up to 2 seconds and a negative correlation between BP and CBFV [27].

Whilst the use of IABP has represented a significant advantage in cardiovascular support and is often a necessary adjuvant for myocardial recovery, it is yet not clear what repercussions in CBF may imply, specifically during the weaning phase. This is the first study to assess cerebral autoregulation in a population at high risk of postoperative neurological damage during weaning of counterpulsation therapy.

Methods

As an exploration of concept, a sample of 20 patients was recruited from two intensive care units. Group one comprised 13 patients requiring an elective insertion of an IABP following angioplasty or cardiac surgery and only one vasopressor to maintain mean arterial pressure (MAP). Group two consisted of seven patients with cardiogenic shock requiring an IABP as well as two or more vasopressors for MAP support, a persistent lactic acidosis and evidence of multiple-organ failure. Vasopressors were chosen at the clinician's discretion. The heterogeneous group size was not a limitation as intergroup comparisons were not the aim of the study. All patients recruited in this study gave informed consent.

A total of ten patients had undergone coronary artery bypass surgery (CABG); ten had been treated with pharmacological measures and coronary angioplasty. The proportion of CAGB patients was balanced between the two groups. All patients were in the acute phase of their illness (Table 1).

Table 1 Clinical characteristics of the study sample

Maintenance of a stable mean arterial pressure during inflation ratio reduction was ensured to allow a stable cerebral perfusion pressure and minimize interference with cerebral autoregulation. Patients were used as their own controls. Two patients were removed from the study due to significant hypotension (pre-defined as a reduction in BP systolic of more than 20% of baseline) during reduction of counterpulsation ratio. Data for 1:3 counterpulsation ratio was only collected in 11 patients, as the permission to wean counterpulsation was based on clinician judgment.

CBFV of middle cerebral arteries (MCAs) were measured using TCD following referenced criteria at the temporal acoustic window [23, 24]. Both MCAs were insonated and the side with best acoustic characteristics chosen for study. TCD machines (DWL® Sipplingen, Germany) used 2-MHz probes. Intra-patient variability was minimized using recordings only collected by one investigator formally trained in TCD [25].

Blood pressure signals from the IABP output channel and CBFV signals from the TCD device were simultaneously recorded using Powerlab (AD technology, model ML880, New South Wales, Australia) to minimise errors and bias which would be present if peripheral arterial pressure had been transduced. Stability of the insonated vessel diameter was assumed by maintaining a stable partial pressure of arterial carbon dioxide (PaCO2) during measurements. This assumption has been used previously in situations where transcranial color coded technology (which enables direct visualization and measurement of vessels diameter), is not available [6]. Patient care, including cardiovascular parameter goals and sedation were at the discretion of the treating clinician. Midazolam with minimal aliquots of fentanyl were used for analgesia and sedation in all cases. Propofolâ„¢ was not used due to the poor cardiac function and precarious haemodynamics in all these patients.

An initial three minute steady-state CBFV was recorded. With the IABP at a ratio of 1:1 a total of six minutes was recorded. Subsequently, the IABP timing was reduced to 1:2 and a further six minute recording of CBFV was taken to ensure the establishment of autoregulatory changes. This data collection was then repeated using a counterpulsation ratio of 1:3. After completion of the 1:3 ratio recordings (Figure 1), counterpulsation was returned to a 1:1 ratio and a further six minute recording collected.

Figure 1
figure 1

Aortic pressures and Transcranial Doppler tracings. Upper figure: IABP pressures (y axis) during 1:3 inflation-ratio over time (x axis). IABP augments every third cardiac beat, showing characteristic traces such as an augmented peak diastole, reduced end-diastolic pressure and decreased non-augmented peak systole; these pulse pressure variations imply a left ventricular afterload reduction whilst increasing diastolic pressures (to improve coronary perfusion). Lower figure: Cerebral blood flow velocities using TCD (y axis) at the Mean Cerebral artery, over time (x axis). Notice that the end-diastolic velocities become negative; known as "reversal flow" and are considered to be the result of a steeling phenomena from the deflation of the IABP in the Aortic root. Both traces are represented in a real time, beat-to-beat correlation, using the integrated software of a Power lab computer.

Institutional Ethics Committee approval for the performance of the study was granted. All patients or their next of kin gave informed consent prior to enrolment in the study.

Data analysis

Cerebral autoregulation was assessed using both linear and non-linear analysis. The former was based on pre-filtered data and application of the cross-correlation statistic. Cross-correlation measures the time between the peak of the CBFV at the level of the MCA and the peak of BP for each heart beat in real time. Several authors have found time delays between BP and CBFV up to a 2 seconds in healthy subjects and delays near to zero when autoregulation is impaired. With loss of cerebral autoregulation, changes in BP are directly transmitted to CBF without any delay in time. Conversely, when autoregulation is intact, buffering mechanisms within the cerebral vessels take place to accommodate BP changes without disturbing CBF dynamics, a "time consuming" process. By giving a trend rather than an absolute value a more realistic description of cerebral autoregulation is achieved rather than a dichotomous description of preserved/non-preserved autoregulation. This test has been extensively used and validated for the assessment of cerebral autoregulation both in dynamic and static conditions [26, 27].

For the non-linear analysis we used a third-order moment test as previously described [28]. The cross-correlation is a second order test. By examining two points in time it is able to measure only the linear delay in time between the two signals. A test in the third-order, examines three points in time, and so is able to detect more complex (non-linear) patterns. It is also able to detect non-stationary or chaotic signals (signals whose mathematical properties change randomly with time).

For the cross-correlation analysis, BP and MCA velocity signals were bandpass filtered using a fifth-order Butterworth filter, into three frequency ranges: very low (0.015-0.07 Hz), low-frequency (0.07-0.15 Hz) and high-frequency (0.15-0.40 Hz) before applying cross-correlation [26]. The time series were normalized so that the autocorrelations at lag zero were one. The peak in the cross-correlation was used to estimate the average delay in seconds (using the xcorr function in Matlab signal processing toolbox). The time delay was estimated for every subject and IABP ratio combination.

Time delays were virtually identical for all three frequency ranges. As such, subsequent analysis was performed on the very low frequency range only. No other data pre-processing was applied [26]. Other researchers have applied windowing (sampling by seconds) to further filter data; however, in this study signals were at risk of being dampened (or even lost) using windowing. Therefore, to minimize the risk of missing important signals, data was kept in a raw form at the expense of some increased noise.

Patterns in time delays among subjects and by IABP ratio were examined using multiple regression modeling with IABP ratio and group as independent variables. A mixed model with a random intercept for each subject was used to control for repeated within-subject measurements [29]. This model was fitted using Proc Mixed in SAS. To give a graphical picture of the overall delay, the average cross-correlation for each balloon and cohort combination was plotted (Figure 2).

Figure 2
figure 2

Time delays and inflation ratios. Time delays in couples of group 1 - left and group 2 - right, showing the relative delay for 1:1(continuous line), 1:2 (dotted line), 1:3 (dashed line) and 1:1-end (intermittent dashed-dotted line) IABP inflation ratio.

Non-linear analysis

Non-linear differences were assessed using a test based in the third-order moment [28]. Initially, series were combined by dividing the IABP by MCA velocity and then log-transformation to give a series with more stable variance. Non-linearities were also assessed in the time domain from zero to 0.1 seconds. Separate tests for each subject and IABP ratio combination were performed and the results combined by overlaying each subject's significant region of non-linearity for the same IABP ratio to identify the times of common non-linear interactions. Regions of non-linear interactions where then plotted for each IABP ratio. Third-order non-linearity was calculated using the "bootstrap" function (Mathworks Australia Pty Ltd).

Beat-to-beat analysis

To visually understand the changes in the IABP and MCA velocity with decreasing balloon ratio, mean time series per beat were plotted (the mean of the log-transformed ratio of IABP to MCA velocity). Each subject's beat length was initially estimated and the mean IABP and MCA velocity subtracted within each beat "window". The first 100 beats were then averaged from the start to the end of the beat in 0.001 second gaps. Mean beats for two subjects (each subject belonging to each study group) were plotted. This was a graphical analysis with no formal statistical testing.

Results

Phase delays between peak CBFV and systolic IABP at the very low frequency range, decreased with a reduction in counterpulsation ratio with mean delays for 1:1, 1:2 and 1:3 counterpulsation ratios of -0.016 seconds (95% CI: -0.023, -0.011), -0.013 seconds (95% CI: -0.019, -0.007) and -0.010 seconds (95% CI: -0.016, -0.004) respectively; 1:1 counterpulsation ratio compared to 1:2, P < 0.0001; 1:1 counterpulsation ratio compared to 1:3, P < 0.0001. Following return to a 1:1 counterpulsation ratio, the phase delay increased between peak CBFV and systolic BP (-0.017 seconds, 95% CI: -0.023 -0.011) being unchanged from initial 1:1 measures (P = 0.62, Table 2). High correlation coefficients were found between counterpulsation ratio and phase delays (Table 3). There were no phase delay differences between patients with cardiogenic shock and those receiving prophylactic counterpulsation.

Table 2 Time delays for IABP ratios Time delay (seconds)
Table 3 Cross-correlation coefficients for time delays and groups

Non-linearities in the ratio of IABP to MCA velocity within lags of 0.1 seconds were noted. As the counterpulsation ratio decreased from 1:1 to 1:3, the common region of non-linearity showed longer time delays demonstrated by areas of the common regions shifting from left to right along the x-axes (Figures 3, 4).

Figure 3
figure 3

Cerebral blood flow variability represented in a scale of colors. Non-linearities represented in dark orange colors, shifting to the right on the (x) axis, for group 1(patients electively supported with IABP): Top-left: 1:1 ratio; top-right: 1:2 ratio; bottom left: 1:3 ratio; bottom right: 1:1-end ratio. Notice the progressive increase of time-lag of the cerebral blood flow variability as the IABP inflation-ratio is weaned; suggesting that cerebral blood flow variability takes longer to be established as the IABP is weaned off.

Figure 4
figure 4

Cerebral blood flow variability represented in a scale of colours. Non-linearities represented in dark orange colours, shifting to the right on the (x) axis, for group 2 (patients in refractory shock): Top-left: 1:1 ratio; top-right: 1:2 ratio; bottom left: 1:3 ratio; bottom right: 1:1-end ratio. Notice the progressive increase of time-lag of the cerebral blood flow variability as the IABP inflation-ratio is weaned; suggesting that cerebral blood flow variability takes longer to be established as the IABP is weaned off. Overall reduction in cerebral blood flow variabilities are shown, when compared with subjects in group 1-fig 3.

The beat-to-beat plots show the mean of the first 100 beats for a patient in groups one (Figure 5) and two (Figure 6). The log-ratio of the two series was always greatest at the area of the IABP end-diastole indicating that the greatest relative difference in the two series occurred at these areas. It is in these areas where the group one subject displays chaotic behavior, as the pattern becomes more erratic when compared with the smoother pattern seen in group two.

Figure 5
figure 5

Representation of cerebral blood flow chaotic pattern. Real tracing from a representative subject in group 1 (patients with elective support of IABP) showing: Upper row: left figure corresponds to 1:1 IABP inflation-ratio trace (in red) superimposed to CBFV (in blue) during the same beat. Middle figure corresponds to 1:2 IABP inflation-ratio trace (in red) superimposed to CBFV (in blue) during another beat. Right figure corresponds to IABP inflation-ratio trace (in red) superimposed to CBFV (in blue) during a cardiac beat. Lower row: simultaneous representation of CBFV non-linearities (variability) expressed as a time function of log ratio (y - axis) and time (x - axis). Notice that the moment of maximum variability corresponds to the end-diastole in each beat

Figure 6
figure 6

Representation of cerebral blood flow chaotic pattern. Real tracing from a representative subject in group 2 (patients in refractory shock) showing: Upper row: left figure corresponds to 1:1 IABP inflation-ratio trace (in red) superimposed to CBFV (in blue) during the same beat. Middle figure corresponds to 1:2 IABP inflation-ratio trace (in red) superimposed to CBFV (in blue) during another beat. Right figure corresponds to IABP inflation-ratio trace (in red) superimposed to CBFV (in blue) during a cardiac beat. Lower row: simultaneous representation of CBFV non-linearities (variability) expressed as a time function of log ratio (y - axis) and time (x - axis). Notice that the moment of maximum variability corresponds to the end-diastole in each beat, however in this case variability is minimum when compared with group 1.

Discussion

This study demonstrates a significant decrease in the phase delay between systolic IABP and peak CBFV as counterpulsation inflation ratio is reduced. In addition, time delays are negative indicating that the changes in CBFV precede those of the systemic arterial tree recorded by the IABP, a behavior previously reported by other authors indicating that cerebral autoregulation is preserved [27]. The progressive shortening of time delays suggest that although preserved, cerebral autoregulation shows a loss of efficacy through the phases of inflation ratio reduction. This pattern differs from the dichotomous description of cerebral autoregulation, i.e. an "all or nothing" state of cerebral autoregulation and represents an assessment in agreement with other authors [26]. Finally, high correlation coefficients found between inflation ratios and time delays of systolic pressures and peak cerebral flow velocities, indicate that there is a changeable autoregulatory response among ratio reduction phases.

Phase delays returned to baseline levels when the counterpulsation ratio returned to a 1:1 ratio, suggesting that there is an immediate recovery of the efficiency of autoregulation. Although the exact reason for this behavior is unclear, two explanations are plausible: Firstly, although not apparent in our study, non-significant reductions in MAP could have accompanied the change of IABP from 1:2 to 1:3, leading to small reductions in CBF which could have then diminished the efficiency of the autoregulatory response. Secondly, the reduction in the frequency of the counterpulsation balloon and therefore the lesser homogenous and more disturbed blood flow pattern exhausts the efficacy (or latency in the response) of cerebral autoregulation. Cheung and colleagues found that CBFV-to-pressure ratio at end-diastole was dependent on the magnitude of augmentation rather than the timing of inflation for both augmented and non-augmented beats [5]. However, early IABP deflation time leads to end-diastolic reversal of CBF [30]. This reversal flow pattern is magnified in patients with left ventricle dysfunction. In this setting of delayed isovolumetric contraction, IABP deflation occurring during a period when there is no blood ejection into the aorta, retrograde carotid artery blood flow occurs.

Although the use of different types of vasopressors could hypothetically interfere in the autoregulatory response, the pattern of progressive reduction in time delays with the reduction of inflation ratio, remain unchanged when compared with patients with minimal vasopressor requirements.

Currently, the two known modes of weaning IABP support are a reduction in ratio of augmented beats and a reduction of balloon inflation pressure while maintaining a support ratio of 1:1. There is no consensus as to which mode of weaning IABP support is superior. A strategy of weaning that avoids carotid flow reversal or impaired cerebral autoregulation could minimize neurological complications.

Cerebral autoregulatory efficiency deteriorates particularly in those patients with severe left ventricle dysfunction; with a reduction of counterpulsation ratio potentially risking cerebral hypoperfusion [7]. This study shows three patients with reversal end-diastolic pattern and a negative CBFV. This phenomenon requires further investigation to define if CBF reductions are related to IABP timing or to balloon augmentation pressure.

Time delays between peak CBFV and systolic IABP show a negative phase consistent with preservation of cerebral autoregulation. This "phase-led property" of cerebral autoregulation has been previously described with CBFV changes preceding BP changes [14].

Non-linear analysis shows that as IABP is weaned to 1:3 ratio, CBFV irregularities spread apart in time, shifting from left to right on the x axis (Figures 3, 4). An increase in chaotic activity at end-diastole was more common in the Group one subject when compared with a Group two subject (Figures 5, 6). This chaotic activity may reflect a more intact system. Several authors have reported beat-to-beat variability in CBF and CBFV [31, 32]; perhaps the result of autonomic cerebral vasomotion [33] or oscillations in central control [34]. Variability in biological systems such as heart rate have been extensively studied and correlated with outcome [35, 36]. Cerebral blood flow has been described as "multifractal" (a behavior that is irregular, mathematically repetitive and explained by its own definition) and chaotic in the normal population while demonstrating monofractal behavior when cerebral autoregulation is impaired [37].

Limitations

This study focuses on the relationship between IABP pressures at the tip of the descending aorta and CBFV assessing their time delay using the cross-correlation approach [26, 27, 38]. This method has been previously used to assess the effect of spontaneous BP oscillations on CBFV in undisturbed systems. Although this study incorporated patients with pressure waveform modifications, these were continuously sustained over days, leading to a stable blood waveform pattern; for this reason, the authors considered that the cross-correlation method could be appropriate for the continuous monitoring of cerebral autoregulation. Similarly to previous studies, data stability was ensured by taking long recordings and allowing a three minute steady state-phase before definitive recording [26].

CBFV was initially recorded in both MCAs, but only the side with better acoustic properties was recorded and analyzed. Although spatial heterogeneity of cerebral autoregulation as well as interhemispheric differences are well described [39], an endpoint of this study was not to define interhemispheric differences, but rather to ensure the best transcranial Doppler recordings in order to minimize the signal-to-noise ratio and increase data reliability [40].

Many authors have found time delays within two seconds in healthy populations when compared with patient cohorts [14]. It could be argued that small differences in time delays, as shown in this study, may not be clinically relevant; however, the emphasis of this study lies upon the conceptual effect of IABP weaning using inflation ratio reduction in both, time delays and chaotic behavior of cerebral blood flow. In a population such as this, cerebral autoregulatory dysfunction has the potential for devastating consequences. Further studies comparing both counterpulsation weaning strategies (inflation-ratio reduction and pressure reduction) on cerebral autoregulation may be needed to define the implications of pulse pressure variation on cerebral autoregulation.

Although this study clinically defines two groups with presumed differences in left ventricular function, such separations may have been arbitrary as there were no autoregulation differences between patients in refractory cardiogenic shock and those receiving supportive counterpulsation.

Time domain cross-correlation functions are often applied to a series of signals in a different range of frequencies. Cerebral autoregulation is known to show high pass filter properties that enable low-frequency oscillations to be efficiently dampened by autoregulatory mechanisms while fluctuations in high-frequency ranges remain unfiltered [15]. Slow spontaneous oscillations occur at low frequencies in time series with continuous BP-CBFV recordings [41]. Our study processed data using bandpass filtering at three frequencies ranges prior to signal analysis, as previously described [42]. No differences in time delays were found among these frequency ranges and as such, results were presented using the very-low frequency filter only. Although the similarity among different frequency ranges has not been commonly reproduced, this finding could be explained by the lack of raw data smoothing, during the processing phase.

Conclusions

This study shows that the use of fully augmented IABP does not adversely affect cerebral autoregulation. Reductions in inflation ratio during counterpulsation weaning progressively worsens cerebral autoregulation efficiency with the potential risk of cerebral hypoperfusion. Reduction in chaotic activity seen at end-diastole in patients with cardiogenic shock and inflation ratio reduction may also indicate impairment of CBF dynamics. Further studies must investigate optimal IABP weaning strategies to minimize the risk of reversed CBF at end-diastole, particularly amongst patients with poor residual left ventricular function.

Abbreviations

IABP:

Intra aortic balloon pump

CBFV:

cerebral blood flow velocity

TCD:

transcranial Doppler

BP:

blood pressure

CBF:

cerebral blood flow

DPTI/TTI:

diastolic pressure time index and tension time index ratio

CPP:

cerebral perfusion pressure

CCF:

cross-correlation function

MAP:

mean arterial pressure

CABG:

coronary artery bypass graft

MCA:

mean cerebral artery

PaC02:

partial pressure of carbon dioxide.

References

  1. Baskett RJ, Ghali WA, Maitland A, Hirsch GM: The intraaortic balloon pump in cardiac surgery. Annals of Thoracic Surgery. 2002, 74: 1276-1287. 10.1016/S0003-4975(02)03663-9.

    Article  PubMed  Google Scholar 

  2. Baskett RJ, O'Connor GT, Hirsch GM, et al: The preoperative intraaortic balloon pump in coronary bypass surgery: a lack of evidence of effectiveness. American Heart Journal. 2005, 150: 1122-1127. 10.1016/j.ahj.2005.01.043.

    Article  PubMed  Google Scholar 

  3. Rosen CL, Sekhar LN, Duong DH: Use of intra-aortic balloon pump counterpulsation for refractory symptomatic vasospasm. Acta Neurochirurgica. 2000, 142: 25-32. 10.1007/s007010050003.

    Article  CAS  PubMed  Google Scholar 

  4. Brass LM: Reversed intracranial blood flow in patients with an intra-aortic balloon-pump. Stroke. 1990, 21: 484-487.

    Article  CAS  PubMed  Google Scholar 

  5. Cheung AT, Levy WJ, Weiss SJ, Barclay DK, Stecker MM: Relationships between cerebral blood flow velocities and arterial pressures during intra-aortic counterpulsation. Journal of Cardiothoracic Vascular Anesthesia. 1998, 12: 51-57. 10.1016/S1053-0770(98)90055-6.

    Article  CAS  PubMed  Google Scholar 

  6. Aaslid R, Lindegaard KF, Sorteberg W, Nornes H: Cerebral autoregulation dynamics in humans. Stroke. 1989, 20: 45-52.

    Article  CAS  PubMed  Google Scholar 

  7. Newman MF, Kirchner JL, Phillips-Bute B, et al: Longitudinal assessment of neurocognitive function after coronary-artery bypass surgery. New England Journal of Medicine. 2001, 344: 395-402. 10.1056/NEJM200102083440601.

    Article  CAS  PubMed  Google Scholar 

  8. Roach GW, Kanchuger M, Mangano CM, et al: Adverse cerebral outcomes after coronary bypass surgery. Multicenter Study of Perioperative Ischemia Research Group and the Ischemia Research and Education Foundation Investigators. New England Journal Medicine. 1996, 335: 1857-1863. 10.1056/NEJM199612193352501.

    Article  CAS  Google Scholar 

  9. Andrews PJ, Sleeman DH, Statham PF, et al: Predicting recovery in patients suffering from traumatic brain injury by using admission variables and physiological data: a comparison between decision tree analysis and logistic-regression. Journal of Neurosurgery. 2002, 97: 326-336. 10.3171/jns.2002.97.2.0326.

    Article  PubMed  Google Scholar 

  10. Enevoldsen EM, Jensen FT: Autoregulation and CO2 responses of cerebral blood flow in patients with acute severe head injury. Journal of Neurosurgery. 1978, 48: 689-703. 10.3171/jns.1978.48.5.0689.

    Article  CAS  PubMed  Google Scholar 

  11. Sahuquillo J, Poca MA, Ausina A, Baguena M, Gracia RM, Rubio E: Arterio-jugular differences of oxygen (AVDO2) for bedside assessment of CO2-reactivity and autoregulation in the acute phase of severe head injury. Acta Neurochirurgica. 1996, 138: 435-444. 10.1007/BF01420306.

    Article  CAS  PubMed  Google Scholar 

  12. Soehle M, Czosnyka M, Pickard JD, Kirkpatrick PJ: Continuous assessment of cerebral autoregulation in subarachnoid hemorrhage. Anesthesia and Analgesia. 2004, 98: 1133-1139. 10.1213/01.ANE.0000111101.41190.99.

    Article  PubMed  Google Scholar 

  13. Czosnyka M, Piechnik S, Richards HK, Kirkpatrick P, Smielewski P, Pickard JD: Contribution of mathematical modelling to the interpretation of bedside tests of cerebrovascular autoregulation. Journal of Neurology Neurosurgery and Psychiatry. 1997, 63: 721-731. 10.1136/jnnp.63.6.721.

    Article  CAS  Google Scholar 

  14. Diehl RR, Linden D, Lucke D, Berlit P: Phase relationship between cerebral blood flow velocity and blood pressure. A clinical test of autoregulation. Stroke. 1995, 26: 1801-1804.

    Article  CAS  PubMed  Google Scholar 

  15. Giller CA: The frequency-dependent behavior of cerebral autoregulation. Neurosurgery. 1990, 27: 362-368. 10.1097/00006123-199009000-00004.

    Article  CAS  PubMed  Google Scholar 

  16. Wolman RL, Nussmeier NA, Aggarwal A, et al: Cerebral injury after cardiac surgery: identification of a group at extraordinary risk. Stroke. 1999, 30: 514-522.

    Article  CAS  PubMed  Google Scholar 

  17. Sendelbach S, Lindquist R, Watanuki S, Savik K: Correlates of neurocognitive function of patients after off-pump coronary artery bypass surgery. American Journal of Critical Care. 2006, 15: 290-298.

    PubMed  Google Scholar 

  18. Selnes OA, Goldsborough MA, Borowicz LM, Enger C, Quaskey SA, McKhann GM: Determinants of cognitive change after coronary artery bypass surgery: a multifactorial problem. Annals of Thoracic Surgery. 1999, 67: 1669-1676. 10.1016/S0003-4975(99)00258-1.

    Article  CAS  PubMed  Google Scholar 

  19. Selnes OA, Goldsborough MA, Borowicz LM, McKhann GM: Neurobehavioural sequelae of cardiopulmonary bypass. Lancet. 1999, 8;353: 1601-1606. 10.1016/S0140-6736(98)07576-X.

    Article  Google Scholar 

  20. Borowicz LM, Goldsborough MA, Selnes OA, McKhann GM: Neuropsychologic change after cardiac surgery: a critical review. Journal Cardiothoracic and Vascular Anesthesia. 1996, 10: 105-11. 10.1016/S1053-0770(96)80185-6.

    Article  CAS  Google Scholar 

  21. Newman MF, Croughwell ND, Blumenthal JA, et al: Effect of aging on cerebral autoregulation during cardiopulmonary bypass. Association with postoperative cognitive dysfunction. Circulation. 1994, 90: II243-II249.

    CAS  PubMed  Google Scholar 

  22. Lang EW, Chesnut RM: Intracranial pressure and cerebral perfusion pressure in severe head injury. New Horizons. 1995, 3: 400-409.

    CAS  PubMed  Google Scholar 

  23. Aaslid R, Markwalder TM, Nornes H: Noninvasive transcranial Doppler ultrasound recording of flow velocity in basal cerebral arteries. Journal of Neurosurgery. 1982, 57: 769-774. 10.3171/jns.1982.57.6.0769.

    Article  CAS  PubMed  Google Scholar 

  24. Lindegaard KF, Lundar T, Wiberg J, Sjoberg D, Aaslid R, Nornes H: Variations in middle cerebral artery blood flow investigated with noninvasive transcranial blood velocity measurements. Stroke. 1987, 18: 1025-1030.

    Article  CAS  PubMed  Google Scholar 

  25. Shen Q, Stuart J, Venkatesh B, Wallace J, Lipman J: Inter observer variability of the transcranial Doppler ultrasound technique: impact of lack of practice on the accuracy of measurement. Journal of Clinical Monitoring and Computing. 1999, 15: 179-184. 10.1023/A:1009925811965.

    Article  CAS  PubMed  Google Scholar 

  26. Christ M, Noack F, Schroeder T, et al: Continuous cerebral autoregulation monitoring by improved cross-correlation analysis: comparison with the cuff deflation test. Intensive Care Medicine. 2007, 33: 246-254. 10.1007/s00134-006-0451-8.

    Article  PubMed  Google Scholar 

  27. Steinmeier R, Bauhuf C, Hubner U, Hofmann RP, Fahlbusch R: Continuous cerebral autoregulation monitoring by cross-correlation analysis. Journal Of Neurotrauma. 2002, 19: 1127-138. 10.1089/08977150260337949.

    Article  PubMed  Google Scholar 

  28. Barnett A: A time-domain test for some types of non-linearity. Wolff R., editor. IEEE Transactions on Signal Processing. 2008, 53: 26-33. 10.1109/TSP.2004.838942.

    Article  Google Scholar 

  29. Dobson AJ, Barnett AG: An Introduction to Generalized Linear Models. Texts in Statistical Science. 2008, Boca Raton, FL: Chapman & Hall/CRC, 3

    Google Scholar 

  30. Applebaum RM, Wun HH, Katz ES, Tunick PA, Kronzon I: Effects of intraaortic balloon counterpulsation on carotid artery blood flow. American HeartJournal. 1998, 135: 850-4. 10.1016/S0002-8703(98)70045-6.

    Article  CAS  Google Scholar 

  31. Daffertshofer M, Hennerici M: Cerebrovascular regulation and vasoneuronal coupling. Journal of Clinical Ultrasound. 1995, 23: 125-138. 10.1002/jcu.1870230207.

    Article  CAS  PubMed  Google Scholar 

  32. Giller CA, Lam M, Roseland A: Periodic variations in transcranial Doppler mean velocities. Journal of Neuroimaging. 1993, 3: 160-162.

    Article  CAS  PubMed  Google Scholar 

  33. Fujii K, Heistad DD, Faraci FM: Vasomotion of basilar arteries in vivo. American Journal of Physiology. 1990, 258: H1829-H1834.

    CAS  PubMed  Google Scholar 

  34. Zernikow B, Michel E, Kohlmann G, Steck J, Schmitt RM, Jorch G: Cerebral autoregulation of preterm neonates--a non-linear control system?. Archives of Disease in Childhood, Fetal and Neonatal. 1994, 70: F166-F173. 10.1136/fn.70.3.F166.

    Article  CAS  Google Scholar 

  35. Bigger JT, Fleiss JL, Rolnitzky LM, Steinman RC: The ability of several short-term measures of RR variability to predict mortality after myocardial infarction. Circulation. 1993, 88: 927-934.

    Article  CAS  PubMed  Google Scholar 

  36. Bigger JT, Fleiss JL, Steinman RC, Rolnitzky LM, Schneider WJ, Stein PK: RR variability in healthy, middle-aged persons compared with patients with chronic coronary heart disease or recent acute myocardial infarction. Circulation. 1995, 1 (91): 1936-1943.

    Article  Google Scholar 

  37. West BJ: Multifractality of cerebral blod flow. Physica A. 2003, 318: 453-460. 10.1016/S0378-4371(02)01377-8.

    Article  Google Scholar 

  38. Steinmeier R, Hofmann RP, Bauhuf C, Hubner U, Fahlbusch R: Continuous cerebral autoregulation monitoring by cross-correlation analysis. Journal of Neurotrauma. 2002, 19: 1127-1138. 10.1089/08977150260337949.

    Article  PubMed  Google Scholar 

  39. Marion DW, Darby J, Yonas H: Acute regional cerebral blood flow changes caused by severe head injuries. Journal of Neurosurgery. 1991, 74: 407-414. 10.3171/jns.1991.74.3.0407.

    Article  CAS  PubMed  Google Scholar 

  40. Lorenz MW, Gonzalez M, Lienerth C, Loesel N, Thoelen N, Sitzer M: Influence of temporal insonation window quality on the assessment of cerebral autoregulation with transcranial Doppler sonography. Ultrasound in Medicine and Biology. 2007, 33: 1540-1545. 10.1016/j.ultrasmedbio.2007.04.010.

    Article  PubMed  Google Scholar 

  41. Lang EW, Diehl RR, Timmermann L, et al: Spontaneous oscillations of arterial blood pressure, cerebral and peripheral blood flow in healthy and comatose subjects. Neurology Research. 1999, 21: 665-669.

    CAS  Google Scholar 

  42. Chiu CC, Yeh SJ: Assessment of cerebral autoregulation using time-domain cross-correlation analysis. Computers in Biology and Medicine. 2001, 31: 471-80. 10.1016/S0010-4825(01)00015-4.

    Article  CAS  PubMed  Google Scholar 

Pre-publication history

Download references

Acknowledgements

We thank Dr Barbara Lingwood for facilitating the use of the Transcranial Doppler device and Dr Matthew Pincus for his review and critique of our manuscript. We thank the High Performance Computing and Research Support Unit, Queensland University of Technology, Brisbane, Australia, for facilitating computational resources and services. We express our most sincere thanks to all the nursing staff at the Intensive Care Units of Prince Charles Hospital and Royal Brisbane Hospital. We thank Dr Lluis Marruecos for all the teaching provided on the use of Transcranial Doppler. This research was supported by a research grant from the Royal Brisbane Hospital Research Foundation (Protocol 2007/076). Additional funding was provided by the Prince Charles Hospital Research Foundation.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Judith Bellapart.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

JB and JF conceived and designed the study; JB and SG undertook patient screen and data acquisition. JB drafted the manuscript which was then reviewed and amended by JF and RB. RB was involved in extensive editing. AB performed all data analysis and supervised the data interpretation. DT facilitated and advised about the use of the powerlab technology. KD resolved all technical problems related to data acquisition. All authors read and approved the final manuscript.

Authors’ original submitted files for images

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Bellapart, J., Geng, S., Dunster, K. et al. Intraaortic Balloon Pump Counterpulsation and Cerebral Autoregulation: an observational study. BMC Anesthesiol 10, 3 (2010). https://doi.org/10.1186/1471-2253-10-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1471-2253-10-3

Keywords